B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930

IF 8.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Imad Tarhoni, Kelsey Reschly, Mackenzie Edward, Margaret Gearhart, Nichole Korpi-Steiner, Steven Cotten
{"title":"B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930","authors":"Imad Tarhoni, Kelsey Reschly, Mackenzie Edward, Margaret Gearhart, Nichole Korpi-Steiner, Steven Cotten","doi":"10.1093/clinchem/hvag086.582","DOIUrl":null,"url":null,"abstract":"Background Accurate hemoglobin A1c (HbA1c) measurement is key for diabetes diagnosis and monitoring. Many methods exist; however, immunoassay methods may be limited in detection of hemoglobin variants, abnormal fractions, and hemoglobin concentrations that impact result measurement. This study evaluated the analytical performance, variant detection, and result reportability of HbA1c measurement using the CAPILLARYS 3 TERA and Siemens Atellica CH930. Methods Studies included precision, accuracy, method comparisons, retesting of flagged samples, variant prevalence analysis, and precision of variant samples. CAPILLARYS 3 TERA within-run precision was evaluated using three residual samples (low, medium, high) analyzed in eight replicates per level and between-run precision was assessed in triplicate over three days. Accuracy was evaluated using 10 NGSP samples with assigned values. Six hundred and thirty-three residual EDTA samples with HbA1c values using the Atellica CH930 sent to McLendon Laboratory as part of routine care were tested using the CAPILLARYS 3 TERA within 24 hours of collection. Retesting of unreportable samples on the Atellica included specimens comprising flagged samples (n=64) and samples (n=65) with total hemoglobin concentrations of 7-12 g/dL. Results were compared to the Abbott Afinion 2 HbA1c diagnostic cartridge when HbA1c values were not generated by the Atellica. An enriched cohort (n=32) was used to challenge hemoglobin variant detection on both platforms. Variant precision was assessed in triplicate for nine specimens. Results CAPILLARYS 3 within-run, and between-run precision demonstrated CVs =1.5% across all levels. All measurements of NGSP samples on both platforms were within 6% relative bias. Slope, intercept, and mean absolute bias were 1.090, -0.373, 0.37% for CAPILLARYS 3 and 1.036, -0.12, and 0.18% for the Atellica compared to NGSP. Method comparison of the CAPILLARYS 3 versus Atellica using residual samples demonstrated a slope of 1.029, intercept of -0.05, correlation of 0.987 and mean bias of 0.12%. The CAPILLARYS 3 generated results for 96.9% (62/64) of Atellica flagged samples. These results showed good agreement (slope 1.025, bias -0.16) compared to the Afinion 2. The two remaining unreportable samples were flagged by the CAPILLARYS 3 as atypical and were previously known hemoglobin variants. For the variant enriched cohort, the CAPILLARYS 3 had a sensitivity of 100% versus 81% for the Atellica in flagging samples with possible interferences. The CAPILLARYS 3 correctly identified variant interferences despite calculating HbA1c values for 13 specimens collected from transfused patients. Six variant samples were not identified by the Atellica yet generated results; five were ß-globin related hemoglobinopathies, with Atellica values systematically lower than CAPILLARYS 3, showing 15-20% underestimation. Additionally, flagged and variant samples showed CV <1%, supporting the stability of HbA1c measurements on CAPILLARYS 3. Notably, in the 7-12 g/dL hemoglobin cohort on the Atellica, 30 samples had flags indicating total hemoglobin <7 g/dL, inconsistent with the total hemoglobin concentration measured. This discrepancy was independent of variant status Conclusion The CAPILLARYS 3 TERA demonstrated accurate resolution of problematic specimens originally tested using the Atellica enzymatic immunoassay. Furthermore, it may prevent inaccurate results in the patient population with hemoglobin variants.