Xiali Gao, Zhichao Zheng, Zhibo Xiao, Xu Sun, Nuomin Zhang, Yang Xiao, Hao Huang, Xuran Zhou, Fan Meng, Xuanhao Wang, Ruimin Chen, Junhui Shi
{"title":"Phase-aberration-corrected coherence factor beamforming for photoacoustic computed tomography.","authors":"Xiali Gao, Zhichao Zheng, Zhibo Xiao, Xu Sun, Nuomin Zhang, Yang Xiao, Hao Huang, Xuran Zhou, Fan Meng, Xuanhao Wang, Ruimin Chen, Junhui Shi","doi":"10.1117/1.JBO.31.10.106003","DOIUrl":null,"url":null,"abstract":"<p><strong>Significance: </strong>Coherence factor (CF) beamforming can suppress sidelobes in photoacoustic computed tomography, but phase aberration can make genuine absorbers appear incoherent and cause conventional CF to downweight real signals.</p><p><strong>Aim: </strong>We aim to reduce photoacoustic volume tomography phase-aberration-induced loss of true signals in conventional coherence-factor reconstruction and validate the method using phantom and <i>in vivo</i> experiments.</p><p><strong>Approach: </strong>PAC-CF extends an ellipsoidal two-region time-of-flight correction to the angle-weighted channel contributions used for both coherent backprojection and final conventional energy-ratio CF evaluation. The ROI-optimized effective internal sound speed is selected from multiframe PA receive data by maximizing the mean CF within a fixed region of interest.</p><p><strong>Results: </strong>Simulations showed that conventional CF at the true source dropped to 0.00794 at a <math><mrow><mn>170</mn> <mtext> </mtext> <mi>m</mi> <mo>/</mo> <mi>s</mi></mrow> </math> sound-speed difference and to <math><mrow><mo>∼</mo> <mn>0.00039</mn></mrow> </math> to 0.00042 at larger sound-speed differences, whereas PAC-CF preserved coherence and localization. The four-method phantom ablation showed that combining two-region TOF correction with final CF weighting produced the most concentrated wire responses and strongest background suppression, whereas mouse experiments showed recovery of vascular features suppressed by conventional CF.</p><p><strong>Conclusions: </strong>By extending two-region time-of-flight correction to the channel contributions evaluated by the final CF weight, PAC-CF reduces the erroneous suppression of true signals caused by propagation-model mismatch and provides a practical PA-only approach for coherence-weighted three-dimensional PACT.</p>","PeriodicalId":15264,"journal":{"name":"Journal of Biomedical Optics","volume":"31 10","pages":"106003"},"PeriodicalIF":3.3000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13652113/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Optics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JBO.31.10.106003","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/10/9 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Significance: Coherence factor (CF) beamforming can suppress sidelobes in photoacoustic computed tomography, but phase aberration can make genuine absorbers appear incoherent and cause conventional CF to downweight real signals.
Aim: We aim to reduce photoacoustic volume tomography phase-aberration-induced loss of true signals in conventional coherence-factor reconstruction and validate the method using phantom and in vivo experiments.
Approach: PAC-CF extends an ellipsoidal two-region time-of-flight correction to the angle-weighted channel contributions used for both coherent backprojection and final conventional energy-ratio CF evaluation. The ROI-optimized effective internal sound speed is selected from multiframe PA receive data by maximizing the mean CF within a fixed region of interest.
Results: Simulations showed that conventional CF at the true source dropped to 0.00794 at a sound-speed difference and to to 0.00042 at larger sound-speed differences, whereas PAC-CF preserved coherence and localization. The four-method phantom ablation showed that combining two-region TOF correction with final CF weighting produced the most concentrated wire responses and strongest background suppression, whereas mouse experiments showed recovery of vascular features suppressed by conventional CF.
Conclusions: By extending two-region time-of-flight correction to the channel contributions evaluated by the final CF weight, PAC-CF reduces the erroneous suppression of true signals caused by propagation-model mismatch and provides a practical PA-only approach for coherence-weighted three-dimensional PACT.