Dual-phase systolic and diastolic cardiac diffusion tensor imaging using higher order motion compensation spin echo sequences at 300 mT/m gradient strength.

IF 5 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
Maryam Afzali, Irvin Teh, Sam Coveney, Lars Mueller, Isaac Watson, May Lwin, Sarah Jones, Fabrizio Fasano, C John Evans, Filip Szczepankiewicz, Derek K Jones, Erica Dall'Armellina, Jürgen E Schneider
{"title":"Dual-phase systolic and diastolic cardiac diffusion tensor imaging using higher order motion compensation spin echo sequences at 300 mT/m gradient strength.","authors":"Maryam Afzali, Irvin Teh, Sam Coveney, Lars Mueller, Isaac Watson, May Lwin, Sarah Jones, Fabrizio Fasano, C John Evans, Filip Szczepankiewicz, Derek K Jones, Erica Dall'Armellina, Jürgen E Schneider","doi":"10.1016/j.jocmr.2026.102805","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Cardiac diffusion tensor imaging (cDTI) provides unique insights into myocardial microstructure and its dynamic changes during the cardiac cycle. However, motion sensitivity remains a major obstacle for spin echo (SE)-based approaches, restricting most implementations to a relatively narrow low-motion window near peak systole.</p><p><strong>Purpose: </strong>To establish whether higher-order motion-compensated (up to 6th order) diffusion gradients beyond standard second order motion compensation (M<sub>2</sub>) on a magnetic resonance imaging (MRI) system with a maximum gradient strength of 300 mT/m improve SE-based cDTI techniques for both systolic and diastolic acquisitions.</p><p><strong>Methods: </strong>Ten healthy volunteers underwent free-breathing, cardiac-gated SE-cDTI acquisitions in systole and in diastole using diffusion gradients with up to second- (M<sub>2</sub>), fourth- (M<sub>4</sub>), and sixth-order (M<sub>6</sub>) motion compensation. Success rate (percentage of left ventricular voxels with (1.1 < MD < 1.9)  × 10<sup>-3</sup>mm<sup>2</sup>∕s) and diffusion measures including mean diffusivity, fractional anisotropy, helix angle, and secondary eigenvector angle were quantified and compared between both cardiac phases.</p><p><strong>Results: </strong>Second-order motion compensation was sufficient for systolic cDTI, whereas fourth-order motion compensation was required for successful diastolic acquisitions in all participants. Mean diffusivity values in systole were (1.53 ± 0.06), (1.60 ± 0.07),   and   (1.62 ± 0.06)  × 10<sup>-3</sup>mm<sup>2</sup>∕s for M<sub>2</sub>, M<sub>4</sub>,   and   M<sub>6</sub> respectively, and in diastole were (2.26 ± 0.67), (1.73 ± 0.19),   and   (1.71 ± 0.12) × 10<sup>-3</sup>mm<sup>2</sup>∕s. No significant differences in cDTI-derived parameters were observed between M<sub>4</sub> and M<sub>6</sub> for systolic and diastolic acquisitions.</p><p><strong>Conclusion: </strong>This study demonstrates the feasibility of dual-phase SE-based cDTI techniques enabled by 300 mT/m gradients. This method will allow reliable quantification of myocardial microstructural changes between systole and diastole and represents a step towards clinical translation of dynamic cDTI.</p>","PeriodicalId":15221,"journal":{"name":"Journal of Cardiovascular Magnetic Resonance","volume":" ","pages":"102805"},"PeriodicalIF":5.0000,"publicationDate":"2026-09-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Magnetic Resonance","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jocmr.2026.102805","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Cardiac diffusion tensor imaging (cDTI) provides unique insights into myocardial microstructure and its dynamic changes during the cardiac cycle. However, motion sensitivity remains a major obstacle for spin echo (SE)-based approaches, restricting most implementations to a relatively narrow low-motion window near peak systole.

