{"title":"Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study.","authors":"Laura Valentina Klüner, Sebastian Zensen, Hannah Peuster, Melanie Ebenau, Denise Bos, Raya Ocker-Serger, Yannick Laurent Thal, Luca Salhöfer, Florian Stehling, Daniel Bernard, Erik Büscher, Matthias Welsner, Lale Umutlu, Michael Forsting, Johannes Haubold, Marcel Klaus Opitz","doi":"10.1055/a-2935-4084","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Photon-counting CT (PCCT) and spectral shaping enable new approaches in pediatric imaging, but the effects of clinically implemented tin-filtered pediatric chest protocols on radiation dose and image noise across image quality levels (IQLs) remain insufficiently characterized. This study compared a clinical Sn100 kVp ultra-high resolution (UHR) protocol with a non-tin-filtered 90 kVp standard-resolution protocol using anthropomorphic pediatric phantoms.</p><p><strong>Materials and methods: </strong>Three pediatric phantoms (1, 5, and 10 years) were scanned on a commercial PCCT system across IQLs 1-300 using the clinical Sn100 kVp UHR and non-tin-filtered 90 kVp standard-resolution protocols. Radiation dose and ROI-based image noise in lung parenchyma and bone were recorded.</p><p><strong>Results: </strong>The Sn100 kVp UHR protocol showed lower lung-parenchymal noise across all phantom sizes and IQLs and generally lower bone noise. In the ultra-low-dose range (IQL 1-10), lung-parenchymal noise reductions ranged from 31% to 52% and bone-noise reductions reached 58%. Across most IQLs, the Sn100 kVp UHR protocol was associated with higher radiation dose. At high IQLs, dose plateaus occurred at approximately IQL 150 and 100 in the 5- and 10-year-old phantoms, followed by dose crossovers at approximately IQL 220 and 150, respectively. At very low IQLs, the Sn100 kVp UHR protocol showed a dose advantage up to IQL 5, 4, and 3 in the 1-, 5-, and 10-year-old phantoms, respectively.</p><p><strong>Conclusions: </strong>The Sn100 kVp UHR protocol was associated with lower image noise but was not universally dose-saving compared with the non-tin-filtered 90 kVp standard-resolution protocol. In addition, the observed dose behavior was influenced not only by spectral shaping but also by detector readout mode and protocol-specific technical constraints. These findings may support future pediatric chest PCCT protocol optimization.</p><p><strong>Key points: </strong>· The clinical Sn100 kVp UHR protocol was associated with lower lung-parenchymal noise and generally lower bone noise compared with the non-tin-filtered 90 kVp standard-resolution protocol.. · The dose advantage of the Sn100 kVp UHR protocol was limited to very low IQLs, while most IQLs showed higher radiation dose compared with the 90 kVp standard-resolution protocol.. · At high image quality levels, the Sn100 kVp UHR protocol exhibited dose plateaus, limiting further dose increase despite higher target image quality..</p><p><strong>Citation format: </strong>· Klüner LV, Zensen S, Peuster H et al. Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study. Rofo 2026; DOI 10.1055/a-2935-4084.</p>","PeriodicalId":21490,"journal":{"name":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2935-4084","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Photon-counting CT (PCCT) and spectral shaping enable new approaches in pediatric imaging, but the effects of clinically implemented tin-filtered pediatric chest protocols on radiation dose and image noise across image quality levels (IQLs) remain insufficiently characterized. This study compared a clinical Sn100 kVp ultra-high resolution (UHR) protocol with a non-tin-filtered 90 kVp standard-resolution protocol using anthropomorphic pediatric phantoms.
Materials and methods: Three pediatric phantoms (1, 5, and 10 years) were scanned on a commercial PCCT system across IQLs 1-300 using the clinical Sn100 kVp UHR and non-tin-filtered 90 kVp standard-resolution protocols. Radiation dose and ROI-based image noise in lung parenchyma and bone were recorded.
Results: The Sn100 kVp UHR protocol showed lower lung-parenchymal noise across all phantom sizes and IQLs and generally lower bone noise. In the ultra-low-dose range (IQL 1-10), lung-parenchymal noise reductions ranged from 31% to 52% and bone-noise reductions reached 58%. Across most IQLs, the Sn100 kVp UHR protocol was associated with higher radiation dose. At high IQLs, dose plateaus occurred at approximately IQL 150 and 100 in the 5- and 10-year-old phantoms, followed by dose crossovers at approximately IQL 220 and 150, respectively. At very low IQLs, the Sn100 kVp UHR protocol showed a dose advantage up to IQL 5, 4, and 3 in the 1-, 5-, and 10-year-old phantoms, respectively.
Conclusions: The Sn100 kVp UHR protocol was associated with lower image noise but was not universally dose-saving compared with the non-tin-filtered 90 kVp standard-resolution protocol. In addition, the observed dose behavior was influenced not only by spectral shaping but also by detector readout mode and protocol-specific technical constraints. These findings may support future pediatric chest PCCT protocol optimization.
Key points: · The clinical Sn100 kVp UHR protocol was associated with lower lung-parenchymal noise and generally lower bone noise compared with the non-tin-filtered 90 kVp standard-resolution protocol.. · The dose advantage of the Sn100 kVp UHR protocol was limited to very low IQLs, while most IQLs showed higher radiation dose compared with the 90 kVp standard-resolution protocol.. · At high image quality levels, the Sn100 kVp UHR protocol exhibited dose plateaus, limiting further dose increase despite higher target image quality..
Citation format: · Klüner LV, Zensen S, Peuster H et al. Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study. Rofo 2026; DOI 10.1055/a-2935-4084.
期刊介绍:
Die RöFo veröffentlicht Originalarbeiten, Übersichtsartikel und Fallberichte aus dem Bereich der Radiologie und den weiteren bildgebenden Verfahren in der Medizin. Es dürfen nur Arbeiten eingereicht werden, die noch nicht veröffentlicht sind und die auch nicht gleichzeitig einer anderen Zeitschrift zur Veröffentlichung angeboten wurden. Alle eingereichten Beiträge unterliegen einer sorgfältigen fachlichen Begutachtung.
Gegründet 1896 – nur knapp 1 Jahr nach der Entdeckung der Röntgenstrahlen durch C.W. Röntgen – blickt die RöFo auf über 100 Jahre Erfahrung als wichtigstes Publikationsmedium in der deutschsprachigen Radiologie zurück. Sie ist damit die älteste radiologische Fachzeitschrift und schafft es erfolgreich, lange Kontinuität mit dem Anspruch an wissenschaftliches Publizieren auf internationalem Niveau zu verbinden. Durch ihren zentralen Platz im Verlagsprogramm stellte die RöFo die Basis für das heute umfassende und erfolgreiche Radiologie-Medienangebot im Georg Thieme Verlag.
Besonders eng verbunden ist die RöFo mit der Geschichte der Röntgengesellschaften in Deutschland und Österreich. Sie ist offizielles Organ von DRG und ÖRG und die Mitglieder der Fachgesellschaften erhalten die Zeitschrift im Rahmen ihrer Mitgliedschaft. Mit ihrem wissenschaftlichen Kernteil und dem eigenen Mitteilungsteil der Fachgesellschaften bietet die RöFo Monat für Monat ein Forum für den Austausch von Inhalten und Botschaften der radiologischen Community im deutschsprachigen Raum.