Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser Enucleation of the Prostate: A Prostate Enucleation Collaborative Initiative.
Jessica Wu, Amanda E Kahn, Benjamin Dalkin, Satvik Kuravi, Adam Cohen, Charles Klose, Eric Qualkenbush, Bryce Baird, Karen L Stern, Smita De, Dane Klett, Rajat Jain, Scott O Quarrier, Chandler D Dora
{"title":"Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser Enucleation of the Prostate: A Prostate Enucleation Collaborative Initiative.","authors":"Jessica Wu, Amanda E Kahn, Benjamin Dalkin, Satvik Kuravi, Adam Cohen, Charles Klose, Eric Qualkenbush, Bryce Baird, Karen L Stern, Smita De, Dane Klett, Rajat Jain, Scott O Quarrier, Chandler D Dora","doi":"10.1002/lsm.70197","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Operating room (OR) noise during holmium laser enucleation of the prostate (HoLEP), particularly from laser activation, may impair surgical workflow and staff performance.</p><p><strong>Methods: </strong>84 HoLEP procedures were recorded at three academic institutions between June 2024 and January 2026. Laser platforms included Lumenis Moses 2.0, Quanta Litho 150, Quanta Magneto, and Dornier Thulio. Sound measurements were obtained using a standardized smartphone-based system during three phases: laser-off (baseline), idle (not firing), and active lasing. Acoustic metrics included average sound level (LAeq), peak sound level (Lcpeak), and sound levels exceeded for 10%, 50%, and 90% of the recording time (L10, L50, and L90). Background noise correction was performed using baseline measurements to account for ambient noise and isolate laser-related sound. OR staff completed a validated survey assessing perceived noise burden (0-10 scale).</p><p><strong>Results: </strong>Background-corrected noise during active lasing differed significantly by platform. Litho demonstrated lower mean noise levels compared to Thulio for LAeq (60.6 vs. 68.2 dB, p = 0.023) and L10 (61.5 vs. 70.6 dB, p = 0.023), while Moses had higher LAeq compared to Litho (63.1 vs. 60.6 dB, p = 0.043). Overall, Litho had the lowest noise levels, whereas Thulio had the highest. Survey responses paralleled these findings. Litho was associated with lower perceived overall effort (2.1 ± 0.9), while Moses demonstrated higher overall effort (3.6 ± 1.2) for communication and concentration (Figure 2). Significant pairwise differences were observed between Moses and Litho (all p < 0.05). Trends were consistent across sites.</p><p><strong>Conclusions: </strong>Noise levels during HoLEP varied by laser technology and correlated with differences in perceived communication and cognitive effort. Litho demonstrated the lowest acoustic burden, while Moses and Thulio had higher noise exposure. These findings highlight the impact of laser selection on the OR environment and intraoperative team performance.</p>","PeriodicalId":17961,"journal":{"name":"Lasers in Surgery and Medicine","volume":" ","pages":"877-884"},"PeriodicalIF":2.1000,"publicationDate":"2026-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Lasers in Surgery and Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/lsm.70197","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/24 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"DERMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Operating room (OR) noise during holmium laser enucleation of the prostate (HoLEP), particularly from laser activation, may impair surgical workflow and staff performance.
Methods: 84 HoLEP procedures were recorded at three academic institutions between June 2024 and January 2026. Laser platforms included Lumenis Moses 2.0, Quanta Litho 150, Quanta Magneto, and Dornier Thulio. Sound measurements were obtained using a standardized smartphone-based system during three phases: laser-off (baseline), idle (not firing), and active lasing. Acoustic metrics included average sound level (LAeq), peak sound level (Lcpeak), and sound levels exceeded for 10%, 50%, and 90% of the recording time (L10, L50, and L90). Background noise correction was performed using baseline measurements to account for ambient noise and isolate laser-related sound. OR staff completed a validated survey assessing perceived noise burden (0-10 scale).
Results: Background-corrected noise during active lasing differed significantly by platform. Litho demonstrated lower mean noise levels compared to Thulio for LAeq (60.6 vs. 68.2 dB, p = 0.023) and L10 (61.5 vs. 70.6 dB, p = 0.023), while Moses had higher LAeq compared to Litho (63.1 vs. 60.6 dB, p = 0.043). Overall, Litho had the lowest noise levels, whereas Thulio had the highest. Survey responses paralleled these findings. Litho was associated with lower perceived overall effort (2.1 ± 0.9), while Moses demonstrated higher overall effort (3.6 ± 1.2) for communication and concentration (Figure 2). Significant pairwise differences were observed between Moses and Litho (all p < 0.05). Trends were consistent across sites.
Conclusions: Noise levels during HoLEP varied by laser technology and correlated with differences in perceived communication and cognitive effort. Litho demonstrated the lowest acoustic burden, while Moses and Thulio had higher noise exposure. These findings highlight the impact of laser selection on the OR environment and intraoperative team performance.
期刊介绍:
Lasers in Surgery and Medicine publishes the highest quality research and clinical manuscripts in areas relating to the use of lasers in medicine and biology. The journal publishes basic and clinical studies on the therapeutic and diagnostic use of lasers in all the surgical and medical specialties. Contributions regarding clinical trials, new therapeutic techniques or instrumentation, laser biophysics and bioengineering, photobiology and photochemistry, outcomes research, cost-effectiveness, and other aspects of biomedicine are welcome. Using a process of rigorous yet rapid review of submitted manuscripts, findings of high scientific and medical interest are published with a minimum delay.