Felipe G Belladonna, Victor T L Vieira, Marco Simões-Carvalho, Carlos Garcia Puentes, Alejandro Jaime, Emmanuel J N L Silva
{"title":"Deformation and Fracture of NiTi Instruments Using Adaptive Reciprocation Intelligent or Fixed-Angle Reciprocation Settings.","authors":"Felipe G Belladonna, Victor T L Vieira, Marco Simões-Carvalho, Carlos Garcia Puentes, Alejandro Jaime, Emmanuel J N L Silva","doi":"10.1016/j.joen.2026.10.004","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>To compare the canal number at first visible deformation, the number of simulated canals completely prepared before fracture, and fractured-fragment length of One RECI instruments operated using fixed-angle conventional reciprocation or the Adaptive Reciprocation Intelligent (ARI) setting.</p><p><strong>Methods: </strong>Twenty-six One RECI instruments were allocated to conventional reciprocation or ARI groups (n = 13). Fixed-angle reciprocation used parameters of 170° counterclockwise/60° clockwise, 4.0 Ncm, and 400 rpm. ARI initially used the same parameters but automatically changed to 90° counterclockwise/30° clockwise, 0.4 Ncm and 500 rpm when resistance to instrument advancement was detected. A single experienced operator repeatedly prepared simulated canals in transparent resin blocks until instrument fracture. Each canal was prepared using successive sequences of three gentle in-and-out pecking movements, with instrument cleaning and inspection, canal irrigation with 5 mL of distilled water, and recapitulation with a size 15 K-file between sequences, until the working length was reached. The canal number at first visible deformation and the number of canals completely prepared before fracture were recorded. Fractured-fragment length was measured digitally from the instrument tip to the fracture line. Groups were compared using two-sided Mann-Whitney U tests (α = 0.05).</p><p><strong>Results: </strong>First visible deformation was detected at a significantly higher simulated canal number with ARI than with conventional reciprocation (median [interquartile range], 3 [1-5] versus 1 [1-2]; P = 0.037). ARI also increased the number of canals completely prepared before fracture (20 [18-30] versus 6 [5-11]; P < 0.001). Fractured-fragment length did not differ significantly between ARI and conventional reciprocation (14.15 [11.65-14.43] mm versus 13.77 [13.61-14.51] mm; P = 0.644).</p><p><strong>Conclusions: </strong>The ARI setting delayed visible deformation and markedly increased the number of simulated canals prepared before fracture of One RECI instruments, without significantly affecting fractured-fragment length.</p>","PeriodicalId":15703,"journal":{"name":"Journal of endodontics","volume":" ","pages":""},"PeriodicalIF":3.9000,"publicationDate":"2026-10-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of endodontics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.joen.2026.10.004","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: To compare the canal number at first visible deformation, the number of simulated canals completely prepared before fracture, and fractured-fragment length of One RECI instruments operated using fixed-angle conventional reciprocation or the Adaptive Reciprocation Intelligent (ARI) setting.
Methods: Twenty-six One RECI instruments were allocated to conventional reciprocation or ARI groups (n = 13). Fixed-angle reciprocation used parameters of 170° counterclockwise/60° clockwise, 4.0 Ncm, and 400 rpm. ARI initially used the same parameters but automatically changed to 90° counterclockwise/30° clockwise, 0.4 Ncm and 500 rpm when resistance to instrument advancement was detected. A single experienced operator repeatedly prepared simulated canals in transparent resin blocks until instrument fracture. Each canal was prepared using successive sequences of three gentle in-and-out pecking movements, with instrument cleaning and inspection, canal irrigation with 5 mL of distilled water, and recapitulation with a size 15 K-file between sequences, until the working length was reached. The canal number at first visible deformation and the number of canals completely prepared before fracture were recorded. Fractured-fragment length was measured digitally from the instrument tip to the fracture line. Groups were compared using two-sided Mann-Whitney U tests (α = 0.05).
Results: First visible deformation was detected at a significantly higher simulated canal number with ARI than with conventional reciprocation (median [interquartile range], 3 [1-5] versus 1 [1-2]; P = 0.037). ARI also increased the number of canals completely prepared before fracture (20 [18-30] versus 6 [5-11]; P < 0.001). Fractured-fragment length did not differ significantly between ARI and conventional reciprocation (14.15 [11.65-14.43] mm versus 13.77 [13.61-14.51] mm; P = 0.644).
Conclusions: The ARI setting delayed visible deformation and markedly increased the number of simulated canals prepared before fracture of One RECI instruments, without significantly affecting fractured-fragment length.
期刊介绍:
The Journal of Endodontics, the official journal of the American Association of Endodontists, publishes scientific articles, case reports and comparison studies evaluating materials and methods of pulp conservation and endodontic treatment. Endodontists and general dentists can learn about new concepts in root canal treatment and the latest advances in techniques and instrumentation in the one journal that helps them keep pace with rapid changes in this field.