Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological Endpoints
IF 2.7 2区 环境科学与生态学Q1 CONSTRUCTION & BUILDING TECHNOLOGY
Jessica Morel, Jean-Michel Lenoir, Camille Ménard, Samuel Le Hir, Clara Bourgeay, Tiago Zambujo, Emma Ensenat, Ibtissam Barhoumi, Mariana Gaboriau, Thomas Julien, Anaïs Proust, Manuel Rosa-Calatrava
{"title":"Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological Endpoints","authors":"Jessica Morel, Jean-Michel Lenoir, Camille Ménard, Samuel Le Hir, Clara Bourgeay, Tiago Zambujo, Emma Ensenat, Ibtissam Barhoumi, Mariana Gaboriau, Thomas Julien, Anaïs Proust, Manuel Rosa-Calatrava","doi":"10.1155/ina/2773156","DOIUrl":null,"url":null,"abstract":"<p>Indoor air quality (IAQ) monitoring in hospital settings is critical to limit airborne transmission of pathogens to vulnerable patients and healthcare workers. However, bioaerosol sampling remains challenging because of the diversity of devices and methodologies and the very low airborne concentration of pathogens in real-world settings, especially when infectivity-based measurements, are required for assessing transmission risk. In the absence of harmonized evaluation procedures, it remains difficult to select sampling strategies that are both reliable and fit for purpose. In this study, we developed a controlled and reproducible experimental framework for the evaluation of bioaerosol samplers under conditions relevant to hospital air monitoring. Two widely used cyclonic bioaerosol samplers, Coriolis+ and Coriolis Compact, were assessed in a calibrated 20 m<sup>3</sup> test chamber generating stable atmospheres containing infectious SARS-CoV-2, <i>Staphylococcus aureus</i>, or <i>Aspergillus fumigatus</i> at different concentration levels. Samplers′ performance was evaluated using complementary molecular and viability-based endpoints and was combined with computational fluid dynamics (CFD) and experimental airflow measurements to characterize airflow dynamics. Our results demonstrated that samplers′ performance strongly depended on the biological endpoint and microorganism considered. For SARS-CoV-2, the two devices showed comparable recovery of infectious virus, at intermediate and high concentrations, but neither detected infectious particles at the lowest level. In contrast, Coriolis+ achieved near-complete recovery of viral genomes, whereas Coriolis Compact recovered only a fraction of the expected signal. Regarding <i>S. aureus</i> and <i>A. fumigatus</i>, Coriolis+ achieved superior recovery of viable microorganisms, whereas Coriolis Compact generally yielded higher genomic recovery at moderate and high concentrations. CFD analyses further showed that local airflow conditions and sampler-specific flow patterns, particularly for the higher flow Coriolis+, are critical for interpreting biological recovery and for defining appropriate placement and operating modes. Overall, this study demonstrates that bioaerosol sampler evaluation must jointly consider molecular, microbiological, and aerodynamic dimensions. The proposed framework offers a basis for more standardized and reproducible performance assessment of bioaerosol samplers and supports evidence-based selection of air monitoring strategies in healthcare environments.</p>","PeriodicalId":13529,"journal":{"name":"Indoor air","volume":"2026 1","pages":""},"PeriodicalIF":2.7000,"publicationDate":"2026-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1155/ina/2773156","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indoor air","FirstCategoryId":"93","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1155/ina/2773156","RegionNum":2,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CONSTRUCTION & BUILDING TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Indoor air quality (IAQ) monitoring in hospital settings is critical to limit airborne transmission of pathogens to vulnerable patients and healthcare workers. However, bioaerosol sampling remains challenging because of the diversity of devices and methodologies and the very low airborne concentration of pathogens in real-world settings, especially when infectivity-based measurements, are required for assessing transmission risk. In the absence of harmonized evaluation procedures, it remains difficult to select sampling strategies that are both reliable and fit for purpose. In this study, we developed a controlled and reproducible experimental framework for the evaluation of bioaerosol samplers under conditions relevant to hospital air monitoring. Two widely used cyclonic bioaerosol samplers, Coriolis+ and Coriolis Compact, were assessed in a calibrated 20 m3 test chamber generating stable atmospheres containing infectious SARS-CoV-2, Staphylococcus aureus, or Aspergillus fumigatus at different concentration levels. Samplers′ performance was evaluated using complementary molecular and viability-based endpoints and was combined with computational fluid dynamics (CFD) and experimental airflow measurements to characterize airflow dynamics. Our results demonstrated that samplers′ performance strongly depended on the biological endpoint and microorganism considered. For SARS-CoV-2, the two devices showed comparable recovery of infectious virus, at intermediate and high concentrations, but neither detected infectious particles at the lowest level. In contrast, Coriolis+ achieved near-complete recovery of viral genomes, whereas Coriolis Compact recovered only a fraction of the expected signal. Regarding S. aureus and A. fumigatus, Coriolis+ achieved superior recovery of viable microorganisms, whereas Coriolis Compact generally yielded higher genomic recovery at moderate and high concentrations. CFD analyses further showed that local airflow conditions and sampler-specific flow patterns, particularly for the higher flow Coriolis+, are critical for interpreting biological recovery and for defining appropriate placement and operating modes. Overall, this study demonstrates that bioaerosol sampler evaluation must jointly consider molecular, microbiological, and aerodynamic dimensions. The proposed framework offers a basis for more standardized and reproducible performance assessment of bioaerosol samplers and supports evidence-based selection of air monitoring strategies in healthcare environments.
期刊介绍:
The quality of the environment within buildings is a topic of major importance for public health.
Indoor Air provides a location for reporting original research results in the broad area defined by the indoor environment of non-industrial buildings. An international journal with multidisciplinary content, Indoor Air publishes papers reflecting the broad categories of interest in this field: health effects; thermal comfort; monitoring and modelling; source characterization; ventilation and other environmental control techniques.
The research results present the basic information to allow designers, building owners, and operators to provide a healthy and comfortable environment for building occupants, as well as giving medical practitioners information on how to deal with illnesses related to the indoor environment.