Hailong Li, Bohan Zhang, Elanchezhian Somasundaram, Mahdieh Shabanian, Zachary Taylor, Neeraja Mahalingam, Zhixiu Lu, Bin Zhang, Stephen W Standage, Gary R Schooler, Lili He, Alexander J Towbin
{"title":"A two-stage deep learning pipeline for automated detection and localization of endotracheal tubes on pediatric chest radiographs: a pilot study.","authors":"Hailong Li, Bohan Zhang, Elanchezhian Somasundaram, Mahdieh Shabanian, Zachary Taylor, Neeraja Mahalingam, Zhixiu Lu, Bin Zhang, Stephen W Standage, Gary R Schooler, Lili He, Alexander J Towbin","doi":"10.1007/s00247-026-06767-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Endotracheal tubes (ETTs) are critical life-support devices for mechanically ventilated pediatric patients, yet automated ETT assessment on pediatric chest radiographs (CXRs) remains limited.</p><p><strong>Objective: </strong>To develop and evaluate a two-stage deep learning pipeline for automated detection and localization of ETTs on pediatric CXRs.</p><p><strong>Materials and methods: </strong>This retrospective study included 1,000 pediatric CXRs (476 ETT-positive, 524 ETT-negative) acquired in 2021 at a single institution. ETT segmentation masks and distal tip coordinates were annotated by trained analysts and verified by pediatric radiologists. A two-stage pipeline consisting of a ResNet classification model followed by a U-Net segmentation model was developed for ETT detection and localization. Performance was evaluated on a held-out test set using multiple metrics, including the area under the receiver operating characteristic curve (AUROC) and the mean absolute error (MAE), with 95% confidence intervals (CI). Inter-observer variability was assessed as a reference for localization performance.</p><p><strong>Results: </strong>Inter-observer variability for ETT tip localization was 2.01 mm MAE on the held-out test set. The pipeline achieved an AUROC of 0.994 (95% CI 0.986, 1.000) for ETT detection. For localization, the pipeline achieved a MAE of 6.59 mm (95% CI 5.19, 8.21 mm). Incorporating the classification stage substantially reduced false-positive segmentations from 17 to 3 among ETT-negative CXRs compared with the standalone segmentation model.</p><p><strong>Conclusion: </strong>A two-stage deep learning pipeline demonstrated high performance for automated ETT detection and promising performance for tip localization on pediatric CXRs in this single-center pilot study. Further evaluation in larger and external pediatric cohorts is needed to assess generalizability.</p>","PeriodicalId":19755,"journal":{"name":"Pediatric Radiology","volume":" ","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00247-026-06767-z","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"PEDIATRICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Endotracheal tubes (ETTs) are critical life-support devices for mechanically ventilated pediatric patients, yet automated ETT assessment on pediatric chest radiographs (CXRs) remains limited.
Objective: To develop and evaluate a two-stage deep learning pipeline for automated detection and localization of ETTs on pediatric CXRs.
Materials and methods: This retrospective study included 1,000 pediatric CXRs (476 ETT-positive, 524 ETT-negative) acquired in 2021 at a single institution. ETT segmentation masks and distal tip coordinates were annotated by trained analysts and verified by pediatric radiologists. A two-stage pipeline consisting of a ResNet classification model followed by a U-Net segmentation model was developed for ETT detection and localization. Performance was evaluated on a held-out test set using multiple metrics, including the area under the receiver operating characteristic curve (AUROC) and the mean absolute error (MAE), with 95% confidence intervals (CI). Inter-observer variability was assessed as a reference for localization performance.
Results: Inter-observer variability for ETT tip localization was 2.01 mm MAE on the held-out test set. The pipeline achieved an AUROC of 0.994 (95% CI 0.986, 1.000) for ETT detection. For localization, the pipeline achieved a MAE of 6.59 mm (95% CI 5.19, 8.21 mm). Incorporating the classification stage substantially reduced false-positive segmentations from 17 to 3 among ETT-negative CXRs compared with the standalone segmentation model.
Conclusion: A two-stage deep learning pipeline demonstrated high performance for automated ETT detection and promising performance for tip localization on pediatric CXRs in this single-center pilot study. Further evaluation in larger and external pediatric cohorts is needed to assess generalizability.
背景:气管内管(ETTs)是机械通气儿童患者的关键生命支持装置,但在儿童胸片(cxr)上自动评估ETTs仍然有限。目的:开发和评估一种两阶段深度学习管道,用于儿科cxr中etas的自动检测和定位。材料和方法:本回顾性研究纳入了2021年在一家机构获得的1,000例儿童cxr(476例et阳性,524例et阴性)。ETT分割掩模和远端尖端坐标由训练有素的分析师注释并由儿科放射科医生验证。提出了一种由ResNet分类模型和U-Net分割模型组成的两阶段管道,用于ETT检测和定位。使用多个指标在一个固定测试集上评估性能,包括受试者工作特征曲线下面积(AUROC)和平均绝对误差(MAE), 95%置信区间(CI)。观察者之间的可变性被评估为定位性能的参考。结果:在伸出测试集中,ETT尖端定位的观察者间变异性为2.01 mm MAE。该管道检测ETT的AUROC为0.994 (95% CI 0.986, 1.000)。对于定位,管道的MAE达到6.59 mm (95% CI 5.19, 8.21 mm)。与独立分割模型相比,纳入分类阶段大大减少了ett阴性cxr中的假阳性分割,从17个减少到3个。结论:在这项单中心试点研究中,两阶段深度学习管道在自动ETT检测和儿科cxr尖端定位方面表现优异。需要在更大的外部儿科队列中进行进一步的评估,以评估其普遍性。
期刊介绍:
Official Journal of the European Society of Pediatric Radiology, the Society for Pediatric Radiology and the Asian and Oceanic Society for Pediatric Radiology
Pediatric Radiology informs its readers of new findings and progress in all areas of pediatric imaging and in related fields. This is achieved by a blend of original papers, complemented by reviews that set out the present state of knowledge in a particular area of the specialty or summarize specific topics in which discussion has led to clear conclusions. Advances in technology, methodology, apparatus and auxiliary equipment are presented, and modifications of standard techniques are described.
Manuscripts submitted for publication must contain a statement to the effect that all human studies have been reviewed by the appropriate ethics committee and have therefore been performed in accordance with the ethical standards laid down in an appropriate version of the 1964 Declaration of Helsinki. It should also be stated clearly in the text that all persons gave their informed consent prior to their inclusion in the study. Details that might disclose the identity of the subjects under study should be omitted.