{"title":"Pressure-flow behavior during RIRS irrigation: a controlled ex vivo porcine kidney study of flow-dependent apparent outflow resistance.","authors":"Baorong Feng, Yan Yang, Shoule Wang, Xiaoming Li, Ziteng Shen, Haoyuan Liu, Wen Zhong","doi":"10.1007/s00345-026-06817-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To measure actual irrigation flow and intrarenal pressure (IRP) under different irrigation modes during retrograde intrarenal surgery (RIRS), and to develop a controlled ex vivo empirical pressure-flow model for the tested endoscope-sheath configurations.</p><p><strong>Methods: </strong>A controlled ex vivo porcine kidney-ureter model was used, with all measurements taken using a 200 μm holmium laser fiber in the working channel of a 7.5Fr or 6.3Fr flexible ureteroscope and 10/12Fr or 12/14Fr suction access sheaths. Manual syringe, gravity, and automatic pump irrigation were tested across three porcine kidney-ureter units. Each condition was measured in triplicate (technical repeats); condition-level means were used for model fitting and internal assessment.</p><p><strong>Results: </strong>Actual flow rose with driving pressure and fell with scope deflection. For the ZSR pump, flow increased from 32.0 to 99.8 mL/min (7.5Fr) and from 9.3 to 49.7 mL/min (6.3Fr) across the tested pressures. Pooling pump and syringe data, apparent outflow resistance increased at higher flow, so the non-suction relation was described by [Formula: see text]. Across 20 condition-level means, the complete model gave an RMSE of 1.04 mmHg and R² of 0.994. In a single-operator full-speed syringe test, back-calculated driving pressure was approximately 600 to 770 mmHg.</p><p><strong>Conclusion: </strong>In this controlled ex vivo model, apparent outflow resistance was flow-dependent, and the empirical model described the tested pressure-flow behavior. Forceful syringe flushing should not be routine when controlled pump, water-injection, or adequate gravity irrigation is available. Clinical safety claims require in vivo and clinical validation.</p>","PeriodicalId":23954,"journal":{"name":"World Journal of Urology","volume":"44 1","pages":""},"PeriodicalIF":3.3000,"publicationDate":"2026-10-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"World Journal of Urology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00345-026-06817-z","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To measure actual irrigation flow and intrarenal pressure (IRP) under different irrigation modes during retrograde intrarenal surgery (RIRS), and to develop a controlled ex vivo empirical pressure-flow model for the tested endoscope-sheath configurations.
Methods: A controlled ex vivo porcine kidney-ureter model was used, with all measurements taken using a 200 μm holmium laser fiber in the working channel of a 7.5Fr or 6.3Fr flexible ureteroscope and 10/12Fr or 12/14Fr suction access sheaths. Manual syringe, gravity, and automatic pump irrigation were tested across three porcine kidney-ureter units. Each condition was measured in triplicate (technical repeats); condition-level means were used for model fitting and internal assessment.
Results: Actual flow rose with driving pressure and fell with scope deflection. For the ZSR pump, flow increased from 32.0 to 99.8 mL/min (7.5Fr) and from 9.3 to 49.7 mL/min (6.3Fr) across the tested pressures. Pooling pump and syringe data, apparent outflow resistance increased at higher flow, so the non-suction relation was described by [Formula: see text]. Across 20 condition-level means, the complete model gave an RMSE of 1.04 mmHg and R² of 0.994. In a single-operator full-speed syringe test, back-calculated driving pressure was approximately 600 to 770 mmHg.
Conclusion: In this controlled ex vivo model, apparent outflow resistance was flow-dependent, and the empirical model described the tested pressure-flow behavior. Forceful syringe flushing should not be routine when controlled pump, water-injection, or adequate gravity irrigation is available. Clinical safety claims require in vivo and clinical validation.
期刊介绍:
The WORLD JOURNAL OF UROLOGY conveys regularly the essential results of urological research and their practical and clinical relevance to a broad audience of urologists in research and clinical practice. In order to guarantee a balanced program, articles are published to reflect the developments in all fields of urology on an internationally advanced level. Each issue treats a main topic in review articles of invited international experts. Free papers are unrelated articles to the main topic.