Comparative Finite Element Analysis of Three Acetabular Plate Designs with Different Bridge Geometries.

IF 2 Q2 ORTHOPEDICS
Orthopedic Research and Reviews Pub Date : 2026-10-06 eCollection Date: 2026-01-01 DOI:10.2147/ORR.S613671
Nurgeldi Manap, Nagmet Mursalov, Mehool Acharya, Boris Gorbunov, Mukhtar Abilmazhinov, Darkhan Sultanov
{"title":"Comparative Finite Element Analysis of Three Acetabular Plate Designs with Different Bridge Geometries.","authors":"Nurgeldi Manap, Nagmet Mursalov, Mehool Acharya, Boris Gorbunov, Mukhtar Abilmazhinov, Darkhan Sultanov","doi":"10.2147/ORR.S613671","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Acetabular fractures are complex injuries of the hip joint. Careful design of fixation plates is crucial for achieving stable osteosynthesis, especially in patients with poor bone quality. Finite element analysis (FEA) is widely used for biomechanical evaluation of orthopedic implants.</p><p><strong>Objective: </strong>The aim of this study was to perform a comparative finite element analysis of three acetabular plate designs differing in bridge geometry and to identify the configuration with the most favourable biomechanical performance. We hypothesized that a plate with two parallel bridges would provide more favourable stress distribution and a higher safety factor compared to single-bridge and non-parallel-bridge designs.</p><p><strong>Methods: </strong>A three-dimensional model of the pelvic bone was reconstructed from CT data. Three titanium alloy (Ti-6Al-4V, thickness 2.5 mm) plates with different bridge configurations (single bridge, two parallel bridges, two non-parallel bridges) were virtually implanted using CAD software (KOMPAS-3D, Autodesk Inventor). The FEA model included cortical and cancellous bone with mechanical properties taken from the literature. A load of 250 N applied at 15° to the inner plate surface simulated partial weight-bearing in the early postoperative period. Von Mises equivalent stress, linear displacement, and safety factor were evaluated.</p><p><strong>Results: </strong>Maximum von Mises equivalent stresses were 207 MPa for the single-bridge plate, 170 MPa for the two-parallel-bridge plate, and 344 MPa for the two-non-parallel-bridge plate. The corresponding minimum safety factors were 1.33, 1.62, and 0.8. The obtained stress values are consistent with findings from similar studies.</p><p><strong>Conclusion: </strong>The plate with two parallel bridges exhibited the highest safety factor and appears to be the most promising candidate for further experimental evaluation based on the present computational findings. The single-bridge and two-non-parallel-bridge designs require further design modification, particularly in the bridge cross-sections, because of their lower static safety margins. FEA is a useful tool for comparative biomechanical assessment of implant designs.</p>","PeriodicalId":19608,"journal":{"name":"Orthopedic Research and Reviews","volume":"18 ","pages":"613671"},"PeriodicalIF":2.0000,"publicationDate":"2026-10-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13653554/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Orthopedic Research and Reviews","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/ORR.S613671","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Acetabular fractures are complex injuries of the hip joint. Careful design of fixation plates is crucial for achieving stable osteosynthesis, especially in patients with poor bone quality. Finite element analysis (FEA) is widely used for biomechanical evaluation of orthopedic implants.

Objective: The aim of this study was to perform a comparative finite element analysis of three acetabular plate designs differing in bridge geometry and to identify the configuration with the most favourable biomechanical performance. We hypothesized that a plate with two parallel bridges would provide more favourable stress distribution and a higher safety factor compared to single-bridge and non-parallel-bridge designs.

Methods: A three-dimensional model of the pelvic bone was reconstructed from CT data. Three titanium alloy (Ti-6Al-4V, thickness 2.5 mm) plates with different bridge configurations (single bridge, two parallel bridges, two non-parallel bridges) were virtually implanted using CAD software (KOMPAS-3D, Autodesk Inventor). The FEA model included cortical and cancellous bone with mechanical properties taken from the literature. A load of 250 N applied at 15° to the inner plate surface simulated partial weight-bearing in the early postoperative period. Von Mises equivalent stress, linear displacement, and safety factor were evaluated.

Results: Maximum von Mises equivalent stresses were 207 MPa for the single-bridge plate, 170 MPa for the two-parallel-bridge plate, and 344 MPa for the two-non-parallel-bridge plate. The corresponding minimum safety factors were 1.33, 1.62, and 0.8. The obtained stress values are consistent with findings from similar studies.

Conclusion: The plate with two parallel bridges exhibited the highest safety factor and appears to be the most promising candidate for further experimental evaluation based on the present computational findings. The single-bridge and two-non-parallel-bridge designs require further design modification, particularly in the bridge cross-sections, because of their lower static safety margins. FEA is a useful tool for comparative biomechanical assessment of implant designs.

三种不同桥梁几何形状髋臼板设计的有限元对比分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Orthopedic Research and Reviews
Orthopedic Research and Reviews Medicine-Orthopedics and Sports Medicine
CiteScore
2.80
自引率
0.00%
发文量
51
审稿时长
16 weeks
期刊介绍: Orthopedic Research and Reviews is an international, peer-reviewed, open-access journal focusing on the patho-physiology of the musculoskeletal system, trauma, surgery and other corrective interventions to restore mobility and function. Advances in new technologies, materials, techniques and pharmacological agents will be particularly welcome. Specific topics covered in the journal include: Patho-physiology and bioengineering, Technologies and materials science, Surgical techniques, including robotics, Trauma management and care, Treatment including pharmacological and non-pharmacological, Rehabilitation and Multidisciplinarian care approaches, Patient quality of life, satisfaction and preference, Health economic evaluations. The journal welcomes submitted papers covering original research, basic science and technology, clinical studies, reviews and evaluations, guidelines, expert opinion and commentary, case reports and extended reports.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书