Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay.

IF 2.3 Q3 MEDICAL LABORATORY TECHNOLOGY
Nina M Diederiks, Jorik H Amesz, Fred P H T M Romijn, Enrico de Koning, Mark J Boogers, Christa M Cobbaert
{"title":"Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay.","authors":"Nina M Diederiks, Jorik H Amesz, Fred P H T M Romijn, Enrico de Koning, Mark J Boogers, Christa M Cobbaert","doi":"10.1093/jalm/jfag148","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>High-sensitivity cardiac troponin T (hs-cTnT) assays must meet strict criteria prior to regulatory approval, as the test is central to myocardial infarction diagnosis. Evaluating analyte stability is essential for ensuring accurate results in stored samples. Here, cTnT stability using the novel sixth-generation assay was evaluated. We also discuss regulatory complexity.</p><p><strong>Methods: </strong>From patients with acute coronary syndrome, serum and lithium heparin plasma (Li-Hep) were collected, pooled per matrix, and aliquoted. One aliquot was immediately analyzed (fresh) while the remainder was stored at ≤-60°C. Selected aliquots were thawed and measured (fresh-frozen) before transfer to 15-25°C for 3 and 5 h, or 2-8°C for 11 and 24 h. Long-term aliquots were transferred frozen to -15 to -25°C for ≥13 months or ≤-60 for 9 and ≥13 months. Measurements were performed in quadruplicate using the Elecsys® hs-cTnT Gen 6 assay on a Cobas® Pure analyzer (Roche Diagnostics). Recovery was calculated relative to fresh measurements. Maximum permissible differences (MPD) were ±1.2 ng/L, ±20%, and ±15% for cTnT concentrations <6 ng/L, ≥6-≤24 ng/L, and >24 ng/L, respectively, allowing one exceedance per condition.</p><p><strong>Results: </strong>In total, >2000 measurements were performed in 18 distinct samples per matrix. Median cTnT concentrations were 49.3 ng/L (8.4-782) in Li-Hep and 49.8 ng/L (8.5-822) in serum. The observed deviation in recovery was below the MPD for all storage conditions and acceptance criteria were met.</p><p><strong>Conclusions: </strong>Stability of cTnT using the sixth-generation hs-cTnT test was demonstrated across different storage conditions for serum and Li-Hep.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2026-09-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/jalm/jfag148","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: High-sensitivity cardiac troponin T (hs-cTnT) assays must meet strict criteria prior to regulatory approval, as the test is central to myocardial infarction diagnosis. Evaluating analyte stability is essential for ensuring accurate results in stored samples. Here, cTnT stability using the novel sixth-generation assay was evaluated. We also discuss regulatory complexity.

Methods: From patients with acute coronary syndrome, serum and lithium heparin plasma (Li-Hep) were collected, pooled per matrix, and aliquoted. One aliquot was immediately analyzed (fresh) while the remainder was stored at ≤-60°C. Selected aliquots were thawed and measured (fresh-frozen) before transfer to 15-25°C for 3 and 5 h, or 2-8°C for 11 and 24 h. Long-term aliquots were transferred frozen to -15 to -25°C for ≥13 months or ≤-60 for 9 and ≥13 months. Measurements were performed in quadruplicate using the Elecsys® hs-cTnT Gen 6 assay on a Cobas® Pure analyzer (Roche Diagnostics). Recovery was calculated relative to fresh measurements. Maximum permissible differences (MPD) were ±1.2 ng/L, ±20%, and ±15% for cTnT concentrations <6 ng/L, ≥6-≤24 ng/L, and >24 ng/L, respectively, allowing one exceedance per condition.

Results: In total, >2000 measurements were performed in 18 distinct samples per matrix. Median cTnT concentrations were 49.3 ng/L (8.4-782) in Li-Hep and 49.8 ng/L (8.5-822) in serum. The observed deviation in recovery was below the MPD for all storage conditions and acceptance criteria were met.

Conclusions: Stability of cTnT using the sixth-generation hs-cTnT test was demonstrated across different storage conditions for serum and Li-Hep.

心脏肌钙蛋白T稳定性研究:第6代高灵敏度检测的监管批准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Applied Laboratory Medicine
Journal of Applied Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
3.70
自引率
5.00%
发文量
137
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书