Designing an Anatomical Plate to Improve Biomechanics and Stability for Distal Humerus Fractures: FEM Investigation and the Influence of Meshing Strategy.

IF 1.6 Q3 ORTHOPEDICS
Fahime Rezazade, Azadeh Ghouchani
{"title":"Designing an Anatomical Plate to Improve Biomechanics and Stability for Distal Humerus Fractures: FEM Investigation and the Influence of Meshing Strategy.","authors":"Fahime Rezazade, Azadeh Ghouchani","doi":"10.22038/abjs.2026.95901.4303","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Distal humerus fractures require stable fixation to restore elbow function, yet conventional plates often exhibit a geometric mismatch with bone morphology, potentially compromising biomechanical performance. Our finite element analysis (FEA) compared a novel anatomical plate designed to conform to native humeral geometry with a conventional plate for treating type B1 distal humerus fractures. The influence of meshing strategy on the outcomes was also evaluated. Both adaptive and uniform approaches were considered.</p><p><strong>Methods: </strong>A CT-based three-dimensional humerus model with heterogeneous bone properties was reconstructed from scans of a 37-year-old woman. An anatomically contoured plate was compared with a conventional locking plate. Adaptive and uniform meshing were applied separately to each model, and each construct was subsequently tested under 100 N tensile and compressive loads.</p><p><strong>Results: </strong>The anatomical plate showed higher stiffness (up to 13%), a broader contact pressure distribution (23--28 MPa vs. conventional: 12--18 MPa), and a larger area of sticking contact, whereas the conventional plate produced higher shear stresses (p=0.113). Uniform meshing generated significantly greater interfragmentary motion than adaptive meshing (p=0.008) but caused unrealistically high screw stresses (up to 1526 MPa). Adaptive meshing predicted higher bone stresses, whereas uniform meshing concentrated stresses in the screws. Although meshing affected the absolute stress and strain magnitudes, the comparative differences between plate designs remained consistent.</p><p><strong>Conclusion: </strong>The anatomically contoured plate offers greater biomechanical stability than conventional designs for distal humerus fractures, as evidenced by reduced displacement and maintained fragment positioning conducive to healing. An adaptive meshing strategy appears more suitable for FEA of complex fixation components because it yields more reliable stress distributions without altering the primary comparative conclusions. Further experimental validation is warranted before clinical application.</p>","PeriodicalId":46704,"journal":{"name":"Archives of Bone and Joint Surgery-ABJS","volume":"14 8","pages":"540-549"},"PeriodicalIF":1.6000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13544424/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of Bone and Joint Surgery-ABJS","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/abjs.2026.95901.4303","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: Distal humerus fractures require stable fixation to restore elbow function, yet conventional plates often exhibit a geometric mismatch with bone morphology, potentially compromising biomechanical performance. Our finite element analysis (FEA) compared a novel anatomical plate designed to conform to native humeral geometry with a conventional plate for treating type B1 distal humerus fractures. The influence of meshing strategy on the outcomes was also evaluated. Both adaptive and uniform approaches were considered.

Methods: A CT-based three-dimensional humerus model with heterogeneous bone properties was reconstructed from scans of a 37-year-old woman. An anatomically contoured plate was compared with a conventional locking plate. Adaptive and uniform meshing were applied separately to each model, and each construct was subsequently tested under 100 N tensile and compressive loads.

Results: The anatomical plate showed higher stiffness (up to 13%), a broader contact pressure distribution (23--28 MPa vs. conventional: 12--18 MPa), and a larger area of sticking contact, whereas the conventional plate produced higher shear stresses (p=0.113). Uniform meshing generated significantly greater interfragmentary motion than adaptive meshing (p=0.008) but caused unrealistically high screw stresses (up to 1526 MPa). Adaptive meshing predicted higher bone stresses, whereas uniform meshing concentrated stresses in the screws. Although meshing affected the absolute stress and strain magnitudes, the comparative differences between plate designs remained consistent.

Conclusion: The anatomically contoured plate offers greater biomechanical stability than conventional designs for distal humerus fractures, as evidenced by reduced displacement and maintained fragment positioning conducive to healing. An adaptive meshing strategy appears more suitable for FEA of complex fixation components because it yields more reliable stress distributions without altering the primary comparative conclusions. Further experimental validation is warranted before clinical application.

设计一种改善肱骨远端骨折生物力学和稳定性的解剖钢板:FEM研究及啮合策略的影响。
目的:肱骨远端骨折需要稳定的固定来恢复肘关节功能,然而传统的钢板通常表现出与骨形态的几何不匹配,潜在地影响生物力学性能。我们的有限元分析(FEA)比较了一种符合天然肱骨几何形状的新型解剖钢板与治疗B1型肱骨远端骨折的传统钢板。并对网格划分策略对结果的影响进行了评价。考虑了自适应方法和统一方法。方法:对一名37岁女性的肱骨进行ct重建,建立了具有异质骨特性的三维模型。解剖轮廓钢板与常规锁定钢板进行比较。每个模型分别采用自适应和均匀网格划分,每个结构随后在100 N拉伸和压缩载荷下进行测试。结果:解剖板显示更高的刚度(高达13%),更广泛的接触压力分布(23- 28 MPa,而传统板:12- 18 MPa),更大的粘接面积,而传统板产生更高的剪切应力(p=0.113)。均匀网格比自适应网格产生更大的碎片间运动(p=0.008),但产生不切实际的高螺钉应力(高达1526 MPa)。自适应网格预测较高的骨应力,而均匀网格集中应力在螺钉。虽然啮合影响绝对应力和应变大小,但不同板设计之间的比较差异保持一致。结论:与传统设计相比,解剖型钢板在肱骨远端骨折治疗中具有更高的生物力学稳定性,可减少移位并保持碎片定位,有利于愈合。自适应网格策略更适合于复杂固定部件的有限元分析,因为它在不改变主要比较结论的情况下产生更可靠的应力分布。在临床应用前需要进一步的实验验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.30
自引率
0.00%
发文量
128
期刊介绍: The Archives of Bone and Joint Surgery (ABJS) aims to encourage a better understanding of all aspects of Orthopedic Sciences. The journal accepts scientific papers including original research, review article, short communication, case report, and letter to the editor in all fields of bone, joint, musculoskeletal surgery and related researches. The Archives of Bone and Joint Surgery (ABJS) will publish papers in all aspects of today`s modern orthopedic sciences including: Arthroscopy, Arthroplasty, Sport Medicine, Reconstruction, Hand and Upper Extremity, Pediatric Orthopedics, Spine, Trauma, Foot and Ankle, Tumor, Joint Rheumatic Disease, Skeletal Imaging, Orthopedic Physical Therapy, Rehabilitation, Orthopedic Basic Sciences (Biomechanics, Biotechnology, Biomaterial..).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书