{"title":"Fluorescence-guided bronchoscopic localization with a novel silicone marker: An ex vivo human proof-of-concept study","authors":"Masaya Tamura MD, Takashi Sakai MD, Naoki Furukawa MD, Yujiro Bunno MD, Marino Yamamoto MD, Ryohei Miyazaki MD, Hironobu Okada MD, Takayuki Sato MD","doi":"10.1016/j.xjtc.2026.102564","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To evaluate the technical feasibility and potential clinical applicability of a novel fluorescence-guided bronchoscopic localization strategy using a solid fluorescent silicone rubber marker for intraoperative localization of small peripheral pulmonary nodules.</div></div><div><h3>Methods</h3><div>An ex vivo human lung model was established using 4 freshly resected human pulmonary lobes. A radiopaque fluorescent silicone rubber marker (LuxSeed) was deployed bronchoscopically into peripheral bronchi without fluoroscopic or navigation guidance. Transpleural fluorescence was evaluated using a commercially available near-infrared thoracoscopic imaging system. Fluorescence detectability was assessed according to lung parenchymal thickness under noncompression and temporary compression conditions, and fluorescence intensity was quantified by signal-to-background ratio analysis.</div></div><div><h3>Results</h3><div>Successful deployment was achieved in all 4 resected pulmonary lobes. Fluorescence was consistently detected through lung parenchyma up to 12-mm thick without compression. Although fluorescence was not visible at depths of 15 and 20 mm under static conditions, temporary lung compression restored fluorescence detectability. Quantitative image analysis demonstrated improved signal-to-background ratios with compression, consistent with visual assessment. The marker was readily visualized using a standard near-infrared thoracoscopic imaging platform without dedicated localization equipment.</div></div><div><h3>Conclusions</h3><div>This proof-of-concept study demonstrates the feasibility of a novel fluorescence-guided bronchoscopic localization technology using a solid fluorescent silicone rubber marker. Compatibility with existing near-infrared thoracoscopic imaging systems and improved detectability during routine lung compression support future clinical evaluation in minimally invasive thoracic surgery.</div></div>","PeriodicalId":53413,"journal":{"name":"JTCVS Techniques","volume":"39 ","pages":"Article 102564"},"PeriodicalIF":2.4000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JTCVS Techniques","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666250726003718","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/3 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
To evaluate the technical feasibility and potential clinical applicability of a novel fluorescence-guided bronchoscopic localization strategy using a solid fluorescent silicone rubber marker for intraoperative localization of small peripheral pulmonary nodules.
Methods
An ex vivo human lung model was established using 4 freshly resected human pulmonary lobes. A radiopaque fluorescent silicone rubber marker (LuxSeed) was deployed bronchoscopically into peripheral bronchi without fluoroscopic or navigation guidance. Transpleural fluorescence was evaluated using a commercially available near-infrared thoracoscopic imaging system. Fluorescence detectability was assessed according to lung parenchymal thickness under noncompression and temporary compression conditions, and fluorescence intensity was quantified by signal-to-background ratio analysis.
Results
Successful deployment was achieved in all 4 resected pulmonary lobes. Fluorescence was consistently detected through lung parenchyma up to 12-mm thick without compression. Although fluorescence was not visible at depths of 15 and 20 mm under static conditions, temporary lung compression restored fluorescence detectability. Quantitative image analysis demonstrated improved signal-to-background ratios with compression, consistent with visual assessment. The marker was readily visualized using a standard near-infrared thoracoscopic imaging platform without dedicated localization equipment.
Conclusions
This proof-of-concept study demonstrates the feasibility of a novel fluorescence-guided bronchoscopic localization technology using a solid fluorescent silicone rubber marker. Compatibility with existing near-infrared thoracoscopic imaging systems and improved detectability during routine lung compression support future clinical evaluation in minimally invasive thoracic surgery.