Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study.
{"title":"Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study.","authors":"Cemal Kocaaslan","doi":"10.14744/nci.2026.08522","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Arteriovenous fistula (AVF) is the gold standard for hemodialysis access, but traditional dual-needle cannulation is limited by a restricted cannulation area and progressive stenosis; single-puncture systems may reduce these risks if hemodynamics can be optimized to prevent recirculation. This study aimed to design a novel single-puncture dual-lumen coaxial catheter and to evaluate its hemodynamic performance, flow efficiency, and recirculation rate using high-resolution computational fluid dynamics (CFD).</p><p><strong>Methods: </strong>In this CFD and design-optimization study, a dual-lumen coaxial catheter with a unique fenestration geometry was designed and analyzed using SolidWorks Flow Simulation at a physiological temperature of 37°C. Boundary conditions included an AVF flow rate of 1000 mL/min and a dialysis pump speed of 300 mL/min. The primary endpoints were flow velocity, shear stress, and recirculation control.</p><p><strong>Results: </strong>CFD simulation (1,030,641 cells) showed a maximum flow velocity of 12.69 m/s and a peak calculated shear stress of 2,602.45 Pa at the narrowest lumen section; because blood damage depends on both stress magnitude and exposure time, these computational data alone do not establish blood compatibility. Thermodilution analysis yielded a theoretical recirculation rate of 0.007% under the modeled conditions, and flow trajectories showed spatial separation of the venous jet from the arterial intake.</p><p><strong>Conclusion: </strong>Under the modeled steady-state conditions (AVF flow, 1000 mL/min; dialysis pump flow, 300 mL/min), CFD predicted stable separation of the return and aspiration streams and a theoretical recirculation rate of 0.007%. These proof-of-concept findings support further geometric optimization and physical testing; they do not establish clinical safety, efficacy, or superiority over standard dual-needle cannulation.</p>","PeriodicalId":94347,"journal":{"name":"Northern clinics of Istanbul","volume":"13 4","pages":"460-468"},"PeriodicalIF":0.9000,"publicationDate":"2026-09-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13591039/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Northern clinics of Istanbul","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14744/nci.2026.08522","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Arteriovenous fistula (AVF) is the gold standard for hemodialysis access, but traditional dual-needle cannulation is limited by a restricted cannulation area and progressive stenosis; single-puncture systems may reduce these risks if hemodynamics can be optimized to prevent recirculation. This study aimed to design a novel single-puncture dual-lumen coaxial catheter and to evaluate its hemodynamic performance, flow efficiency, and recirculation rate using high-resolution computational fluid dynamics (CFD).
Methods: In this CFD and design-optimization study, a dual-lumen coaxial catheter with a unique fenestration geometry was designed and analyzed using SolidWorks Flow Simulation at a physiological temperature of 37°C. Boundary conditions included an AVF flow rate of 1000 mL/min and a dialysis pump speed of 300 mL/min. The primary endpoints were flow velocity, shear stress, and recirculation control.
Results: CFD simulation (1,030,641 cells) showed a maximum flow velocity of 12.69 m/s and a peak calculated shear stress of 2,602.45 Pa at the narrowest lumen section; because blood damage depends on both stress magnitude and exposure time, these computational data alone do not establish blood compatibility. Thermodilution analysis yielded a theoretical recirculation rate of 0.007% under the modeled conditions, and flow trajectories showed spatial separation of the venous jet from the arterial intake.
Conclusion: Under the modeled steady-state conditions (AVF flow, 1000 mL/min; dialysis pump flow, 300 mL/min), CFD predicted stable separation of the return and aspiration streams and a theoretical recirculation rate of 0.007%. These proof-of-concept findings support further geometric optimization and physical testing; they do not establish clinical safety, efficacy, or superiority over standard dual-needle cannulation.