Clinical Implementation and Performance of Real-Time AI System for Device Tracking in Neurointervention: First U.S. Evaluation of Neuro-Vascular Assist.
João Victor Sanders, Joshua Jimenez, Marion Oliver, Kenichi Kono, Krishna C Joshi, Demetrius Lopes
{"title":"Clinical Implementation and Performance of Real-Time AI System for Device Tracking in Neurointervention: First U.S. Evaluation of Neuro-Vascular Assist.","authors":"João Victor Sanders, Joshua Jimenez, Marion Oliver, Kenichi Kono, Krishna C Joshi, Demetrius Lopes","doi":"10.1007/s00062-026-01723-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the first U.S.-based implementation, notification accuracy, and temporal operator response of Neuro-Vascular Assist.</p><p><strong>Methods: </strong>In a single-center study, 21 consecutive neurointerventional cases were performed using the AI system integrated with a biplane angiographic system. Procedures were classified as \"Standard\", \"Coil\", \"Filter\", or \"Venous\". Neuro-Vascular Assist tracked devices including guidewires, stents, coils, and filters in real-time. Visual and audio notifications were made based on device entry and exit in a determined region of interest. We classified these notifications as true positives (TP), false positives (FP), and false negatives (FN), and further assessed for temporal operator response (TOR).</p><p><strong>Results: </strong>The AI system was successfully integrated into the workflow without procedural delays or system failures affecting notification delivery. A total of 232 TPs, 13 FPs, and 17 FNs were recorded, resulting in an overall precision of 94.7% and recall of 93.2%. Among the 59 TP notifications evaluated for TOR, 25 (42.4%) were followed by an observable operator response within 10 s. These 25 positive TOR events represented 10.2% of all 245 notifications.</p><p><strong>Conclusion: </strong>Neuro-Vascular Assist was successfully implemented and demonstrated preliminary accuracy and feasibility in a U.S.-based neurointerventional setting. Larger prospective studies are warranted to determine whether real-time AI-assisted device tracking influences procedural decision-making, workflow, safety, or clinical outcomes.</p>","PeriodicalId":10391,"journal":{"name":"Clinical Neuroradiology","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00062-026-01723-8","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the first U.S.-based implementation, notification accuracy, and temporal operator response of Neuro-Vascular Assist.
Methods: In a single-center study, 21 consecutive neurointerventional cases were performed using the AI system integrated with a biplane angiographic system. Procedures were classified as "Standard", "Coil", "Filter", or "Venous". Neuro-Vascular Assist tracked devices including guidewires, stents, coils, and filters in real-time. Visual and audio notifications were made based on device entry and exit in a determined region of interest. We classified these notifications as true positives (TP), false positives (FP), and false negatives (FN), and further assessed for temporal operator response (TOR).
Results: The AI system was successfully integrated into the workflow without procedural delays or system failures affecting notification delivery. A total of 232 TPs, 13 FPs, and 17 FNs were recorded, resulting in an overall precision of 94.7% and recall of 93.2%. Among the 59 TP notifications evaluated for TOR, 25 (42.4%) were followed by an observable operator response within 10 s. These 25 positive TOR events represented 10.2% of all 245 notifications.
Conclusion: Neuro-Vascular Assist was successfully implemented and demonstrated preliminary accuracy and feasibility in a U.S.-based neurointerventional setting. Larger prospective studies are warranted to determine whether real-time AI-assisted device tracking influences procedural decision-making, workflow, safety, or clinical outcomes.