Influences of design aspects on recording functionality of cuff-type peripheral nerve interfaces.

IF 3.8
Vera A Oppelt, Emma Mueller, Fabian Pfeiffer, Seline Mahler, Ronny Pfeifer, Martin Schuettler, Thomas Stieglitz
{"title":"Influences of design aspects on recording functionality of cuff-type peripheral nerve interfaces.","authors":"Vera A Oppelt, Emma Mueller, Fabian Pfeiffer, Seline Mahler, Ronny Pfeifer, Martin Schuettler, Thomas Stieglitz","doi":"10.1088/1741-2552/aea36d","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective.</i>High-quality recording of peripheral nerve signals is a critical requirement for implantable neural interfaces, yet the influence of contact configuration and electrical insulation on cuff electrode recording quality remains poorly understood.<i>Approach.</i>Using a physical nerve model<i>in vitro</i>, the transfer function (TF) of commercially available split-cylinder and spiral cuff electrodes was measured for systematically varied contact configurations, including contact size, spacing, and size ratio of cathode to anode, as well as degree of electrical insulation. To derive single-fiber and compound action potentials, the TF was convolved with modeled action currents, enabling direct comparison of recording quality across designs.<i>Main results.</i>Contact edge spacing was found to be more decisive than contact surface area for recording amplitude, regardless of contact segmentation. In quasi-tripolar configurations, common mode suppression additionally reduced sensitivity to geometric variations. Both the length of the longitudinal insulation and the integrity of the electrode seal were found to strongly influence recording amplitude, with complete removal of the seal reducing amplitude by approximately 50%, outweighing all geometric design parameters combined.<i>Significance.</i>These findings establish electrical insulation integrity as the primary design criterion for cuff electrode recording performance, shifting focus away from contact geometry alone, and provide a practical framework for the development of next-generation peripheral nerve interfaces.</p>","PeriodicalId":94096,"journal":{"name":"Journal of neural engineering","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2026-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neural engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/1741-2552/aea36d","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective.High-quality recording of peripheral nerve signals is a critical requirement for implantable neural interfaces, yet the influence of contact configuration and electrical insulation on cuff electrode recording quality remains poorly understood.Approach.Using a physical nerve modelin vitro, the transfer function (TF) of commercially available split-cylinder and spiral cuff electrodes was measured for systematically varied contact configurations, including contact size, spacing, and size ratio of cathode to anode, as well as degree of electrical insulation. To derive single-fiber and compound action potentials, the TF was convolved with modeled action currents, enabling direct comparison of recording quality across designs.Main results.Contact edge spacing was found to be more decisive than contact surface area for recording amplitude, regardless of contact segmentation. In quasi-tripolar configurations, common mode suppression additionally reduced sensitivity to geometric variations. Both the length of the longitudinal insulation and the integrity of the electrode seal were found to strongly influence recording amplitude, with complete removal of the seal reducing amplitude by approximately 50%, outweighing all geometric design parameters combined.Significance.These findings establish electrical insulation integrity as the primary design criterion for cuff electrode recording performance, shifting focus away from contact geometry alone, and provide a practical framework for the development of next-generation peripheral nerve interfaces.

设计方面对袖带型周围神经接口记录功能的影响。
目的:高质量的周围神经信号记录是植入式神经接口的关键要求,但接触配置和电绝缘对袖带电极记录质量的影响尚不清楚。方法:使用体外物理神经模型,测量了市面上可买到的分离式圆柱体和螺旋袖带电极的传递函数,以系统地改变接触配置,包括接触尺寸、间距、阴极与阳极的尺寸比以及电绝缘程度。为了得到单纤维和复合动作电位,传递函数与模拟的动作电流进行卷积,从而可以直接比较不同设计的记录质量。主要结果:发现接触边缘间距比接触表面积对记录振幅更有决定性,而不考虑接触分割。在准三极性配置中,共模抑制还降低了对几何变化的灵敏度。研究发现,纵向绝缘的长度和电极密封的完整性都对记录振幅有很大影响,完全去除密封会使振幅降低约50%,超过所有几何设计参数的总和。意义:这些发现建立了电绝缘完整性作为袖带电极记录性能的主要设计标准,将焦点从单独的接触几何转移,并为下一代周围神经接口的开发提供了实用框架。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书