{"title":"Measuring adult breathing pattern: devices, outcomes, and evidence-a scoping review.","authors":"Rodrigo Martins, Fátima Rodrigues, Ana Martins, Bruno Neves, Helena Silva, João Guerra, Susana Costa, Nélson Costa","doi":"10.1007/s11325-026-03834-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To map instruments and protocols used to assess adult breathing pattern (BP), describe the outcomes they yield, appraise reported measurement properties, and identify whether any single modality is \"clinically ready.\" The review asks which instruments and protocols quantify adult BP, which outcomes are reported (e.g., rate, amplitude, I:E ratio, thoracoabdominal mobility), and what evidence exists for reliability, sensitivity, or specificity.</p><p><strong>Methods: </strong>A scoping review, guided by Joanna Briggs Institute methods and reported per PRISMA-ScR, was undertaken. Searches in Scopus, Web of Science, and PubMed were run to 27 Aug 2024 and updated on 21 Aug 2025 with identical strategies. Five reviewers screened records in three stages against predefined eligibility criteria (adult human samples; instrumented BP assessment). Data were charted for population, device and sampling, reference standards, tasks, outcomes, and clinimetric reporting. Thirty-one studies met eligibility.</p><p><strong>Results: </strong>Evidence (1994-2025; 31 studies) clusters into contact and contactless families: optical/vision and motion capture, respiratory inductance plethysmography and other belts/pressure arrays, wearable inertial sensors, radar, and depth/thermal cameras. Outcomes concentrate on respiratory rate, tidal/volume proxies, and thoraco-abdominal contributions/coordination; fewer studies report event timing, displacement, or phase/asynchrony. Validation is chiefly against spirometry/pneumotachography; agreement for respiratory rate is generally high, while absolute volume agreement varies with calibration and posture. Reliability indices and diagnostic metrics are inconsistently reported. Most samples are small and posture-bound (laboratory or seated/supine), with heterogeneous preprocessing and reference-frame conventions limiting comparability.</p><p><strong>Conclusion: </strong>Multiple modalities can quantify BP, but no single instrument currently satisfies validity, reliability, responsiveness, interpretability, feasibility, and equity across settings. Near-term use is strongest for respiratory-rate and trend-level expansion under controlled conditions. Progress depends on standardized reference frames for wearables, transparent breath-cycle definitions, cross-instrument calibration against common protocols, and larger real-world studies with systematic reporting.</p>","PeriodicalId":520777,"journal":{"name":"Sleep & breathing = Schlaf & Atmung","volume":"30 6","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-10-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sleep & breathing = Schlaf & Atmung","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s11325-026-03834-5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To map instruments and protocols used to assess adult breathing pattern (BP), describe the outcomes they yield, appraise reported measurement properties, and identify whether any single modality is "clinically ready." The review asks which instruments and protocols quantify adult BP, which outcomes are reported (e.g., rate, amplitude, I:E ratio, thoracoabdominal mobility), and what evidence exists for reliability, sensitivity, or specificity.
Methods: A scoping review, guided by Joanna Briggs Institute methods and reported per PRISMA-ScR, was undertaken. Searches in Scopus, Web of Science, and PubMed were run to 27 Aug 2024 and updated on 21 Aug 2025 with identical strategies. Five reviewers screened records in three stages against predefined eligibility criteria (adult human samples; instrumented BP assessment). Data were charted for population, device and sampling, reference standards, tasks, outcomes, and clinimetric reporting. Thirty-one studies met eligibility.
Results: Evidence (1994-2025; 31 studies) clusters into contact and contactless families: optical/vision and motion capture, respiratory inductance plethysmography and other belts/pressure arrays, wearable inertial sensors, radar, and depth/thermal cameras. Outcomes concentrate on respiratory rate, tidal/volume proxies, and thoraco-abdominal contributions/coordination; fewer studies report event timing, displacement, or phase/asynchrony. Validation is chiefly against spirometry/pneumotachography; agreement for respiratory rate is generally high, while absolute volume agreement varies with calibration and posture. Reliability indices and diagnostic metrics are inconsistently reported. Most samples are small and posture-bound (laboratory or seated/supine), with heterogeneous preprocessing and reference-frame conventions limiting comparability.
Conclusion: Multiple modalities can quantify BP, but no single instrument currently satisfies validity, reliability, responsiveness, interpretability, feasibility, and equity across settings. Near-term use is strongest for respiratory-rate and trend-level expansion under controlled conditions. Progress depends on standardized reference frames for wearables, transparent breath-cycle definitions, cross-instrument calibration against common protocols, and larger real-world studies with systematic reporting.
目的:绘制用于评估成人呼吸模式(BP)的仪器和方案,描述它们产生的结果,评估报告的测量特性,并确定是否有任何单一模式“临床就绪”。该综述询问了哪些仪器和方案可以量化成人血压,报告了哪些结果(例如,发病率、幅度、I:E比、胸腹活动度),以及存在哪些可靠性、敏感性或特异性的证据。方法:在Joanna Briggs研究所的方法指导下,根据PRISMA-ScR报告进行范围审查。Scopus, Web of Science和PubMed的搜索运行到2024年8月27日,并以相同的策略更新到2025年8月21日。五名审稿人根据预定义的资格标准(成人样本;仪器血压评估)分三个阶段筛选记录。将总体、设备和抽样、参考标准、任务、结果和临床计量报告的数据绘制成图表。31项研究符合资格。结果:证据(1994-2025;31项研究)分为接触式和非接触式两大类:光学/视觉和运动捕捉、呼吸感应体积脉搏仪和其他带式/压力阵列、可穿戴惯性传感器、雷达和深度/热像仪。结果集中于呼吸频率、潮汐/容积指标和胸腹贡献/协调;较少的研究报告事件时间、位移或相位/异步。验证主要针对肺活量测定法/气相造影术;呼吸频率的一致性一般较高,而绝对容积的一致性因校准和姿势而异。可靠性指标和诊断指标报告不一致。大多数样本都很小且有姿势限制(实验室或坐着/仰卧),不同的预处理和参考框架惯例限制了可比性。结论:多种方法可以量化BP,但目前没有一种方法能够满足有效性、可靠性、响应性、可解释性、可行性和公平性。在受控条件下,短期使用对呼吸速率和趋势水平扩张最有效。进展取决于可穿戴设备的标准化参考框架、透明的呼吸周期定义、针对通用协议的跨仪器校准以及更大规模的真实世界研究和系统报告。