Updated systematic review and meta-analysis comparing Myval versus Sapien 3 transcatheter heart valves in severe aortic stenosis: focus on early and one-year mortality.
{"title":"Updated systematic review and meta-analysis comparing Myval versus Sapien 3 transcatheter heart valves in severe aortic stenosis: focus on early and one-year mortality.","authors":"Lin Yang, Guisong Yu, Mingchen Pan, Yandong Liu, Yuxi Zhu, Mengyu Zhang, Haigen Yu, Zhaohui Pei","doi":"10.1080/00015385.2026.2742689","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Transcatheter aortic valve implantation (TAVI) is a guideline-recommended treatment for severe aortic stenosis (AS). While the balloon-expandable Edwards Sapien 3 valve has long been the benchmark, the newer Myval valve has emerged as a potential alternative. Comparative data on mid-term survival, specifically one-year all-cause mortality, are lacking but crucial for clinical evaluation.</p><p><strong>Methods: </strong>This systematic review and meta-analysis synthesised direct comparative evidence on the Myval versus Sapien 3 valves in patients undergoing TAVI, searching PubMed, Embase, and Cochrane Central up to November 2025. The primary outcomes included all-cause mortality within 30 days and 1 year. Secondary outcomes were 30-day safety events and hemodynamic parameters.</p><p><strong>Results: </strong>Five studies (<i>n</i> = 2771) were included. The pooled analysis indicated no statistical difference in 30-day mortality (RR: 1.05, 95% CI: 0.53-2.07; <i>p</i> = 0.89; I<sup>2</sup>=0%) or in one-year all-cause mortality (RR: 0.93, 95% CI: 0.64-1.35; <i>p</i> = 0.70; I<sup>2</sup>=0%). Rates of stroke, major bleeding, pacemaker implantation, and other 30-day complications were also comparable (all <i>p</i> > 0.05). Hemodynamically, the Myval valve was superior at 30 days, with a lower mean gradient (MD: -2.37, 95% CI: -3.45 to -1.30; <i>p</i> < 0.0001) and larger effective orifice area (MD: 0.26, 95% CI: 0.17-0.35; <i>p</i> < 0.00001). Available one-year data from two randomised trials indicated this hemodynamic benefit was sustained.</p><p><strong>Conclusion: </strong>This updated meta-analysis demonstrates that the newer Myval valve offers comparable early safety and one-year survival to the established Sapien 3 valve, with superior and sustained hemodynamic performance. These findings support its role as a safe and effective alternative in the TAVI landscape.</p>","PeriodicalId":6979,"journal":{"name":"Acta cardiologica","volume":" ","pages":"1-12"},"PeriodicalIF":4.0000,"publicationDate":"2026-10-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta cardiologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00015385.2026.2742689","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Transcatheter aortic valve implantation (TAVI) is a guideline-recommended treatment for severe aortic stenosis (AS). While the balloon-expandable Edwards Sapien 3 valve has long been the benchmark, the newer Myval valve has emerged as a potential alternative. Comparative data on mid-term survival, specifically one-year all-cause mortality, are lacking but crucial for clinical evaluation.
Methods: This systematic review and meta-analysis synthesised direct comparative evidence on the Myval versus Sapien 3 valves in patients undergoing TAVI, searching PubMed, Embase, and Cochrane Central up to November 2025. The primary outcomes included all-cause mortality within 30 days and 1 year. Secondary outcomes were 30-day safety events and hemodynamic parameters.
Results: Five studies (n = 2771) were included. The pooled analysis indicated no statistical difference in 30-day mortality (RR: 1.05, 95% CI: 0.53-2.07; p = 0.89; I2=0%) or in one-year all-cause mortality (RR: 0.93, 95% CI: 0.64-1.35; p = 0.70; I2=0%). Rates of stroke, major bleeding, pacemaker implantation, and other 30-day complications were also comparable (all p > 0.05). Hemodynamically, the Myval valve was superior at 30 days, with a lower mean gradient (MD: -2.37, 95% CI: -3.45 to -1.30; p < 0.0001) and larger effective orifice area (MD: 0.26, 95% CI: 0.17-0.35; p < 0.00001). Available one-year data from two randomised trials indicated this hemodynamic benefit was sustained.
Conclusion: This updated meta-analysis demonstrates that the newer Myval valve offers comparable early safety and one-year survival to the established Sapien 3 valve, with superior and sustained hemodynamic performance. These findings support its role as a safe and effective alternative in the TAVI landscape.