Analytical interference by IgM paraprotein causing platform-dependent electrolyte discrepancies: a case-based laboratory investigation.

IF 1.8
Biochemia medica Pub Date : 2026-10-15 Epub Date: 2026-08-10 DOI:10.11613/BM.2026.030901
Sheng Wang, Xiaoyong Duan, Yulin Zou, Jianxiong Gui
{"title":"Analytical interference by IgM paraprotein causing platform-dependent electrolyte discrepancies: a case-based laboratory investigation.","authors":"Sheng Wang, Xiaoyong Duan, Yulin Zou, Jianxiong Gui","doi":"10.11613/BM.2026.030901","DOIUrl":null,"url":null,"abstract":"<p><p>Accurate measurement of serum electrolytes and calcium is essential for clinical management; however, monoclonal immunoglobulins, particularly IgM paraproteins, may cause misleading analytical interference. We report the case of an 85-year-old man admitted for low back pain whose serum sample showed marked inter-platform discordance in sodium and total calcium results. The Hitachi platform reported a \"hypercalcemic crisis\" (total calcium 4.95 mmol/L) and severe hyponatremia (125 mmol/L), whereas the Beckman Coulter AU platform showed only mildly elevated total calcium (2.89 mmol/L) and milder hyponatremia (133 mmol/L). This corresponded to a 71.3% difference in total calcium, despite both platforms using the Arsenazo III method. Direct ion-selective electrode measurements excluded true hyponatremia and hypercalcemia. Dilution studies, polyethylene glycol precipitation, reaction-curve inspection, and simulation experiments confirmed that IgM-related precipitation and optical scattering were the underlying mechanisms. Specifically, platform-specific mixing dynamics determined the magnitude of interference. Recognition of biochemical-clinical discordance and targeted method comparison prompted hematological evaluation and led to a diagnosis of Waldenström macroglobulinemia. This case highlights that unexplained analytical discrepancies may delay recognition of underlying disease if analytical interference is not appropriately investigated.</p>","PeriodicalId":94370,"journal":{"name":"Biochemia medica","volume":"36 3","pages":"030901"},"PeriodicalIF":1.8000,"publicationDate":"2026-10-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13489491/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemia medica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11613/BM.2026.030901","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/10 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Accurate measurement of serum electrolytes and calcium is essential for clinical management; however, monoclonal immunoglobulins, particularly IgM paraproteins, may cause misleading analytical interference. We report the case of an 85-year-old man admitted for low back pain whose serum sample showed marked inter-platform discordance in sodium and total calcium results. The Hitachi platform reported a "hypercalcemic crisis" (total calcium 4.95 mmol/L) and severe hyponatremia (125 mmol/L), whereas the Beckman Coulter AU platform showed only mildly elevated total calcium (2.89 mmol/L) and milder hyponatremia (133 mmol/L). This corresponded to a 71.3% difference in total calcium, despite both platforms using the Arsenazo III method. Direct ion-selective electrode measurements excluded true hyponatremia and hypercalcemia. Dilution studies, polyethylene glycol precipitation, reaction-curve inspection, and simulation experiments confirmed that IgM-related precipitation and optical scattering were the underlying mechanisms. Specifically, platform-specific mixing dynamics determined the magnitude of interference. Recognition of biochemical-clinical discordance and targeted method comparison prompted hematological evaluation and led to a diagnosis of Waldenström macroglobulinemia. This case highlights that unexplained analytical discrepancies may delay recognition of underlying disease if analytical interference is not appropriately investigated.

IgM副蛋白引起的平台依赖性电解质差异的分析干扰:一项基于病例的实验室调查。
准确测定血清电解质和钙对临床管理至关重要;然而,单克隆免疫球蛋白,特别是IgM副蛋白,可能引起误导性的分析干扰。我们报告一例85岁男性因腰痛入院,其血清样本显示明显的平台间钠和总钙结果不一致。日立平台报告了“高钙危机”(总钙4.95 mmol/L)和严重低钠血症(125 mmol/L),而Beckman Coulter AU平台仅显示轻度总钙升高(2.89 mmol/L)和轻度低钠血症(133 mmol/L)。尽管两种平台都使用了偶氮胂III方法,但总钙含量相差71.3%。直接离子选择电极测量排除了真正的低钠血症和高钙血症。稀释研究、聚乙二醇沉淀、反应曲线检验和模拟实验证实,igm相关的沉淀和光散射是潜在的机制。具体来说,平台特定的混合动力学决定了干扰的大小。认识到生化-临床不一致和有针对性的方法比较,促使血液学评估并导致Waldenström巨球蛋白血症的诊断。本病例强调,如果分析干扰没有得到适当的调查,无法解释的分析差异可能会延迟对潜在疾病的识别。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书