Automated chemiluminescence immunoassay for the detection of anti-nephrin autoantibodies.

IF 21.8 1区 医学 Q1 UROLOGY & NEPHROLOGY
Rui Gu, Jiayu Zhang, Chungen Qian, Gang Wang, Mingyu Lai, Jiayi Shen, Jianhua Mao, Qing Ye
{"title":"Automated chemiluminescence immunoassay for the detection of anti-nephrin autoantibodies.","authors":"Rui Gu, Jiayu Zhang, Chungen Qian, Gang Wang, Mingyu Lai, Jiayi Shen, Jianhua Mao, Qing Ye","doi":"10.1016/j.kint.2026.07.036","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Circulating anti-nephrin autoantibodies are being increasingly recognized in autoimmune podocytopathies, particularly minimal change disease (MCD) and focal segmental glomerulosclerosis (FSGS). However, circulating anti-nephrin reactivity is often low, and currently available assays differ substantially in design, antigen production, and analytical performance. This methodological heterogeneity highlights the need for standardized and quantitative assays for clinical application.</p><p><strong>Methods: </strong>Recombinant human nephrin ectodomain-coated magnetic microparticles were used to establish an automated chemiluminescence immunoassay (CLIA). Serum anti-nephrin IgG was measured in 121 children with biopsy-confirmed MCD (n=103) or primary FSGS (n=18), together with disease controls (genetically confirmed MCD/FSGS, IgA nephropathy, IgA vasculitis with nephritis, lupus nephritis, and ANCA-associated glomerulonephritis) and healthy individuals. Samples were tested in parallel by immunoprecipitation-Western blot (IP-WB). Antigen inhibition assays and cohort-based clinical analyses were performed.</p><p><strong>Results: </strong>The CLIA met predefined analytical performance criteria and showed substantial agreement with IP-WB, with an overall concordance of 86.8% and Cohen's kappa coefficient of 0.714. CLIA detected anti-nephrin IgG more frequently than IP-WB in patients with MCD or FSGS (41.3% vs. 28.1%), including additional low-level positive samples. Antigen inhibition assays showed dose-dependent inhibition in CLIA-positive samples. Among patients with baseline nephrotic-range proteinuria, antibody positivity was associated with greater proteinuria, lower serum albumin, and steroid-sensitive nephrotic syndrome. Patients with elevated anti-nephrin IgG levels had significantly shorter relapse-free survival. During complete remission, anti-nephrin IgG positivity was associated with relapse within three months, with relapse risk increasing according to antibody level.</p><p><strong>Conclusions: </strong>Together, our automated CLIA may help move anti-nephrin IgG testing from specialized research workflows toward standardized clinical application, supporting future use in disease stratification and longitudinal monitoring.</p>","PeriodicalId":17801,"journal":{"name":"Kidney international","volume":" ","pages":""},"PeriodicalIF":21.8000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kidney international","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.kint.2026.07.036","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Circulating anti-nephrin autoantibodies are being increasingly recognized in autoimmune podocytopathies, particularly minimal change disease (MCD) and focal segmental glomerulosclerosis (FSGS). However, circulating anti-nephrin reactivity is often low, and currently available assays differ substantially in design, antigen production, and analytical performance. This methodological heterogeneity highlights the need for standardized and quantitative assays for clinical application.

Methods: Recombinant human nephrin ectodomain-coated magnetic microparticles were used to establish an automated chemiluminescence immunoassay (CLIA). Serum anti-nephrin IgG was measured in 121 children with biopsy-confirmed MCD (n=103) or primary FSGS (n=18), together with disease controls (genetically confirmed MCD/FSGS, IgA nephropathy, IgA vasculitis with nephritis, lupus nephritis, and ANCA-associated glomerulonephritis) and healthy individuals. Samples were tested in parallel by immunoprecipitation-Western blot (IP-WB). Antigen inhibition assays and cohort-based clinical analyses were performed.

Results: The CLIA met predefined analytical performance criteria and showed substantial agreement with IP-WB, with an overall concordance of 86.8% and Cohen's kappa coefficient of 0.714. CLIA detected anti-nephrin IgG more frequently than IP-WB in patients with MCD or FSGS (41.3% vs. 28.1%), including additional low-level positive samples. Antigen inhibition assays showed dose-dependent inhibition in CLIA-positive samples. Among patients with baseline nephrotic-range proteinuria, antibody positivity was associated with greater proteinuria, lower serum albumin, and steroid-sensitive nephrotic syndrome. Patients with elevated anti-nephrin IgG levels had significantly shorter relapse-free survival. During complete remission, anti-nephrin IgG positivity was associated with relapse within three months, with relapse risk increasing according to antibody level.

Conclusions: Together, our automated CLIA may help move anti-nephrin IgG testing from specialized research workflows toward standardized clinical application, supporting future use in disease stratification and longitudinal monitoring.

自动化学发光免疫分析法检测抗肾素自身抗体。
导论:循环抗肾素自身抗体在自身免疫性足细胞病变,特别是微小改变病(MCD)和局灶节段性肾小球硬化(FSGS)中被越来越多地发现。然而,循环抗肾素反应性通常较低,目前可用的检测方法在设计、抗原产生和分析性能方面存在很大差异。这种方法的异质性强调了临床应用标准化和定量分析的必要性。方法:采用重组人肾素外膜结构域磁微粒建立自动化学发光免疫分析法(CLIA)。对121例活检证实的MCD(103例)或原发性FSGS(18例)患儿以及疾病对照(遗传证实的MCD/FSGS、IgA肾病、IgA血管炎伴肾炎、狼疮性肾炎和anca相关性肾小球肾炎)和健康人进行血清抗肾素IgG检测。采用免疫沉淀- western blot (IP-WB)平行检测。进行抗原抑制试验和基于队列的临床分析。结果:CLIA符合预先设定的分析性能标准,与IP-WB基本一致,总体一致性为86.8%,Cohen’s kappa系数为0.714。在MCD或FSGS患者中,CLIA检测到抗肾素IgG的频率高于IP-WB(41.3%对28.1%),包括额外的低水平阳性样本。抗原抑制试验显示在clia阳性样品中存在剂量依赖性抑制。在基线肾范围蛋白尿患者中,抗体阳性与蛋白尿增多、血清白蛋白降低和类固醇敏感性肾病综合征相关。抗肾素IgG水平升高的患者无复发生存期明显缩短。在完全缓解期,抗肾素IgG阳性与3个月内复发相关,且随抗体水平升高复发风险增加。结论:总之,我们的自动化CLIA可能有助于将抗肾素IgG检测从专门的研究工作流程转向标准化的临床应用,支持未来在疾病分层和纵向监测中的应用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Kidney international
Kidney international 医学-泌尿学与肾脏学
CiteScore
23.30
自引率
3.10%
发文量
490
审稿时长
3-6 weeks
期刊介绍: Kidney International (KI), the official journal of the International Society of Nephrology, is led by Dr. Pierre Ronco (Paris, France) and stands as one of nephrology's most cited and esteemed publications worldwide. KI provides exceptional benefits for both readers and authors, featuring highly cited original articles, focused reviews, cutting-edge imaging techniques, and lively discussions on controversial topics. The journal is dedicated to kidney research, serving researchers, clinical investigators, and practicing nephrologists.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书