Sophia Hohenstatt, Gabriele Maliandi, Dominik F Vollherbst, Noah G Bahce, Susanne Bonekamp, Martin Bendszus, Markus A Möhlenbruch
{"title":"Workflow level release of microplastics, nanoplastics, and perfluoroalkyl and polyfluoroalkyl substances during simulated neurovascular interventions.","authors":"Sophia Hohenstatt, Gabriele Maliandi, Dominik F Vollherbst, Noah G Bahce, Susanne Bonekamp, Martin Bendszus, Markus A Möhlenbruch","doi":"10.1136/jnis-2026-025917","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Catheter based neurovascular interventions rely on polymer based, coated, and fluorinated devices. Mechanical manipulation may release particulate and molecular contaminants, but integrated workflow level quantification of microplastic (MP), nanoplastic (NP), and perfluoroalkyl and polyfluoroalkyl substance (PFAS) release during neurovascular procedures remains limited. This study quantified MP, NP, and PFAS release in an in vitro simulation of neurointerventional workflows.</p><p><strong>Methods: </strong>20 neurovascular procedures were simulated under standardized in vitro conditions, including 10 mechanical thrombectomy and 10 aneurysm treatment workflows. Procedures were performed in 5 L glass containers using warmed saline, contrast medium, and clinically applied neurovascular devices. Procedural fluid was collected after each simulation and aliquoted for MP, NP, and PFAS analysis using micro-Fourier transform infrared spectroscopy, pyrolysis-gas chromatography-mass spectrometry, and ultra high performance liquid chromatography-tandem mass spectrometry, respectively.</p><p><strong>Results: </strong>Contaminants were detected across all simulated procedures. Median MP particle count was 15 particles per analyzed sample volume (IQR 9.0-27.3; range 0-58), median NP concentration was 18.19 µg/L (IQR 5.85-33.35; range 2.19-1671.10), and median sum PFAS concentration was 127 µg/L (IQR 93.5-272.0; range 47-458). MP and NP levels did not differ between thrombectomy and aneurysm simulations. Sum PFAS concentrations were higher during thrombectomy than during aneurysm treatment simulations (median 209.5 vs 98.0 µg/L; P=0.010), particularly in stent retriever compared with direct aspiration simulations (median 287.0 vs 120.0 µg/L; P=0.030). NP and sum PFAS concentrations were strongly correlated (ρ=0.689; P=0.001).</p><p><strong>Conclusions: </strong>Simulated neurovascular interventions generated measurable release of MP, NP, and PFAS into procedural fluid. Workflow level assessment may support future source specific testing, clinical validation, and material safety evaluation.</p>","PeriodicalId":16411,"journal":{"name":"Journal of NeuroInterventional Surgery","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-09-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of NeuroInterventional Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1136/jnis-2026-025917","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"NEUROIMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Catheter based neurovascular interventions rely on polymer based, coated, and fluorinated devices. Mechanical manipulation may release particulate and molecular contaminants, but integrated workflow level quantification of microplastic (MP), nanoplastic (NP), and perfluoroalkyl and polyfluoroalkyl substance (PFAS) release during neurovascular procedures remains limited. This study quantified MP, NP, and PFAS release in an in vitro simulation of neurointerventional workflows.
Methods: 20 neurovascular procedures were simulated under standardized in vitro conditions, including 10 mechanical thrombectomy and 10 aneurysm treatment workflows. Procedures were performed in 5 L glass containers using warmed saline, contrast medium, and clinically applied neurovascular devices. Procedural fluid was collected after each simulation and aliquoted for MP, NP, and PFAS analysis using micro-Fourier transform infrared spectroscopy, pyrolysis-gas chromatography-mass spectrometry, and ultra high performance liquid chromatography-tandem mass spectrometry, respectively.
Results: Contaminants were detected across all simulated procedures. Median MP particle count was 15 particles per analyzed sample volume (IQR 9.0-27.3; range 0-58), median NP concentration was 18.19 µg/L (IQR 5.85-33.35; range 2.19-1671.10), and median sum PFAS concentration was 127 µg/L (IQR 93.5-272.0; range 47-458). MP and NP levels did not differ between thrombectomy and aneurysm simulations. Sum PFAS concentrations were higher during thrombectomy than during aneurysm treatment simulations (median 209.5 vs 98.0 µg/L; P=0.010), particularly in stent retriever compared with direct aspiration simulations (median 287.0 vs 120.0 µg/L; P=0.030). NP and sum PFAS concentrations were strongly correlated (ρ=0.689; P=0.001).
Conclusions: Simulated neurovascular interventions generated measurable release of MP, NP, and PFAS into procedural fluid. Workflow level assessment may support future source specific testing, clinical validation, and material safety evaluation.
期刊介绍:
The Journal of NeuroInterventional Surgery (JNIS) is a leading peer review journal for scientific research and literature pertaining to the field of neurointerventional surgery. The journal launch follows growing professional interest in neurointerventional techniques for the treatment of a range of neurological and vascular problems including stroke, aneurysms, brain tumors, and spinal compression.The journal is owned by SNIS and is also the official journal of the Interventional Chapter of the Australian and New Zealand Society of Neuroradiology (ANZSNR), the Canadian Interventional Neuro Group, the Hong Kong Neurological Society (HKNS) and the Neuroradiological Society of Taiwan.