Under pressure- test, test, maybe? A survey of heat exchanger integrity testing at cardiac centres in Australia and New Zealand.

IF 1.3 4区 医学 Q4 CARDIAC & CARDIOVASCULAR SYSTEMS
Martin Gill
{"title":"Under pressure- test, test, maybe? A survey of heat exchanger integrity testing at cardiac centres in Australia and New Zealand.","authors":"Martin Gill","doi":"10.1177/02676591261495416","DOIUrl":null,"url":null,"abstract":"<p><p>IntroductionHeat exchanger (HE) integrity is vital for patient safety during cardiopulmonary bypass (CPB), as failures can lead to severe complications including haemolysis and toxemia. While manufacturers mandate pre-use integrity testing, adherence and methods vary significantly. This study aimed to identify current practices across Australia and New Zealand and assess the mechanical efficacy of manufacturers recommended testing protocols.MethodsA telephone survey was conducted of the most senior perfusionist available at all 80 cardiac surgical hospitals in Australia and New Zealand regarding HE testing customs and known failures. Simultaneously, an in vitro assessment measured the pressures generated by two common heater cooler units (HCUs)- the 3T and HCU40, in conjunction with oxygenator and cardioplegia HEs to evaluate the generated pressures during usage.ResultsNine unverified historical HE failure events were recalled by surveyed senior perfusionists across the region. While 100% of programs were successfully surveyed, only 49% complete perfusion teams consistently perform pre-clinical integrity testing. Of those that test, the most common method (82%) is running water through the HE, as per manufacturers instructions. However, <i>in vitro</i> data revealed that while the HCU40 maintains positive pressure, the 3T HCU generates negative pressure at the HE outlet during use. This negative pressure gradient creates a theoretical risk of masking defects near the outlet during a standard 'water-run' test, as a positive pressure gradient is not uniformly maintained across the HE. Such an occurrence would render the manufacturers \"water-run\" test insufficient for detecting defects near the outlet.ConclusionCurrent HE testing practices are non-uniform and often limited by logistical barriers. Reliance on fluid-based testing is hemodynamically flawed in certain HCU configurations. To enhance adherence and safety, manufacturers should validate gas pressure testing for end users, and perfusion departments should adopt standardised, positive-pressure testing protocols and documented emergency failure algorithms.</p>","PeriodicalId":49707,"journal":{"name":"Perfusion-Uk","volume":" ","pages":"2676591261495416"},"PeriodicalIF":1.3000,"publicationDate":"2026-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Perfusion-Uk","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/02676591261495416","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

IntroductionHeat exchanger (HE) integrity is vital for patient safety during cardiopulmonary bypass (CPB), as failures can lead to severe complications including haemolysis and toxemia. While manufacturers mandate pre-use integrity testing, adherence and methods vary significantly. This study aimed to identify current practices across Australia and New Zealand and assess the mechanical efficacy of manufacturers recommended testing protocols.MethodsA telephone survey was conducted of the most senior perfusionist available at all 80 cardiac surgical hospitals in Australia and New Zealand regarding HE testing customs and known failures. Simultaneously, an in vitro assessment measured the pressures generated by two common heater cooler units (HCUs)- the 3T and HCU40, in conjunction with oxygenator and cardioplegia HEs to evaluate the generated pressures during usage.ResultsNine unverified historical HE failure events were recalled by surveyed senior perfusionists across the region. While 100% of programs were successfully surveyed, only 49% complete perfusion teams consistently perform pre-clinical integrity testing. Of those that test, the most common method (82%) is running water through the HE, as per manufacturers instructions. However, in vitro data revealed that while the HCU40 maintains positive pressure, the 3T HCU generates negative pressure at the HE outlet during use. This negative pressure gradient creates a theoretical risk of masking defects near the outlet during a standard 'water-run' test, as a positive pressure gradient is not uniformly maintained across the HE. Such an occurrence would render the manufacturers "water-run" test insufficient for detecting defects near the outlet.ConclusionCurrent HE testing practices are non-uniform and often limited by logistical barriers. Reliance on fluid-based testing is hemodynamically flawed in certain HCU configurations. To enhance adherence and safety, manufacturers should validate gas pressure testing for end users, and perfusion departments should adopt standardised, positive-pressure testing protocols and documented emergency failure algorithms.

在压力下——测试,测试,也许?澳大利亚和新西兰心脏中心热交换器完整性测试的调查。
在体外循环(CPB)期间,热交换器(HE)的完整性对患者安全至关重要,因为故障可能导致严重的并发症,包括溶血和毒血症。虽然制造商要求使用前进行完整性测试,但依从性和方法差异很大。本研究旨在确定澳大利亚和新西兰目前的做法,并评估制造商推荐的测试方案的机械功效。方法对澳大利亚和新西兰80家心脏外科医院的最资深灌注师进行电话调查,了解HE检测的习惯和已知的失败。同时,一项体外评估测量了两种常见的加热冷却器(hcu)- 3T和HCU40产生的压力,并结合氧合器和心脏停搏器he来评估使用过程中产生的压力。结果被调查的区域内高级灌注医师回忆了9例未经证实的历史HE失败事件。虽然100%的项目成功调查,但只有49%的完整灌注团队坚持进行临床前完整性测试。在这些测试中,最常见的方法(82%)是根据制造商的说明将水通过HE。然而,体外数据显示,HCU40在使用过程中保持正压,3T HCU在HE出口产生负压。在标准的“水运行”测试中,这种负压梯度造成了在出口附近掩盖缺陷的理论风险,因为在整个HE中没有均匀地保持正压梯度。这种情况将使制造商的“水运行”试验不足以检测出口附近的缺陷。结论目前的HE检测实践不统一,且常常受到物流障碍的限制。在某些HCU配置中,依赖基于流体的测试存在血流动力学缺陷。为了提高依从性和安全性,制造商应验证最终用户的气体压力测试,灌注部门应采用标准化的正压测试方案和记录的紧急故障算法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Perfusion-Uk
Perfusion-Uk 医学-外周血管病
CiteScore
3.00
自引率
8.30%
发文量
203
审稿时长
6-12 weeks
期刊介绍: Perfusion is an ISI-ranked, peer-reviewed scholarly journal, which provides current information on all aspects of perfusion, oxygenation and biocompatibility and their use in modern cardiac surgery. The journal is at the forefront of international research and development and presents an appropriately multidisciplinary approach to perfusion science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书