Gianluca Cappuccini, Kenneth J Hoffer, Arianna Grendele, Catarina C P Coutinho, Domenico Schiano-Lomoriello, Giacomo Savini
{"title":"Repeatability of an SD-OCT biometer (MS-39 AL) and agreement with a validated biometer (IOLMaster 700).","authors":"Gianluca Cappuccini, Kenneth J Hoffer, Arianna Grendele, Catarina C P Coutinho, Domenico Schiano-Lomoriello, Giacomo Savini","doi":"10.1097/j.jcrs.0000000000002064","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the repeatability of biometric measurements obtained with the MS-39 AL (CSO) and to assess their agreement with those provided by the IOLMaster 700 (Carl Zeiss).</p><p><strong>Setting: </strong>IRCCS Bietti Foundation, Rome, Italy.</p><p><strong>Design: </strong>Prospective evaluation of diagnostic test.</p><p><strong>Methods: </strong>Patients were classified in Group A (with cataract) and Group B (with clear lens). For repeatability analysis, three measurements were obtained with the MS-39 AL. Agreement was analyzed by comparing the first MS-39 AL measurement with the corresponding measurement by the IOLMaster 700.</p><p><strong>Results: </strong>111 eyes were enrolled. Group A included 53 eyes (mean age: 69.4 ±7.3 years) and Group B 58 eyes (mean age 35.5 ±9.1 years). In both groups, the MS-39 AL demonstrated high repeatability for all biometric parameters, as the coefficient of variation ranged from 0.04% for axial length to 2.10% for lens thickness, except in the case of corneal astigmatism. Agreement between biometers was high for most parameters, with narrow 95% limits of agreement. Significant differences were observed for anterior chamber depth (MS-39 AL = 3.52±0.44 mm, IOLMaster 700 = 3.42±0.45 mm, p <0.0001), corneal diameter (MS-39 AL = 11.90±0.46 mm, IOLMaster 700 = 12.19±0.49 mm, p <0.0001), central corneal thickness (MS-39 AL = 531.75±31.20 µm, IOLMaster 700 = 534.80±32.79 µm, p <0.001) and lens thickness (MS-39 AL = 4.18±0.55 mm, IOLMaster 700 = 4.13±0.52 mm, p = 0.002).</p><p><strong>Conclusions: </strong>The MS-39 AL provides highly repeatable biometric measurements and shows good agreement with the IOLMaster 700 for most parameters in both cataractous and clear-lens eyes.</p>","PeriodicalId":15214,"journal":{"name":"Journal of cataract and refractive surgery","volume":" ","pages":""},"PeriodicalIF":3.7000,"publicationDate":"2026-09-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cataract and refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/j.jcrs.0000000000002064","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the repeatability of biometric measurements obtained with the MS-39 AL (CSO) and to assess their agreement with those provided by the IOLMaster 700 (Carl Zeiss).
Setting: IRCCS Bietti Foundation, Rome, Italy.
Design: Prospective evaluation of diagnostic test.
Methods: Patients were classified in Group A (with cataract) and Group B (with clear lens). For repeatability analysis, three measurements were obtained with the MS-39 AL. Agreement was analyzed by comparing the first MS-39 AL measurement with the corresponding measurement by the IOLMaster 700.
Results: 111 eyes were enrolled. Group A included 53 eyes (mean age: 69.4 ±7.3 years) and Group B 58 eyes (mean age 35.5 ±9.1 years). In both groups, the MS-39 AL demonstrated high repeatability for all biometric parameters, as the coefficient of variation ranged from 0.04% for axial length to 2.10% for lens thickness, except in the case of corneal astigmatism. Agreement between biometers was high for most parameters, with narrow 95% limits of agreement. Significant differences were observed for anterior chamber depth (MS-39 AL = 3.52±0.44 mm, IOLMaster 700 = 3.42±0.45 mm, p <0.0001), corneal diameter (MS-39 AL = 11.90±0.46 mm, IOLMaster 700 = 12.19±0.49 mm, p <0.0001), central corneal thickness (MS-39 AL = 531.75±31.20 µm, IOLMaster 700 = 534.80±32.79 µm, p <0.001) and lens thickness (MS-39 AL = 4.18±0.55 mm, IOLMaster 700 = 4.13±0.52 mm, p = 0.002).
Conclusions: The MS-39 AL provides highly repeatable biometric measurements and shows good agreement with the IOLMaster 700 for most parameters in both cataractous and clear-lens eyes.