Aberrant, Elevated Papillary Muscle and LV Wall Forces Associated With Mitral Valve Prolapse Are Normalized by Surgical Neochordal Repair but Not by Transcatheter Edge-to-Edge Repair.
Matthew H Park, Antonia van Kampen, Chris Huynh, Katrina Liou, Lisa Lee, Serguei Melnitchouk, Yuanjia Zhu, Stefan Elde, Alexander K Reed, Gabe Weininger, Michael A Borger, Robert A Levine, Y Joseph Woo
{"title":"Aberrant, Elevated Papillary Muscle and LV Wall Forces Associated With Mitral Valve Prolapse Are Normalized by Surgical Neochordal Repair but Not by Transcatheter Edge-to-Edge Repair.","authors":"Matthew H Park, Antonia van Kampen, Chris Huynh, Katrina Liou, Lisa Lee, Serguei Melnitchouk, Yuanjia Zhu, Stefan Elde, Alexander K Reed, Gabe Weininger, Michael A Borger, Robert A Levine, Y Joseph Woo","doi":"10.1161/CIRCINTERVENTIONS.125.016161","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Recent studies have linked mitral valve prolapse to increased papillary muscle (PM) traction in mitral valve prolapse and MitraClip patients. We aimed to biomechanically compare edge-to-edge repair via MitraClip implantation and neochordal repair for mitral valve prolapse on PM forces ex vivo.</p><p><strong>Methods: </strong>We developed a force transduction system to measure the forces exerted on the PMs throughout the cardiac cycle in a left heart simulator. Using a randomized, paired procedure with porcine mitral valves (n=8), we measured the hemodynamic parameters and PM forces for control, prolapse, chordal rupture, MitraClip edge-to-edge repair, surgical edge-to-edge repair, and neochordal repair conditions.</p><p><strong>Results: </strong>Hemodynamic parameters, including maximum ventricular pressure and transmitral pressure gradients, remained consistent across all experimental conditions. The rupture condition induced pathological mitral regurgitation relative to control valves (23.7±18.5% versus 3.1±1.0%; <i>P</i><0.001). In the chordal rupture model, composite peak PM forces were 7.8±1.3 N, ≈16% higher than baseline control valves (6.7±0.7 N, <i>P</i>=0.005). Both surgical and transcatheter edge-to-edge repair conditions resulted in significantly higher peak PM forces compared with both baseline (<i>P</i><0.001) and diseased prolapse conditions (<i>P</i>=0.03). MitraClip repair valves recorded peak PM forces ≈23% higher than those of baseline control valves. Neochordal repair of the prolapsed model resulted in significant reductions of peak PM forces to near-baseline levels (6.8±1.3 N; <i>P</i>=0.53). All repair conditions reduced mitral regurgitation levels to moderate (<15%).</p><p><strong>Conclusions: </strong>We demonstrated that MitraClip, while reducing regurgitation, increases aberrant forces on the PMs. In contrast, neochordal repair both reduces regurgitation and restores PM forces back to physiological levels. Our work suggests that these elevated forces, associated with mitral valve prolapse and MitraClip, predispose the mitral valve to continued traction, altering the mechanical load on the subvalvular apparatus.</p>","PeriodicalId":10330,"journal":{"name":"Circulation: Cardiovascular Interventions","volume":" ","pages":"e016161"},"PeriodicalIF":7.2000,"publicationDate":"2026-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation: Cardiovascular Interventions","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCINTERVENTIONS.125.016161","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Recent studies have linked mitral valve prolapse to increased papillary muscle (PM) traction in mitral valve prolapse and MitraClip patients. We aimed to biomechanically compare edge-to-edge repair via MitraClip implantation and neochordal repair for mitral valve prolapse on PM forces ex vivo.
Methods: We developed a force transduction system to measure the forces exerted on the PMs throughout the cardiac cycle in a left heart simulator. Using a randomized, paired procedure with porcine mitral valves (n=8), we measured the hemodynamic parameters and PM forces for control, prolapse, chordal rupture, MitraClip edge-to-edge repair, surgical edge-to-edge repair, and neochordal repair conditions.
Results: Hemodynamic parameters, including maximum ventricular pressure and transmitral pressure gradients, remained consistent across all experimental conditions. The rupture condition induced pathological mitral regurgitation relative to control valves (23.7±18.5% versus 3.1±1.0%; P<0.001). In the chordal rupture model, composite peak PM forces were 7.8±1.3 N, ≈16% higher than baseline control valves (6.7±0.7 N, P=0.005). Both surgical and transcatheter edge-to-edge repair conditions resulted in significantly higher peak PM forces compared with both baseline (P<0.001) and diseased prolapse conditions (P=0.03). MitraClip repair valves recorded peak PM forces ≈23% higher than those of baseline control valves. Neochordal repair of the prolapsed model resulted in significant reductions of peak PM forces to near-baseline levels (6.8±1.3 N; P=0.53). All repair conditions reduced mitral regurgitation levels to moderate (<15%).
Conclusions: We demonstrated that MitraClip, while reducing regurgitation, increases aberrant forces on the PMs. In contrast, neochordal repair both reduces regurgitation and restores PM forces back to physiological levels. Our work suggests that these elevated forces, associated with mitral valve prolapse and MitraClip, predispose the mitral valve to continued traction, altering the mechanical load on the subvalvular apparatus.
背景:最近的研究将二尖瓣脱垂与二尖瓣脱垂和MitraClip患者乳头肌(PM)牵引力增加联系起来。我们的目的是在生物力学上比较二尖瓣脱垂在PM力作用下通过MitraClip植入和新索索修复的边缘到边缘修复。方法:我们开发了一种力传导系统来测量在左心模拟器中施加在pm上的力。采用猪二尖瓣随机配对手术(n=8),我们测量了血流动力学参数和PM力,用于控制、脱垂、脊索破裂、MitraClip边缘对边缘修复、手术边缘对边缘修复和新脊索修复。结果:血流动力学参数,包括最大心室压和透射压梯度,在所有实验条件下保持一致。相对于控制阀,破裂情况引起病理性二尖瓣反流(23.7±18.5% vs 3.1±1.0%;PP=0.005)。与基线相比,手术和经导管边缘对边缘修复条件导致峰值PM力显著更高(PP=0.03)。MitraClip修复阀记录的峰值PM力比基线控制阀高约23%。脱垂模型的新脊索修复导致峰值PM力显著降低至接近基线水平(6.8±1.3 N; P=0.53)。所有的修复条件都将二尖瓣反流水平降低到中等水平(结论:我们证明了MitraClip在减少反流的同时,增加了对pm的异常力。相反,新脊索修复既可以减少反流,又可以将PM力恢复到生理水平。我们的工作表明,这些升高的力,与二尖瓣脱垂和MitraClip相关,使二尖瓣易于持续牵引,改变了瓣下装置的机械负荷。
期刊介绍:
Circulation: Cardiovascular Interventions, an American Heart Association journal, focuses on interventional techniques pertaining to coronary artery disease, structural heart disease, and vascular disease, with priority placed on original research and on randomized trials and large registry studies. In addition, pharmacological, diagnostic, and pathophysiological aspects of interventional cardiology are given special attention in this online-only journal.