Avshalom Shaffer, Daniel Possti, Yizhaq Shmayahu, Deganit Barak-Shinar, Roy Beigel, David Hochstein, Rona Haker, Yael Hanein, Hila Meiri, Oliana Vazhgovsky, Shai Tejman-Yarden
{"title":"Interval-Level Validation of a Wearable Dry-Electrode ECG Biosensor Platform: A Proof-of-Concept Study.","authors":"Avshalom Shaffer, Daniel Possti, Yizhaq Shmayahu, Deganit Barak-Shinar, Roy Beigel, David Hochstein, Rona Haker, Yael Hanein, Hila Meiri, Oliana Vazhgovsky, Shai Tejman-Yarden","doi":"10.3390/bios16090481","DOIUrl":null,"url":null,"abstract":"<p><p>Wearable dry-electrode electrocardiographic (ECG) systems may enable longer and more comfortable rhythm monitoring, but their interval-level measurement fidelity requires validation against reference acquisition systems. In this single-center proof-of-concept study, 20 participants (10 healthy volunteers and 10 cardiology patients) underwent simultaneous ECG recordings with a U.S. Food and Drug Administration (FDA)-cleared wearable dry-electrode platform (X-trodes System M) and a wired reference system (MP150 with ECG100C module, Biopac Systems). Three-minute paired segments were recorded and underwent temporal synchronization, bandpass filtering, baseline correction, automated discrete-wavelet peak detection in NeuroKit2 (version 0.2.10), and standardized manual adjudication. After predefined exclusions for temporal alignment and P-wave detection quality, agreement was assessed in nine cardiac and ten healthy participants for RR intervals and in seven cardiac and eight healthy participants for P-to-R peak intervals. RR interval agreement was excellent in the healthy and cardiac cohorts, with a mean bias of 0.4 ± 3.1 ms and 0.1 ± 4.6 ms and mean absolute error (MAE) of 0.97 ms and 1.28 ms, respectively. Intraclass correlation coefficient (ICC) was over 0.99 for both cohorts. P-to-R peak interval agreement was strong, with an MAE of 1.76 ms and 7.14 ms and ICC of 0.96 and 0.98, respectively. These findings provide preliminary evidence of interval-level agreement for the X-trodes wearable dry-electrode ECG biosensor platform under controlled, short-duration daytime conditions. Because the matched interval pairs were clustered within a small number of participants, the findings should be regarded as proof-of-concept evidence supporting larger ambulatory validation studies rather than as definitive clinical validation.</p>","PeriodicalId":48608,"journal":{"name":"Biosensors-Basel","volume":"16 9","pages":""},"PeriodicalIF":6.2000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13604097/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosensors-Basel","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3390/bios16090481","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Wearable dry-electrode electrocardiographic (ECG) systems may enable longer and more comfortable rhythm monitoring, but their interval-level measurement fidelity requires validation against reference acquisition systems. In this single-center proof-of-concept study, 20 participants (10 healthy volunteers and 10 cardiology patients) underwent simultaneous ECG recordings with a U.S. Food and Drug Administration (FDA)-cleared wearable dry-electrode platform (X-trodes System M) and a wired reference system (MP150 with ECG100C module, Biopac Systems). Three-minute paired segments were recorded and underwent temporal synchronization, bandpass filtering, baseline correction, automated discrete-wavelet peak detection in NeuroKit2 (version 0.2.10), and standardized manual adjudication. After predefined exclusions for temporal alignment and P-wave detection quality, agreement was assessed in nine cardiac and ten healthy participants for RR intervals and in seven cardiac and eight healthy participants for P-to-R peak intervals. RR interval agreement was excellent in the healthy and cardiac cohorts, with a mean bias of 0.4 ± 3.1 ms and 0.1 ± 4.6 ms and mean absolute error (MAE) of 0.97 ms and 1.28 ms, respectively. Intraclass correlation coefficient (ICC) was over 0.99 for both cohorts. P-to-R peak interval agreement was strong, with an MAE of 1.76 ms and 7.14 ms and ICC of 0.96 and 0.98, respectively. These findings provide preliminary evidence of interval-level agreement for the X-trodes wearable dry-electrode ECG biosensor platform under controlled, short-duration daytime conditions. Because the matched interval pairs were clustered within a small number of participants, the findings should be regarded as proof-of-concept evidence supporting larger ambulatory validation studies rather than as definitive clinical validation.
Biosensors-BaselBiochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
6.60
自引率
14.80%
发文量
983
审稿时长
11 weeks
期刊介绍:
Biosensors (ISSN 2079-6374) provides an advanced forum for studies related to the science and technology of biosensors and biosensing. It publishes original research papers, comprehensive reviews and communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced. Electronic files and software regarding the full details of the calculation or experimental procedure, if unable to be published in a normal way, can be deposited as supplementary electronic material.