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"12 1","pages":""},"PeriodicalIF":8.3000,"publicationDate":"2026-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvag086.582","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background Accurate hemoglobin A1c (HbA1c) measurement is key for diabetes diagnosis and monitoring. Many methods exist; however, immunoassay methods may be limited in detection of hemoglobin variants, abnormal fractions, and hemoglobin concentrations that impact result measurement. This study evaluated the analytical performance, variant detection, and result reportability of HbA1c measurement using the CAPILLARYS 3 TERA and Siemens Atellica CH930. Methods Studies included precision, accuracy, method comparisons, retesting of flagged samples, variant prevalence analysis, and precision of variant samples. CAPILLARYS 3 TERA within-run precision was evaluated using three residual samples (low, medium, high) analyzed in eight replicates per level and between-run precision was assessed in triplicate over three days. Accuracy was evaluated using 10 NGSP samples with assigned values. Six hundred and thirty-three residual EDTA samples with HbA1c values using the Atellica CH930 sent to McLendon Laboratory as part of routine care were tested using the CAPILLARYS 3 TERA within 24 hours of collection. Retesting of unreportable samples on the Atellica included specimens comprising flagged samples (n=64) and samples (n=65) with total hemoglobin concentrations of 7-12 g/dL. Results were compared to the Abbott Afinion 2 HbA1c diagnostic cartridge when HbA1c values were not generated by the Atellica. An enriched cohort (n=32) was used to challenge hemoglobin variant detection on both platforms. Variant precision was assessed in triplicate for nine specimens. Results CAPILLARYS 3 within-run, and between-run precision demonstrated CVs =1.5% across all levels. All measurements of NGSP samples on both platforms were within 6% relative bias. Slope, intercept, and mean absolute bias were 1.090, -0.373, 0.37% for CAPILLARYS 3 and 1.036, -0.12, and 0.18% for the Atellica compared to NGSP. Method comparison of the CAPILLARYS 3 versus Atellica using residual samples demonstrated a slope of 1.029, intercept of -0.05, correlation of 0.987 and mean bias of 0.12%. The CAPILLARYS 3 generated results for 96.9% (62/64) of Atellica flagged samples. These results showed good agreement (slope 1.025, bias -0.16) compared to the Afinion 2. The two remaining unreportable samples were flagged by the CAPILLARYS 3 as atypical and were previously known hemoglobin variants. For the variant enriched cohort, the CAPILLARYS 3 had a sensitivity of 100% versus 81% for the Atellica in flagging samples with possible interferences. The CAPILLARYS 3 correctly identified variant interferences despite calculating HbA1c values for 13 specimens collected from transfused patients. Six variant samples were not identified by the Atellica yet generated results; five were ß-globin related hemoglobinopathies, with Atellica values systematically lower than CAPILLARYS 3, showing 15-20% underestimation. Additionally, flagged and variant samples showed CV <1%, supporting the stability of HbA1c measurements on CAPILLARYS 3. Notably, in the 7-12 g/dL hemoglobin cohort on the Atellica, 30 samples had flags indicating total hemoglobin <7 g/dL, inconsistent with the total hemoglobin concentration measured. This discrepancy was independent of variant status Conclusion The CAPILLARYS 3 TERA demonstrated accurate resolution of problematic specimens originally tested using the Atellica enzymatic immunoassay. Furthermore, it may prevent inaccurate results in the patient population with hemoglobin variants.
B-299相对于西门子Atellica CH930的CAPILLARYS 3 TERA HbA1c检测的分析性能、变异检测和结果报告性
背景准确的糖化血红蛋白(HbA1c)检测是糖尿病诊断和监测的关键。存在许多方法;然而,免疫分析方法在检测血红蛋白变异、异常组分和血红蛋白浓度方面可能受到限制,这些会影响结果的测量。本研究评估了使用CAPILLARYS 3 TERA和Siemens Atellica CH930测量HbA1c的分析性能、变异检测和结果报告性。方法研究包括精密度、准确度、方法比较、标记样本复检、变异流行率分析和变异样本的精密度。在每个水平的8个重复中分析3个剩余样本(低、中、高),评估CAPILLARYS 3 TERA的运行内精度,并在三天内评估三个重复的运行间精度。使用10个具有指定值的NGSP样本评估准确性。633份HbA1c值的残留EDTA样本,作为常规护理的一部分,使用Atellica CH930送到McLendon实验室,在收集的24小时内使用CAPILLARYS 3 TERA进行检测。在Atellica上重新检测的不可报告样本包括标记样本(n=64)和总血红蛋白浓度为7-12 g/dL的样本(n=65)。当HbA1c值不是由Atellica产生时,将结果与雅培Afinion 2 HbA1c诊断盒进行比较。一个富集队列(n=32)用于挑战两个平台上的血红蛋白变异检测。对9个标本进行了三次变异精度评估。结果capillillys 3运行内和运行间精度在所有水平上的CVs =1.5%。两个平台上NGSP样本的所有测量值的相对偏差都在6%以内。与NGSP相比,CAPILLARYS 3的斜率、截距和平均绝对偏差分别为1.090、-0.373和0.37%,Atellica的斜率、截距和平均绝对偏差分别为1.036、-0.12和0.18%。用剩余样本比较CAPILLARYS 3和Atellica的方法,斜率为1.029,截距为-0.05,相关性为0.987,平均偏差为0.12%。CAPILLARYS 3对96.9%(62/64)的Atellica标记样本产生了结果。与Afinion 2相比,这些结果显示出良好的一致性(斜率1.025,偏差-0.16)。剩下的两个不可报告的样本被CAPILLARYS 3标记为非典型,并且是以前已知的血红蛋白变体。对于变异富集的队列,CAPILLARYS 3在标记有可能干扰的样品中对Atellica的敏感性为100%,而对Atellica的敏感性为81%。尽管CAPILLARYS 3计算了从输血患者收集的13个标本的HbA1c值,但它正确地识别了变异干扰。Atellica尚未识别出6个变异样本,但尚未产生结果;5例为ß-globin相关的血红蛋白病,Atellica值系统性低于CAPILLARYS 3,低估15-20%。此外,标记和变异样本显示CV &;lt;1%,支持在CAPILLARYS上测量HbA1c的稳定性。值得注意的是,在Atellica的7-12 g/dL血红蛋白队列中,有30个样本有标记总血红蛋白&;lt;7 g/dL,与测定的总血红蛋白浓度不一致。结论CAPILLARYS 3 TERA能够准确地解决最初使用Atellica酶促免疫分析法检测的问题标本。此外,它可以防止在血红蛋白变异患者群体中出现不准确的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书