A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care Glucometer

IF 8.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Pit Hung Lim, Sian Foong Lim, Kelly Lim, Lizhen Ong
{"title":"A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care Glucometer","authors":"Pit Hung Lim, Sian Foong Lim, Kelly Lim, Lizhen Ong","doi":"10.1093/clinchem/hvag086.308","DOIUrl":null,"url":null,"abstract":"Background Point-of-care testing (POCT) for glucose monitoring plays a crucial role in enabling timely clinical decisions within healthcare settings. The Roche cobas® pulse is a POCT device for glucose measurement and is compatible with cobas® GLU test strips and cobas® GLU QC kit for internal quality control (Roche Diagnostics GmbH). This new generation glucometer utilises FAD-dependent glucose dehydrogenase (GDH) enzyme technology and is intended to replace the current Roche Inform II system. This study provided a real-world analytical and clinical accuracy evaluation of the Roche cobas® pulse glucometer compared to its predecessor Roche Inform II. Methods Between June to October 2025, we evaluated Roche cobas® pulse glucometer (FAD-GDH electrochemical method) against its predecessor Roche Inform II (mutant quinoprotein GDH electrochemical method) and referencing to central laboratory platform testing on Abbott CI hexokinase glucose method for method comparison, imprecision, linearity and accuracy. Within-day and between-day imprecision were assessed using two QC levels (Low level: 3.3 mmol/L; High level: 17.7 mmol/L). Linearity was verified using commercial linearity material ranging from 1.5 to 29.7 mmol/L. Method correlation studies were performed using 77 patient samples for each comparison (Alinity CI and Inform II) across the reportable range of 0.6 to 33.3 mmol/L. Accuracy was verified using CAP WBGQ-B 2025 EQA samples. Performance was evaluated using Royal College of Pathologists of Australasia (RCPA) allowable performance specifications (±0.4 mmol/L for glucose =5.0 mmol/L or ±8% for glucose >5.0 mmol/L) and Food and Drug Administration (FDA) recommendations of 98% accuracy within ±15% for glucose = 4.2 mmol/L or ±0.83 mmol/L for glucose < 4.2 mmol/L. Results The Roche cobas® pulse showed excellent precision with within-day CV of 1.0% and 0.5% and between-day CV of 1.3% and 0.4% for low and high levels respectively, fulfilling manufacturer claims. Linearity was verified across the tested range (R²=0.9998). Correlation with Abbott Alinity CI showed strong agreement (cobas® pulse = 0.9291[Alinity CI] + 0.2634, R²=0.9979, n=76), with 71 of 76 samples (93.4%) meeting RCPA allowable performance specifications (APS) after excluding 1 outlier. Comparison with Roche Inform II showed excellent correlation (cobas® pulse = 0.9693[Inform II] + 0.013, R²=0.9940, n=77), with 73 of 77 samples (94.8%) meeting RCPA APS. Both comparisons achieved 100% compliance with FDA hospital-use accuracy criteria, exceeding the 98% requirement. Accuracy verification using CAP EQA samples showed all results within ±2 SDI and CAP evaluation criteria (peer group mean ±12.5%). The Roche cobas® pulse glucometer demonstrated analytical equivalence to Inform II, with improved precision, equivalent linearity across the clinical range and comparable accuracy. Conclusion The Roche cobas® pulse glucometer demonstrated excellent analytical performance, meeting all manufacturer's claims, RCPA performance specifications, FDA hospital use recommendation and CAP evaluation criteria. It performed comparably to Roche Inform II and to central laboratory platform on Abbott Alinity, >93% of samples within RCPA APS criteria, confirms its suitability for replacing the Roche Inform II glucometer for point-of-care glucose monitoring in professional healthcare settings. Collectively, these findings reinforce cobas® pulse as a clinically dependable and technologically advanced successor to the Inform II in professional healthcare environments.