{"title":"B-415 Agreement Between Colorimetric Roche Cobas PRO and ICP-MS Zinc Assays in Blood Samples","authors":"Waleed Tamimi, Khaled Ahmad, Monzer Elsafdi, Ahmad Kamaleddine, Mohammed Alabdullatif, Majd Alharbi, Khalid Sumaily, Khalid Al-Qahtani, Ahmed Alakedi, Osamah khojah, Omar Almugren, Mohamed Babiker, Sara Zubair, Kareem Shehata, Duaa BinKhalaf","doi":"10.1093/clinchem/hvag086.682","DOIUrl":null,"url":null,"abstract":"Background Accurate measurement of zinc concentrations in blood is essential for assessing nutritional status and diagnosing deficiency. This study compared zinc results obtained using the Roche Cobas Pro colorimetric assay with those generated by inductively coupled plasma mass spectrometry (ICP-MS), the analytical reference method. Methods A total of 22 patient serum blood samples collected in royal-blue top tubes, spanning a concentration range of 55–97?µg/dL, were analyzed by both techniques. This study was conducted on January 2026 between the laboratory of King Abdelaziz Medical City and Medical Diagnostic laboratory of Dr Suliman Al-Habib Sahafah Hospital, both in Riyadh of Saudi Arabia. Direct colorimetric test without deproteinization of the sample. At pH 8.6, in a buffered media, zinc reacts with the specific compound (5-Br-PAPS): Bromo-pyridylazo propyl-N-(3-sulfopropyl) amino] phenol, forms a stable-colored complex and measured at wavelength 560nm. The color intensity is proportional to the amount of zinc present in the sample. The results of CAP proficiency samples were also compared for the two assays. Results Method comparison demonstrated strong agreement, with a correlation coefficient of ??= 0.9269, indicating a high degree of linear association across the tested range. There were 6 samples (27%) with zinc deficiency by colorimetric method with 83% agreement with ICP-MS method. On the other hand, a 100% agreement was found in the normal samples between the two assays. The reference interval was verified for the colorimetric Roche Cobas and found to be between 68 to 107 ug/dL. The CAP proficiency of 3 samples (R-01, R-02, R-03) for years 2024 and 2025 have shown a good agreement between the two assays. Conclusion While minor proportional differences were observed, the overall performance of the Cobas colorimetric method showed good concordance with ICP-MS, supporting its suitability for routine clinical zinc measurement when rapid, high-throughput analysis is required.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"35 1","pages":""},"PeriodicalIF":8.3000,"publicationDate":"2026-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvag086.682","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background Accurate measurement of zinc concentrations in blood is essential for assessing nutritional status and diagnosing deficiency. This study compared zinc results obtained using the Roche Cobas Pro colorimetric assay with those generated by inductively coupled plasma mass spectrometry (ICP-MS), the analytical reference method. Methods A total of 22 patient serum blood samples collected in royal-blue top tubes, spanning a concentration range of 55–97?µg/dL, were analyzed by both techniques. This study was conducted on January 2026 between the laboratory of King Abdelaziz Medical City and Medical Diagnostic laboratory of Dr Suliman Al-Habib Sahafah Hospital, both in Riyadh of Saudi Arabia. Direct colorimetric test without deproteinization of the sample. At pH 8.6, in a buffered media, zinc reacts with the specific compound (5-Br-PAPS): Bromo-pyridylazo propyl-N-(3-sulfopropyl) amino] phenol, forms a stable-colored complex and measured at wavelength 560nm. The color intensity is proportional to the amount of zinc present in the sample. The results of CAP proficiency samples were also compared for the two assays. Results Method comparison demonstrated strong agreement, with a correlation coefficient of ??= 0.9269, indicating a high degree of linear association across the tested range. There were 6 samples (27%) with zinc deficiency by colorimetric method with 83% agreement with ICP-MS method. On the other hand, a 100% agreement was found in the normal samples between the two assays. The reference interval was verified for the colorimetric Roche Cobas and found to be between 68 to 107 ug/dL. The CAP proficiency of 3 samples (R-01, R-02, R-03) for years 2024 and 2025 have shown a good agreement between the two assays. Conclusion While minor proportional differences were observed, the overall performance of the Cobas colorimetric method showed good concordance with ICP-MS, supporting its suitability for routine clinical zinc measurement when rapid, high-throughput analysis is required.
期刊介绍:
Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM).
The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics.
In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology.
The journal is indexed in databases such as MEDLINE and Web of Science.