Peter M Maloca, Philippe C Cattin, Nadja Inglin, Claudia Bruè, Pascal W Hasler, Simon Rothenbuehler, Hanna Zuche, Javier Zarranz-Ventura, Philippe Valmaggia
{"title":"Noninterchangeability of Vessel Density Measurements Across Next-Generation SS-OCTA Systems.","authors":"Peter M Maloca, Philippe C Cattin, Nadja Inglin, Claudia Bruè, Pascal W Hasler, Simon Rothenbuehler, Hanna Zuche, Javier Zarranz-Ventura, Philippe Valmaggia","doi":"10.1167/tvst.15.9.24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Comparing swept-source optical coherence tomography angiography (OCTA) vessel density in the superficial and deep capillary plexuses across leading OCTA systems.</p><p><strong>Methods: </strong>OCTAs were acquired from healthy participants using two OCTA systems (BMizar 400 kHz and Yalkaid 100 kHz, TowardPi, Beijing, China) with scan sizes of 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm2. The vessel density was quantified using the manufacturer's software.</p><p><strong>Results: </strong>Across 30 subjects comprising 60 eyes and a total of 225 OCTA scans, significant interdevice differences were observed at the Early Treatment Diabetic Retinopathy Study sector level in all eight scan size × plexus comparisons (three-way analysis of variance; P < 1.16 × 10e-4). The mean BMizar vessel density values for the superficial and deep plexuses were 43.4% ± 1.6% and 44.9% ± 1.5%, 44.7% ± 2.7% and 47.9% ± 1.1%, 44.2% ± 1.3% and 45.7% ± 1.1%, and 42.4 ± 1.6 % and 44.3 ± 1.3% for the 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm² scans; the Yalkaid values were lower. BMizar showed greater vessel density than Yalkaid, particularly in 3 × 3 mm² deep slab scans, with differences of approximately 5% to 10%. Across scans larger than 3 × 3 mm², BMizar showed greater central vessel density; the peripheral patterns varied, with reversals favoring Yalkaid and greater border effects in larger fields of view.</p><p><strong>Conclusions: </strong>Despite interdevice differences, both devices showed consistent OCTA acquisition.</p><p><strong>Translational relevance: </strong>Device differences may affect clinical interpretation, longitudinal follow-up, and multicenter studies.</p>","PeriodicalId":23322,"journal":{"name":"Translational Vision Science & Technology","volume":"15 9","pages":"24"},"PeriodicalIF":2.7000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13641204/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Translational Vision Science & Technology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1167/tvst.15.9.24","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Comparing swept-source optical coherence tomography angiography (OCTA) vessel density in the superficial and deep capillary plexuses across leading OCTA systems.
Methods: OCTAs were acquired from healthy participants using two OCTA systems (BMizar 400 kHz and Yalkaid 100 kHz, TowardPi, Beijing, China) with scan sizes of 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm2. The vessel density was quantified using the manufacturer's software.
Results: Across 30 subjects comprising 60 eyes and a total of 225 OCTA scans, significant interdevice differences were observed at the Early Treatment Diabetic Retinopathy Study sector level in all eight scan size × plexus comparisons (three-way analysis of variance; P < 1.16 × 10e-4). The mean BMizar vessel density values for the superficial and deep plexuses were 43.4% ± 1.6% and 44.9% ± 1.5%, 44.7% ± 2.7% and 47.9% ± 1.1%, 44.2% ± 1.3% and 45.7% ± 1.1%, and 42.4 ± 1.6 % and 44.3 ± 1.3% for the 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm² scans; the Yalkaid values were lower. BMizar showed greater vessel density than Yalkaid, particularly in 3 × 3 mm² deep slab scans, with differences of approximately 5% to 10%. Across scans larger than 3 × 3 mm², BMizar showed greater central vessel density; the peripheral patterns varied, with reversals favoring Yalkaid and greater border effects in larger fields of view.
Conclusions: Despite interdevice differences, both devices showed consistent OCTA acquisition.
Translational relevance: Device differences may affect clinical interpretation, longitudinal follow-up, and multicenter studies.
期刊介绍:
Translational Vision Science & Technology (TVST), an official journal of the Association for Research in Vision and Ophthalmology (ARVO), an international organization whose purpose is to advance research worldwide into understanding the visual system and preventing, treating and curing its disorders, is an online, open access, peer-reviewed journal emphasizing multidisciplinary research that bridges the gap between basic research and clinical care. A highly qualified and diverse group of Associate Editors and Editorial Board Members is led by Editor-in-Chief Marco Zarbin, MD, PhD, FARVO.
The journal covers a broad spectrum of work, including but not limited to:
Applications of stem cell technology for regenerative medicine,
Development of new animal models of human diseases,
Tissue bioengineering,
Chemical engineering to improve virus-based gene delivery,
Nanotechnology for drug delivery,
Design and synthesis of artificial extracellular matrices,
Development of a true microsurgical operating environment,
Refining data analysis algorithms to improve in vivo imaging technology,
Results of Phase 1 clinical trials,
Reverse translational ("bedside to bench") research.
TVST seeks manuscripts from scientists and clinicians with diverse backgrounds ranging from basic chemistry to ophthalmic surgery that will advance or change the way we understand and/or treat vision-threatening diseases. TVST encourages the use of color, multimedia, hyperlinks, program code and other digital enhancements.