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"49 1","pages":""},"PeriodicalIF":8.3000,"publicationDate":"2026-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvag086.308","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background Point-of-care testing (POCT) for glucose monitoring plays a crucial role in enabling timely clinical decisions within healthcare settings. The Roche cobas® pulse is a POCT device for glucose measurement and is compatible with cobas® GLU test strips and cobas® GLU QC kit for internal quality control (Roche Diagnostics GmbH). This new generation glucometer utilises FAD-dependent glucose dehydrogenase (GDH) enzyme technology and is intended to replace the current Roche Inform II system. This study provided a real-world analytical and clinical accuracy evaluation of the Roche cobas® pulse glucometer compared to its predecessor Roche Inform II. Methods Between June to October 2025, we evaluated Roche cobas® pulse glucometer (FAD-GDH electrochemical method) against its predecessor Roche Inform II (mutant quinoprotein GDH electrochemical method) and referencing to central laboratory platform testing on Abbott CI hexokinase glucose method for method comparison, imprecision, linearity and accuracy. Within-day and between-day imprecision were assessed using two QC levels (Low level: 3.3 mmol/L; High level: 17.7 mmol/L). Linearity was verified using commercial linearity material ranging from 1.5 to 29.7 mmol/L. Method correlation studies were performed using 77 patient samples for each comparison (Alinity CI and Inform II) across the reportable range of 0.6 to 33.3 mmol/L. Accuracy was verified using CAP WBGQ-B 2025 EQA samples. Performance was evaluated using Royal College of Pathologists of Australasia (RCPA) allowable performance specifications (±0.4 mmol/L for glucose =5.0 mmol/L or ±8% for glucose >5.0 mmol/L) and Food and Drug Administration (FDA) recommendations of 98% accuracy within ±15% for glucose = 4.2 mmol/L or ±0.83 mmol/L for glucose < 4.2 mmol/L. Results The Roche cobas® pulse showed excellent precision with within-day CV of 1.0% and 0.5% and between-day CV of 1.3% and 0.4% for low and high levels respectively, fulfilling manufacturer claims. Linearity was verified across the tested range (R²=0.9998). Correlation with Abbott Alinity CI showed strong agreement (cobas® pulse = 0.9291[Alinity CI] + 0.2634, R²=0.9979, n=76), with 71 of 76 samples (93.4%) meeting RCPA allowable performance specifications (APS) after excluding 1 outlier. Comparison with Roche Inform II showed excellent correlation (cobas® pulse = 0.9693[Inform II] + 0.013, R²=0.9940, n=77), with 73 of 77 samples (94.8%) meeting RCPA APS. Both comparisons achieved 100% compliance with FDA hospital-use accuracy criteria, exceeding the 98% requirement. Accuracy verification using CAP EQA samples showed all results within ±2 SDI and CAP evaluation criteria (peer group mean ±12.5%). The Roche cobas® pulse glucometer demonstrated analytical equivalence to Inform II, with improved precision, equivalent linearity across the clinical range and comparable accuracy. Conclusion The Roche cobas® pulse glucometer demonstrated excellent analytical performance, meeting all manufacturer's claims, RCPA performance specifications, FDA hospital use recommendation and CAP evaluation criteria. It performed comparably to Roche Inform II and to central laboratory platform on Abbott Alinity, >93% of samples within RCPA APS criteria, confirms its suitability for replacing the Roche Inform II glucometer for point-of-care glucose monitoring in professional healthcare settings. Collectively, these findings reinforce cobas® pulse as a clinically dependable and technologically advanced successor to the Inform II in professional healthcare environments.
罗氏Cobas®脉搏即时血糖仪的A-378性能评估
背景:用于血糖监测的即时检测(POCT)在医疗保健环境中实现及时的临床决策方面发挥着至关重要的作用。Roche cobas®pulse是一种用于葡萄糖测量的POCT设备,与cobas®GLU试笔和cobas®GLU QC试剂盒兼容,用于内部质量控制(Roche Diagnostics GmbH)。这种新一代血糖仪利用fad依赖性葡萄糖脱氢酶(GDH)酶技术,旨在取代目前的罗氏Inform II系统。与之前的罗氏Inform II相比,本研究提供了对罗氏cobas®脉搏血糖仪的实际分析和临床准确性评估。方法于2025年6月至10月,将Roche cobas®脉冲血糖仪(FAD-GDH电化学法)与其前身Roche Inform II(突变型醌蛋白GDH电化学法)进行比较,并参考雅培CI己糖激酶葡萄糖法的中心实验室平台测试,进行方法比较、不精密度、线性度和准确性。采用两种QC水平(低水平:3.3 mmol/L;高水平:17.7 mmol/L)评估日内和日间不精确度。采用1.5 ~ 29.7 mmol/L的商业线性材料进行线性验证。方法相关性研究使用77例患者样本进行每次比较(Alinity CI和Inform II),报告范围为0.6至33.3 mmol/L。使用CAP WBGQ-B 2025 EQA样品验证准确性。使用澳大利亚皇家病理学院(RCPA)允许的性能规范(葡萄糖±0.4 mmol/L =5.0 mmol/L或±8%葡萄糖+ 5.0 mmol/L)和美国食品和药物管理局(FDA)推荐的98%准确度在±15%范围内葡萄糖= 4.2 mmol/L或±0.83 mmol/L葡萄糖+ amp;4.2更易/ L。结果罗氏cobas®脉冲具有良好的精度,低、高水平的日内CV分别为1.0%和0.5%,日间CV分别为1.3%和0.4%,符合制造商的要求。在测试范围内证实线性(R²=0.9998)。与雅培Alinity CI的相关性显示出很强的一致性(cobas®脉冲= 0.9291[Alinity CI] + 0.2634, R²=0.9979,n=76),排除1个异常值后,76个样本中有71个(93.4%)符合RCPA允许性能规格(APS)。与罗氏Inform II比较,相关性极好(cobas®脉冲= 0.9693[Inform II] + 0.013, R²=0.9940,n=77), 77份样本中有73份(94.8%)符合RCPA APS。两种比较都达到了100%符合FDA医院使用准确性标准,超过了98%的要求。使用CAP EQA样本进行准确性验证,结果均在±2 SDI和CAP评价标准范围内(同行组平均值±12.5%)。罗氏cobas®脉搏血糖仪与Inform II的分析等效性,具有更高的精度,在整个临床范围内的等效线性和相当的准确性。结论罗氏cobas®脉搏血糖仪具有优异的分析性能,符合所有制造商的声明、RCPA性能规范、FDA医院使用建议和CAP评估标准。它的表现与罗氏Inform II和雅培Alinity的中心实验室平台相当。93%的样本符合RCPA APS标准,证实其适合替代罗氏Inform II血糖仪,用于专业医疗机构的即时血糖监测。总的来说,这些发现加强了cobas®pulse作为专业医疗保健环境中Inform II的临床可靠和技术先进的继任者。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书