Spurious Fingertip Oxygen Desaturation During the 6-Minute Walk Test in a Patient With NSIP and Raynaud Phenomenon: The Value of Forehead Pulse Oximetry and Transcutaneous Oxygen Tension.

IF 0.7 Q4 RESPIRATORY SYSTEM
Respirology Case Reports Pub Date : 2026-08-25 eCollection Date: 2026-08-01 DOI:10.1002/rcr2.70737
Yusuke Nakamura, Masayoshi Takamori, Honoka Sunaga, Kosuke Natsume, Yuri Nomura, Yuto Goto, Nobuhiko Tsukada, Azusa Tsukada, Nobuhiko Uchida, Nana Yazawa, Hiromi Hoshi, Yasuo Shimizu, Seiji Niho
{"title":"Spurious Fingertip Oxygen Desaturation During the 6-Minute Walk Test in a Patient With NSIP and Raynaud Phenomenon: The Value of Forehead Pulse Oximetry and Transcutaneous Oxygen Tension.","authors":"Yusuke Nakamura, Masayoshi Takamori, Honoka Sunaga, Kosuke Natsume, Yuri Nomura, Yuto Goto, Nobuhiko Tsukada, Azusa Tsukada, Nobuhiko Uchida, Nana Yazawa, Hiromi Hoshi, Yasuo Shimizu, Seiji Niho","doi":"10.1002/rcr2.70737","DOIUrl":null,"url":null,"abstract":"<p><p>Reduced digital perfusion associated with Raynaud phenomenon may cause spuriously low fingertip pulse oximetry readings during the 6-minute walk test (6MWT), potentially leading to overestimation of exercise-induced hypoxaemia in patients with interstitial lung disease. We report a 39-year-old woman with non-specific interstitial pneumonia and Raynaud phenomenon who underwent repeated 6MWTs. Fingertip SpO<sub>2</sub> progressively decreased from 96% to 82%, whereas forehead SpO<sub>2</sub> remained stable (93%-98%). Simultaneously, transcutaneous oxygen tension (TcPO<sub>2</sub>) measured on the volar forearm remained unchanged. Progressive finger cooling and digital pallor were also observed. Fingertip SpO<sub>2</sub> rapidly returned to baseline during recovery. These findings suggested that the apparent desaturation reflected reduced digital perfusion rather than true systemic hypoxaemia. This case highlights the potential limitations of fingertip pulse oximetry during exercise in patients with Raynaud phenomenon. Alternative methods of assessing oxygenation, such as forehead pulse oximetry or TcPO<sub>2</sub>, may be useful during the 6MWT when fingertip SpO<sub>2</sub> readings are discordant with the clinical presentation.</p>","PeriodicalId":45846,"journal":{"name":"Respirology Case Reports","volume":"14 8","pages":"e70737"},"PeriodicalIF":0.7000,"publicationDate":"2026-08-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13504363/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Respirology Case Reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/rcr2.70737","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0

Abstract

Reduced digital perfusion associated with Raynaud phenomenon may cause spuriously low fingertip pulse oximetry readings during the 6-minute walk test (6MWT), potentially leading to overestimation of exercise-induced hypoxaemia in patients with interstitial lung disease. We report a 39-year-old woman with non-specific interstitial pneumonia and Raynaud phenomenon who underwent repeated 6MWTs. Fingertip SpO2 progressively decreased from 96% to 82%, whereas forehead SpO2 remained stable (93%-98%). Simultaneously, transcutaneous oxygen tension (TcPO2) measured on the volar forearm remained unchanged. Progressive finger cooling and digital pallor were also observed. Fingertip SpO2 rapidly returned to baseline during recovery. These findings suggested that the apparent desaturation reflected reduced digital perfusion rather than true systemic hypoxaemia. This case highlights the potential limitations of fingertip pulse oximetry during exercise in patients with Raynaud phenomenon. Alternative methods of assessing oxygenation, such as forehead pulse oximetry or TcPO2, may be useful during the 6MWT when fingertip SpO2 readings are discordant with the clinical presentation.

NSIP患者6分钟步行试验中假指尖氧饱和度和雷诺现象:前额脉搏血氧仪和经皮氧张力的价值。
在6分钟步行试验(6MWT)中,与雷诺现象相关的手指灌注减少可能导致指尖脉搏血氧仪读数虚低,可能导致对间质性肺疾病患者运动性低氧血症的高估。我们报告一位患有非特异性间质性肺炎和雷诺现象的39岁女性,她接受了多次6MWTs。指尖SpO2从96%逐渐下降到82%,而前额SpO2保持稳定(93%-98%)。同时,经皮氧张力(TcPO2)测量前臂掌侧保持不变。还观察到渐进式手指冷却和手指苍白。指尖SpO2在恢复过程中迅速恢复到基线水平。这些结果表明,明显的去饱和反映了数字灌注减少,而不是真正的全身低氧血症。本病例强调了雷诺现象患者运动时指尖脉搏血氧测定的潜在局限性。在6MWT期间,当指尖SpO2读数与临床表现不一致时,评估氧合的替代方法,如前额脉搏血氧仪或TcPO2,可能是有用的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Respirology Case Reports
Respirology Case Reports RESPIRATORY SYSTEM-
CiteScore
1.40
自引率
0.00%
发文量
178
审稿时长
8 weeks
期刊介绍: Respirology Case Reports is an open-access online journal dedicated to the publication of original clinical case reports, case series, clinical images and clinical videos in all fields of respiratory medicine. The Journal encourages the international exchange between clinicians and researchers of experiences in diagnosing and treating uncommon diseases or diseases with unusual presentations. All manuscripts are peer-reviewed through a streamlined process that aims at providing a rapid turnaround time from submission to publication.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书