Flexural Stiffness and Force Retention of Six 3D-Printed Thermoplastics for Orthodontic Appliances: An In Vitro Screening Study.

IF 3.4 Q2 DENTISTRY, ORAL SURGERY & MEDICINE
Marco Serafin, Marina Borgese, Elisa Boccalari, Gilberto Binda, Mario Raspanti, Piero Antonio Zecca
{"title":"Flexural Stiffness and Force Retention of Six 3D-Printed Thermoplastics for Orthodontic Appliances: An In Vitro Screening Study.","authors":"Marco Serafin, Marina Borgese, Elisa Boccalari, Gilberto Binda, Mario Raspanti, Piero Antonio Zecca","doi":"10.3390/dj14090565","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background/Objectives:</b> Directly printed orthodontic appliances rely mostly on vat-photopolymerisation resins; pre-polymerised thermoplastics processed by fused deposition modelling (FDM) and selective laser sintering (SLS) are a less-explored, single-component route, but are poorly characterised. This study compared six thermoplastic material-process systems, supplied for biomedical or dental use, using a single intraoral-simulating protocol. <b>Methods:</b> Thin specimens (25 × 5 × 0.6 mm) of polyethylene terephthalate glycol (PETG), polycarbonate (PC), polyamide (PA), polyether-ether-ketone (PEEK) and a thermoplastic copolyester elastomer (TPC) printed by FDM, and of polyamide 12 (PA12) by SLS, underwent three-point bending and 10 min stress relaxation at 37 °C in artificial saliva (<i>n</i> = 5). Differences were assessed using the Kruskal-Wallis test with Dunn post hoc comparisons (Holm correction), and Fourier-transform infrared (FTIR) spectroscopy was used to compare virgin and printed materials for the five FDM polymers. <b>Results:</b> Apparent flexural moduli were 989 (PA12), 750 (PC), 687 (PEEK), 509 (PETG), 85 (PA) and 32 MPa (TPC), a very large material effect (ε<sup>2</sup> = 0.95), with post hoc comparisons separating only the extremes of that range. Relaxation over 10 min ranged from 2.6% (PC) to 22.1% (PA12) among the five materials in which it could be quantified, with TPC falling below the load-cell resolution; expressed as force, PA12 lost the most (0.86 N) and yet still delivered 3.01 N, which was above the initial force of PEEK, PETG, PA and TPC, though not separable from that of PC, which retained 97.4% of its own. Virgin and printed spectra were concordant (Pearson r ≥ 0.986, against ≤ 0.80 between different materials). <b>Conclusions:</b> The PA12/SLS system was the stiffest, followed by PC and PEEK, but, as the only laser-sintered material, its polymer and process effects cannot be separated. PETG and PC combine useful stiffness with accessible FDM; PA and TPC are far more compliant. These specimen-level findings provide a screening basis for selecting material-process systems for printed orthodontic appliances, preceding device-level validation.</p>","PeriodicalId":11269,"journal":{"name":"Dentistry Journal","volume":"14 9","pages":""},"PeriodicalIF":3.4000,"publicationDate":"2026-09-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13605013/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dentistry Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/dj14090565","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Background/Objectives: Directly printed orthodontic appliances rely mostly on vat-photopolymerisation resins; pre-polymerised thermoplastics processed by fused deposition modelling (FDM) and selective laser sintering (SLS) are a less-explored, single-component route, but are poorly characterised. This study compared six thermoplastic material-process systems, supplied for biomedical or dental use, using a single intraoral-simulating protocol. Methods: Thin specimens (25 × 5 × 0.6 mm) of polyethylene terephthalate glycol (PETG), polycarbonate (PC), polyamide (PA), polyether-ether-ketone (PEEK) and a thermoplastic copolyester elastomer (TPC) printed by FDM, and of polyamide 12 (PA12) by SLS, underwent three-point bending and 10 min stress relaxation at 37 °C in artificial saliva (n = 5). Differences were assessed using the Kruskal-Wallis test with Dunn post hoc comparisons (Holm correction), and Fourier-transform infrared (FTIR) spectroscopy was used to compare virgin and printed materials for the five FDM polymers. Results: Apparent flexural moduli were 989 (PA12), 750 (PC), 687 (PEEK), 509 (PETG), 85 (PA) and 32 MPa (TPC), a very large material effect (ε2 = 0.95), with post hoc comparisons separating only the extremes of that range. Relaxation over 10 min ranged from 2.6% (PC) to 22.1% (PA12) among the five materials in which it could be quantified, with TPC falling below the load-cell resolution; expressed as force, PA12 lost the most (0.86 N) and yet still delivered 3.01 N, which was above the initial force of PEEK, PETG, PA and TPC, though not separable from that of PC, which retained 97.4% of its own. Virgin and printed spectra were concordant (Pearson r ≥ 0.986, against ≤ 0.80 between different materials). Conclusions: The PA12/SLS system was the stiffest, followed by PC and PEEK, but, as the only laser-sintered material, its polymer and process effects cannot be separated. PETG and PC combine useful stiffness with accessible FDM; PA and TPC are far more compliant. These specimen-level findings provide a screening basis for selecting material-process systems for printed orthodontic appliances, preceding device-level validation.

六种3d打印热塑性塑料正畸矫治器的弯曲刚度和力保持:体外筛选研究。
背景/目的:直接打印的正畸矫治器主要依赖于光聚合树脂;通过熔融沉积建模(FDM)和选择性激光烧结(SLS)加工的预聚合热塑性塑料是一种较少探索的单组分路线,但表征不佳。本研究比较了用于生物医学或牙科用途的六种热塑性材料工艺系统,使用单一的口内模拟方案。方法:FDM打印的聚对苯二甲酸乙二醇酯(PETG)、聚碳酸酯(PC)、聚酰胺(PA)、聚醚醚酮(PEEK)和热塑性共聚弹性体(TPC)和SLS打印的聚酰胺12 (PA12)的薄样品(25 × 5 × 0.6 mm)在37℃的人工唾液中进行三点弯曲和10 min应力松弛(n = 5)。使用Kruskal-Wallis测试和Dunn事后比较(Holm校正)来评估差异,并使用傅里叶变换红外(FTIR)光谱来比较五种FDM聚合物的原始材料和印刷材料。结果:表观弯曲模量分别为989 (PA12)、750 (PC)、687 (PEEK)、509 (PETG)、85 (PA)和32 MPa (TPC),这是一个非常大的材料效应(ε2 = 0.95),事后比较只分离了该范围的极端值。5种材料在10分钟内的弛豫在2.6% (PC)到22.1% (PA12)之间,其中TPC低于测力元件分辨率;以力表示,PA12损失最大(0.86 N),但仍输出3.01 N,高于PEEK、PETG、PA和TPC的初始力,但与PC的初始力不可分,PC的初始力保留了97.4%。原始光谱与印刷光谱一致(皮尔逊r≥0.986,不同材料间皮尔逊r≤0.80)。结论:PA12/SLS体系的硬度最高,其次是PC和PEEK,但作为唯一的激光烧结材料,其高分子效应和工艺效应是不可分割的。PETG和PC结合了有用的刚度和可访问的FDM;PA和TPC的兼容性要高得多。这些标本水平的发现为选择打印正畸矫治器的材料-工艺系统提供了筛选基础,然后进行设备级验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Dentistry Journal
Dentistry Journal Dentistry-Dentistry (all)
CiteScore
3.70
自引率
7.70%
发文量
213
审稿时长
11 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书