Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations.
IF 2 4区 医学Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Christoph G Lisson, Jochen Breitruck, Konrad Görich, Benedikt Haggenmüller, Nico Sollmann, Arthur P Wunderlich, Meinrad Beer, Catharina S Lisson
{"title":"Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations.","authors":"Christoph G Lisson, Jochen Breitruck, Konrad Görich, Benedikt Haggenmüller, Nico Sollmann, Arthur P Wunderlich, Meinrad Beer, Catharina S Lisson","doi":"10.1055/a-2944-0126","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate radiation exposure in peripheral, abdominal, and cerebral digital subtraction angiography (DSA) across three consecutive equipment generations and to identify technological potential for dose reduction.</p><p><strong>Material and methods: </strong>In this retrospective single-center study, 450 patients were examined using three different platforms (150 per system): a conventional system (Allura Xper) and two subsequent generations featuring advanced noise reduction (Allura Clarity and Azurion). The patient cohort was equally distributed among head, abdomen, and lower limb regions. Primary endpoints included total dose-area product (DAP), fluoroscopic dose rate, dose per DSA frame, and the DAP per 3D rotational angiography (3D-RA) for the head. Subjective image quality was evaluated using a 5-point Likert scale.</p><p><strong>Results: </strong>Transitioning from the conventional system to the first generation with advanced image processing (Allura Clarity) resulted in a significant reduction in total DAP of 70% (abdomen), 65% (head), and 69% (lower limbs; each p<0.001). This effect was driven by a significant decrease in fluoroscopic dose rate (34-50%) and dose per DSA frame (68-74%). Transitioning to the latest system generation (Azurion) yielded a non-significant yet further reduction of 6-17%. Cerebral 3D-RA demonstrated stable dose requirements across all three generations (p=0.123). Subjective image quality improved significantly with the introduction of noise reduction algorithms (p<0.001), with no significant difference between the two most recent systems (p>0.05).</p><p><strong>Conclusions: </strong>Advanced image processing algorithms are an effective tool for radiation dose reduction in angiography. However, a technological dose plateau appears to have been reached, as the current equipment generation has not shown further significant improvements in radiation dose requirements compared to its immediate predecessor. Dose requirements for 3D-RA do not, at present, benefit significantly from the advantages of modern image processing algorithms.</p><p><strong>Key points: </strong>· State-of-the-art image processing markedly reduces radiation exposure.. · The largest effect was found with Allura Clarity.. · Azurion showed no additional significant dose reduction.. · DSA benefited more than fluoroscopy.. · 3D rotational angiography showed no significant difference in dose..</p><p><strong>Citation format: </strong>· Lisson CG, Breitruck J, Görich K et al. Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations. Rofo 2026; DOI 10.1055/a-2944-0126.</p>","PeriodicalId":21490,"journal":{"name":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2944-0126","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate radiation exposure in peripheral, abdominal, and cerebral digital subtraction angiography (DSA) across three consecutive equipment generations and to identify technological potential for dose reduction.
Material and methods: In this retrospective single-center study, 450 patients were examined using three different platforms (150 per system): a conventional system (Allura Xper) and two subsequent generations featuring advanced noise reduction (Allura Clarity and Azurion). The patient cohort was equally distributed among head, abdomen, and lower limb regions. Primary endpoints included total dose-area product (DAP), fluoroscopic dose rate, dose per DSA frame, and the DAP per 3D rotational angiography (3D-RA) for the head. Subjective image quality was evaluated using a 5-point Likert scale.
Results: Transitioning from the conventional system to the first generation with advanced image processing (Allura Clarity) resulted in a significant reduction in total DAP of 70% (abdomen), 65% (head), and 69% (lower limbs; each p<0.001). This effect was driven by a significant decrease in fluoroscopic dose rate (34-50%) and dose per DSA frame (68-74%). Transitioning to the latest system generation (Azurion) yielded a non-significant yet further reduction of 6-17%. Cerebral 3D-RA demonstrated stable dose requirements across all three generations (p=0.123). Subjective image quality improved significantly with the introduction of noise reduction algorithms (p<0.001), with no significant difference between the two most recent systems (p>0.05).
Conclusions: Advanced image processing algorithms are an effective tool for radiation dose reduction in angiography. However, a technological dose plateau appears to have been reached, as the current equipment generation has not shown further significant improvements in radiation dose requirements compared to its immediate predecessor. Dose requirements for 3D-RA do not, at present, benefit significantly from the advantages of modern image processing algorithms.
Key points: · State-of-the-art image processing markedly reduces radiation exposure.. · The largest effect was found with Allura Clarity.. · Azurion showed no additional significant dose reduction.. · DSA benefited more than fluoroscopy.. · 3D rotational angiography showed no significant difference in dose..
Citation format: · Lisson CG, Breitruck J, Görich K et al. Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations. Rofo 2026; DOI 10.1055/a-2944-0126.
期刊介绍:
Die RöFo veröffentlicht Originalarbeiten, Übersichtsartikel und Fallberichte aus dem Bereich der Radiologie und den weiteren bildgebenden Verfahren in der Medizin. Es dürfen nur Arbeiten eingereicht werden, die noch nicht veröffentlicht sind und die auch nicht gleichzeitig einer anderen Zeitschrift zur Veröffentlichung angeboten wurden. Alle eingereichten Beiträge unterliegen einer sorgfältigen fachlichen Begutachtung.
Gegründet 1896 – nur knapp 1 Jahr nach der Entdeckung der Röntgenstrahlen durch C.W. Röntgen – blickt die RöFo auf über 100 Jahre Erfahrung als wichtigstes Publikationsmedium in der deutschsprachigen Radiologie zurück. Sie ist damit die älteste radiologische Fachzeitschrift und schafft es erfolgreich, lange Kontinuität mit dem Anspruch an wissenschaftliches Publizieren auf internationalem Niveau zu verbinden. Durch ihren zentralen Platz im Verlagsprogramm stellte die RöFo die Basis für das heute umfassende und erfolgreiche Radiologie-Medienangebot im Georg Thieme Verlag.
Besonders eng verbunden ist die RöFo mit der Geschichte der Röntgengesellschaften in Deutschland und Österreich. Sie ist offizielles Organ von DRG und ÖRG und die Mitglieder der Fachgesellschaften erhalten die Zeitschrift im Rahmen ihrer Mitgliedschaft. Mit ihrem wissenschaftlichen Kernteil und dem eigenen Mitteilungsteil der Fachgesellschaften bietet die RöFo Monat für Monat ein Forum für den Austausch von Inhalten und Botschaften der radiologischen Community im deutschsprachigen Raum.