Purpose: To establish whether higher-order motion-compensated (up to 6th order) diffusion gradients beyond standard second order motion compensation (M2) on a magnetic resonance imaging (MRI) system with a maximum gradient strength of 300 mT/m improve SE-based cDTI techniques for both systolic and diastolic acquisitions.

Methods: Ten healthy volunteers underwent free-breathing, cardiac-gated SE-cDTI acquisitions in systole and in diastole using diffusion gradients with up to second- (M2), fourth- (M4), and sixth-order (M6) motion compensation. Success rate (percentage of left ventricular voxels with (1.1 < MD < 1.9)  × 10-3mm2∕s) and diffusion measures including mean diffusivity, fractional anisotropy, helix angle, and secondary eigenvector angle were quantified and compared between both cardiac phases.

Results: Second-order motion compensation was sufficient for systolic cDTI, whereas fourth-order motion compensation was required for successful diastolic acquisitions in all participants. Mean diffusivity values in systole were (1.53 ± 0.06), (1.60 ± 0.07),   and   (1.62 ± 0.06)  × 10-3mm2∕s for M2, M4,   and   M6 respectively, and in diastole were (2.26 ± 0.67), (1.73 ± 0.19),   and   (1.71 ± 0.12) × 10-3mm2∕s. No significant differences in cDTI-derived parameters were observed between M4 and M6 for systolic and diastolic acquisitions.

Conclusion: This study demonstrates the feasibility of dual-phase SE-based cDTI techniques enabled by 300 mT/m gradients. This method will allow reliable quantification of myocardial microstructural changes between systole and diastole and represents a step towards clinical translation of dynamic cDTI.

采用高阶运动补偿自旋回波序列在300 mT/m梯度强度下进行心脏收缩和舒张双期弥散张量成像。
背景:心脏弥散张量成像(cDTI)为心肌微结构及其在心脏周期内的动态变化提供了独特的见解。然而,运动灵敏度仍然是基于自旋回波(SE)方法的主要障碍,限制了大多数实现在峰值收缩期附近相对狭窄的低运动窗口。目的:确定在最大梯度强度为300 mT/m的磁共振成像(MRI)系统上,高阶运动补偿(高达6阶)扩散梯度是否超过标准二阶运动补偿(M2),从而改善基于se的cDTI技术在收缩期和舒张期的采集。方法:10名健康志愿者在收缩期和舒张期使用扩散梯度进行自由呼吸、心控SE-cDTI采集,最高可达二级(M2)、四级(M4)和六级(M6)运动补偿。对两个心脏相的成功率(左室体素(1.1 -3mm2∕s)百分比)和扩散措施(包括平均扩散率、分数各向异性、螺旋角和次级特征向量角)进行量化并进行比较。结果:二级运动补偿对收缩期cDTI足够,而四级运动补偿对所有参与者的舒张期取得成功都是必需的。意味着收缩的扩散系数值(1.53±0.06),(1.60±0.07), 和 (1.62±0.06)×10-3mm2平方米∕年代,M4,分别 和 M6在心脏舒张期(2.26±0.67),(1.73±0.19),  (1.71±0.12)×10-3mm2∕年代。M4和M6在收缩期和舒张期获得的cdti衍生参数无显著差异。结论:本研究证明了300mt /m梯度下双相se基cDTI技术的可行性。这种方法将允许可靠的定量心肌收缩和舒张之间的微结构变化,并代表了动态cDTI临床翻译的一步。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
10.90
自引率
12.50%
发文量
61
审稿时长
6-12 weeks
期刊介绍: Journal of Cardiovascular Magnetic Resonance (JCMR) publishes high-quality articles on all aspects of basic, translational and clinical research on the design, development, manufacture, and evaluation of cardiovascular magnetic resonance (CMR) methods applied to the cardiovascular system. Topical areas include, but are not limited to: New applications of magnetic resonance to improve the diagnostic strategies, risk stratification, characterization and management of diseases affecting the cardiovascular system. New methods to enhance or accelerate image acquisition and data analysis. Results of multicenter, or larger single-center studies that provide insight into the utility of CMR. Basic biological perceptions derived by CMR methods.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书