Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.

IF 1 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE
Chung-Yu Lin, Yu-Hsuan Chien, I-Chih Huang
{"title":"Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.","authors":"Chung-Yu Lin, Yu-Hsuan Chien, I-Chih Huang","doi":"10.1097/MBP.0000000000000831","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the clinical accuracy of the AViTA BPM636 oscillometric upper-arm blood pressure monitor in the tested pediatric and adult population, including children aged 5-12 years and adults aged 21-82 years with arm circumferences ranging from 15 to 42 cm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2018/Amd2:2024).</p><p><strong>Methods: </strong>The AViTA BPM636 was evaluated in 85 participants (including 35 pediatric and 50 adult subjects) with upper-arm circumferences ranging from 15 to 42 cm, and two cuffs of different sizes (15 to 22 and 22 to 42 cm) were used in this study. The recruitment fulfilled the requirements of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) for a combined pediatric and adult population using the same arm sequential blood pressure measurement method.</p><p><strong>Results: </strong>For criterion 1, the mean ± SD of the differences between the test device and reference blood pressure measurements was 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP. For criterion 2, the SD of individual mean differences for SBP (5.44 mmHg) and DBP (4.98 mmHg) remained well within the maximum permissible limits of 6.88 and 6.95 mmHg, respectively, thereby satisfying the validation criteria.</p><p><strong>Conclusion: </strong>The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":""},"PeriodicalIF":1.0000,"publicationDate":"2026-10-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000831","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To evaluate the clinical accuracy of the AViTA BPM636 oscillometric upper-arm blood pressure monitor in the tested pediatric and adult population, including children aged 5-12 years and adults aged 21-82 years with arm circumferences ranging from 15 to 42 cm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2018/Amd2:2024).

Methods: The AViTA BPM636 was evaluated in 85 participants (including 35 pediatric and 50 adult subjects) with upper-arm circumferences ranging from 15 to 42 cm, and two cuffs of different sizes (15 to 22 and 22 to 42 cm) were used in this study. The recruitment fulfilled the requirements of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) for a combined pediatric and adult population using the same arm sequential blood pressure measurement method.

Results: For criterion 1, the mean ± SD of the differences between the test device and reference blood pressure measurements was 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP. For criterion 2, the SD of individual mean differences for SBP (5.44 mmHg) and DBP (4.98 mmHg) remained well within the maximum permissible limits of 6.88 and 6.95 mmHg, respectively, thereby satisfying the validation criteria.

Conclusion: The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.

根据美国国家标准协会/医疗器械进步协会/国际标准化组织81060-2:2018/ am2d:2024, AViTA上臂血压监测仪BPM636在儿科和成人人群中的验证。
目的:根据美国国家标准协会/医疗器械进步协会/国际标准化组织(ANSI/AAMI/ISO)通用标准(ISO 81060-2:2018/ am2d:2024),评估AViTA BPM636上臂血压计在儿童和成人(包括5-12岁儿童和21-82岁成人)上臂围15 - 42 cm的临床准确性。方法:选取上臂围15 ~ 42 cm的85名受试者(35名儿童,50名成人),采用15 ~ 22 cm和22 ~ 42 cm两种不同尺寸的袖口对AViTA BPM636进行评估。招募符合ANSI/AAMI/ISO通用标准(ISO 81060-2:2018/ am2d:2024)的要求,使用相同的臂序贯血压测量方法,将儿童和成人结合使用。结果:对于标准1,测试装置与参考血压测量值之间差异的平均值±SD为收缩压0.93±6.11 mmHg,舒张压-0.20±5.64 mmHg。对于标准2,收缩压(5.44 mmHg)和舒张压(4.98 mmHg)的个体平均差异的SD分别保持在最大允许范围6.88和6.95 mmHg内,从而满足验证标准。结论:AViTA BPM636通过了ANSI/AAMI/ISO通用标准(ISO 81060-2:2018/ am2d:2024)协议的标准。它可以推荐用于测试人群的家庭血压测量,包括5-12岁的儿童和21-82岁的成年人,手臂周长从15到42厘米。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书