{"title":"High-Precision Rb Isotope Analysis by MC-ICP-MS Using Sr Empirical External Normalization","authors":"Haoming Yin, Qi Feng, Tao Li","doi":"10.1002/rcm.70185","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Rationale</h3>\n \n <p>Stable rubidium (Rb) isotopes have received increasing attention as emerging geochemical tracers for investigating planetary evolution, water–silicate interactions, and global Rb cycling. High-precision Rb isotope measurements by multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) are commonly performed using the standard-sample bracketing (SSB) method, but this approach is sensitive to short-term instrumental drift.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Building upon the previously reported empirical external normalization (EEN) with <sup>92</sup>Zr/<sup>90</sup>Zr combined with the SSB method, we adopted <sup>88</sup>Sr/<sup>86</sup>Sr for EEN + SSB MC-ICP-MS Rb isotope measurement, owing to the closer mass discrimination between Sr and Rb. In this study, we correted the isobaric interference of <sup>87</sup>Sr on <sup>87</sup>Rb using an exponential law, compared Sr-EEN with Zr-EEN using the in-house USTC-Rb standard, and evaluated elemental doping effects (K, Ca, Na, and Mn).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The results show that the USTC-Rb standard yielded δ<sup>87/85</sup>Rb values of −0.15‰ ± 0.04‰ (2SD, <i>n</i> = 33) by Sr-EEN and −0.16‰ ± 0.04‰ (2SD, <i>n</i> = 33) by Zr-EEN, indistinguishable within analytical uncertainty. Elemental doping experiments suggest that the measured Rb isotope data were not significantly affected when the K/Rb, Ca/Rb, Na/Rb, and Mn/Rb ratios were up to 100. The long-term intermediate reproducibility for NIST SRM 984 was 0.00‰ ± 0.04‰ (2SD, <i>n</i> = 121).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Because Sr can be effectively separated from Rb using Sr-spec resin, this Sr-EEN method provides a precise and reliable alternative for Rb isotope determination by MC-ICP-MS and exhibits high tolerance to matrix elements.</p>\n </section>\n </div>","PeriodicalId":225,"journal":{"name":"Rapid Communications in Mass Spectrometry","volume":"40 23","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2026-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rapid Communications in Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/rcm.70185","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Rationale
Stable rubidium (Rb) isotopes have received increasing attention as emerging geochemical tracers for investigating planetary evolution, water–silicate interactions, and global Rb cycling. High-precision Rb isotope measurements by multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) are commonly performed using the standard-sample bracketing (SSB) method, but this approach is sensitive to short-term instrumental drift.
Methods
Building upon the previously reported empirical external normalization (EEN) with 92Zr/90Zr combined with the SSB method, we adopted 88Sr/86Sr for EEN + SSB MC-ICP-MS Rb isotope measurement, owing to the closer mass discrimination between Sr and Rb. In this study, we correted the isobaric interference of 87Sr on 87Rb using an exponential law, compared Sr-EEN with Zr-EEN using the in-house USTC-Rb standard, and evaluated elemental doping effects (K, Ca, Na, and Mn).
Results
The results show that the USTC-Rb standard yielded δ87/85Rb values of −0.15‰ ± 0.04‰ (2SD, n = 33) by Sr-EEN and −0.16‰ ± 0.04‰ (2SD, n = 33) by Zr-EEN, indistinguishable within analytical uncertainty. Elemental doping experiments suggest that the measured Rb isotope data were not significantly affected when the K/Rb, Ca/Rb, Na/Rb, and Mn/Rb ratios were up to 100. The long-term intermediate reproducibility for NIST SRM 984 was 0.00‰ ± 0.04‰ (2SD, n = 121).
Conclusions
Because Sr can be effectively separated from Rb using Sr-spec resin, this Sr-EEN method provides a precise and reliable alternative for Rb isotope determination by MC-ICP-MS and exhibits high tolerance to matrix elements.
期刊介绍:
Rapid Communications in Mass Spectrometry is a journal whose aim is the rapid publication of original research results and ideas on all aspects of the science of gas-phase ions; it covers all the associated scientific disciplines. There is no formal limit on paper length ("rapid" is not synonymous with "brief"), but papers should be of a length that is commensurate with the importance and complexity of the results being reported. Contributions may be theoretical or practical in nature; they may deal with methods, techniques and applications, or with the interpretation of results; they may cover any area in science that depends directly on measurements made upon gaseous ions or that is associated with such measurements.