{"title":"Validity of Bone Mineral Density Measurements Using SONIALVISION G4: A Quantitative CT-based Evaluation.","authors":"Junichiro Morita, Keisuke Uemura, Sotaro Kono, Yi Gu, Yuto Nishioka, Tomoki Asano, Ryo Higuchi, Hirokazu Mae, Makoto Iwasa, Tetsuro Tani, Kazuma Takashima, Kazunori Tamura, Takeshi Ogawa, Wataru Ando, Takeshi Ogawa, Yoshito Otake, Seiji Okada, Hidetoshi Hamada","doi":"10.1007/s00774-026-01766-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>SONIALVISION G4 is a fluoroscopic imaging system equipped with a bone densitometry function calibrated against GE Healthcare dual-energy X-ray absorptiometry (DXA) measurements. However, evidence supporting the validity of in vivo bone mineral density (BMD) measurements remains limited. This study compared BMD measurements obtained using SONIALVISION G4 with GE DXA reference data and evaluated their validity using quantitative computed tomography (CT).</p><p><strong>Materials and methods: </strong>A total of 887 BMD examinations performed with SONIALVISION G4 in women aged ≥45 years were compared with published Japanese reference data obtained using GE DXA. CT-derived BMD was quantified using a validated deep learning-based method and compared with measurements obtained using SONIALVISION G4 in 30 patients. To validate the CT method against GE DXA, 697 patients from four additional institutions underwent GE DXA and CT. Analyses were conducted for both the femoral neck and total hip.</p><p><strong>Results: </strong>Femoral neck BMD generally showed small differences from the reference values, whereas total hip BMD was consistently higher in women aged ≥55 years. CT-derived BMD correlated with SONIALVISION G4 measurements (r = 0.762 and 0.842 for the femoral neck and total hip, respectively), with median absolute differences of 0.022 and 0.133 g/cm<sup>2</sup>. In contrast, CT-derived BMD showed very strong correlations with GE DXA (r = 0.913 and 0.959) and smaller absolute differences (0.043 and 0.033 g/cm<sup>2</sup>).</p><p><strong>Conclusion: </strong>SONIALVISION G4 showed small differences in femoral neck BMD measurements compared with CT-derived measurements but may systematically overestimate total hip BMD. Further validation studies are warranted.</p>","PeriodicalId":15116,"journal":{"name":"Journal of Bone and Mineral Metabolism","volume":" ","pages":""},"PeriodicalIF":2.9000,"publicationDate":"2026-09-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Bone and Mineral Metabolism","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00774-026-01766-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: SONIALVISION G4 is a fluoroscopic imaging system equipped with a bone densitometry function calibrated against GE Healthcare dual-energy X-ray absorptiometry (DXA) measurements. However, evidence supporting the validity of in vivo bone mineral density (BMD) measurements remains limited. This study compared BMD measurements obtained using SONIALVISION G4 with GE DXA reference data and evaluated their validity using quantitative computed tomography (CT).

Materials and methods: A total of 887 BMD examinations performed with SONIALVISION G4 in women aged ≥45 years were compared with published Japanese reference data obtained using GE DXA. CT-derived BMD was quantified using a validated deep learning-based method and compared with measurements obtained using SONIALVISION G4 in 30 patients. To validate the CT method against GE DXA, 697 patients from four additional institutions underwent GE DXA and CT. Analyses were conducted for both the femoral neck and total hip.

Results: Femoral neck BMD generally showed small differences from the reference values, whereas total hip BMD was consistently higher in women aged ≥55 years. CT-derived BMD correlated with SONIALVISION G4 measurements (r = 0.762 and 0.842 for the femoral neck and total hip, respectively), with median absolute differences of 0.022 and 0.133 g/cm2. In contrast, CT-derived BMD showed very strong correlations with GE DXA (r = 0.913 and 0.959) and smaller absolute differences (0.043 and 0.033 g/cm2).

Conclusion: SONIALVISION G4 showed small differences in femoral neck BMD measurements compared with CT-derived measurements but may systematically overestimate total hip BMD. Further validation studies are warranted.

使用sonalvision G4测量骨密度的有效性:一种基于ct的定量评估。
简介:sonalvision G4是一种配备骨密度测量功能的透视成像系统,根据GE医疗双能x射线吸收仪(DXA)测量校准。然而,支持体内骨密度(BMD)测量的有效性的证据仍然有限。本研究将使用sonalvision G4获得的骨密度测量值与GE DXA参考数据进行比较,并使用定量计算机断层扫描(CT)评估其有效性。材料和方法:使用sonalvision G4对年龄≥45岁的女性进行887例骨密度检查,并与使用GE DXA获得的日本发表的参考数据进行比较。使用经过验证的基于深度学习的方法对ct衍生的骨密度进行量化,并与使用sonalvision G4获得的30例患者的测量结果进行比较。为了验证CT方法对GE DXA的有效性,来自另外四个机构的697名患者接受了GE DXA和CT治疗。对股骨颈和全髋进行分析。结果:股骨颈骨密度与参考值的差异较小,而总髋骨密度在年龄≥55岁的女性中一直较高。ct得出的骨密度与sonalvision G4测量值相关(股骨颈和全髋r分别为0.762和0.842),中位绝对差值分别为0.022和0.133 g/cm2。相比之下,ct衍生的BMD与GE DXA的相关性非常强(r = 0.913和0.959),绝对差异较小(0.043和0.033 g/cm2)。结论:sonalvision G4显示股骨颈骨密度测量值与ct测量值差异不大,但可能系统性地高估了全髋骨密度。进一步的验证研究是必要的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Bone and Mineral Metabolism
Journal of Bone and Mineral Metabolism 医学-内分泌学与代谢
CiteScore
6.30
自引率
3.00%
发文量
89
审稿时长
6-12 weeks
期刊介绍: The Journal of Bone and Mineral Metabolism (JBMM) provides an international forum for researchers and clinicians to present and discuss topics relevant to bone, teeth, and mineral metabolism, as well as joint and musculoskeletal disorders. The journal welcomes the submission of manuscripts from any country. Membership in the society is not a prerequisite for submission. Acceptance is based on the originality, significance, and validity of the material presented. The journal is aimed at researchers and clinicians dedicated to improvements in research, development, and patient-care in the fields of bone and mineral metabolism.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书