Increased False Positives vs Increased Interference Flags: Comparison of 2 Ethyl Glucuronide Immunoassays.

IF 2.3 Q3 MEDICAL LABORATORY TECHNOLOGY
Alexandria N Jackson, Sara P Wyness, Sonia L La'ulu, Jessica M Boyd, Kamisha L Johnson-Davis
{"title":"Increased False Positives vs Increased Interference Flags: Comparison of 2 Ethyl Glucuronide Immunoassays.","authors":"Alexandria N Jackson, Sara P Wyness, Sonia L La'ulu, Jessica M Boyd, Kamisha L Johnson-Davis","doi":"10.1093/jalm/jfag114","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Screening for alcohol use in a population to treat pain and substance use disorders is important for monitoring adherence to treatment plans. Urine screening for ethanol exposure by immunoassay can provide a cost-effective testing option, with a faster turnaround time to result in comparison to LC-MS/MS. The purpose of this study was to compare two ethyl glucuronide (EtG) immunoassays to determine which method would be better suited for our laboratory work flow and patient population.</p><p><strong>Methods: </strong>Two EtG immunoassays were evaluated to assess false-positive results and interference flags for alcohol biomarker testing. The EtG enzyme immunoassays were verified and put into use for patient testing during 2 separate timeframes.</p><p><strong>Results: </strong>Both assays compared qualitatively well to LC-MS/MS for positive EtG (>500 ng/mL) patient samples. In our patient population, the ARK EtG assay had a presumptive positivity rate of 13.81%, a false-positive rate of 1.35%, and an interference rate of 0.45%. The Immunalysis assay had a 9.87% presumptive positivity rate, a false-positive rate of 0.15%, and a 2.86% interference rate.</p><p><strong>Conclusions: </strong>Selecting an assay for the laboratory requires understanding the strengths and limitations of the method. For our work flow, our lab chose the assay with more interferences, since our work flow reflexes presumptive positive immunoassay results and interferences to confirmation by mass spectrometry. For labs that use an immunoassay screen-only approach, selecting an immunoassay that generates a higher rate of interference flags could make result interpretation difficult for the client and raise concerns for specimen adulteration.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2026-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/jalm/jfag114","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Screening for alcohol use in a population to treat pain and substance use disorders is important for monitoring adherence to treatment plans. Urine screening for ethanol exposure by immunoassay can provide a cost-effective testing option, with a faster turnaround time to result in comparison to LC-MS/MS. The purpose of this study was to compare two ethyl glucuronide (EtG) immunoassays to determine which method would be better suited for our laboratory work flow and patient population.

Methods: Two EtG immunoassays were evaluated to assess false-positive results and interference flags for alcohol biomarker testing. The EtG enzyme immunoassays were verified and put into use for patient testing during 2 separate timeframes.

Results: Both assays compared qualitatively well to LC-MS/MS for positive EtG (>500 ng/mL) patient samples. In our patient population, the ARK EtG assay had a presumptive positivity rate of 13.81%, a false-positive rate of 1.35%, and an interference rate of 0.45%. The Immunalysis assay had a 9.87% presumptive positivity rate, a false-positive rate of 0.15%, and a 2.86% interference rate.

Conclusions: Selecting an assay for the laboratory requires understanding the strengths and limitations of the method. For our work flow, our lab chose the assay with more interferences, since our work flow reflexes presumptive positive immunoassay results and interferences to confirmation by mass spectrometry. For labs that use an immunoassay screen-only approach, selecting an immunoassay that generates a higher rate of interference flags could make result interpretation difficult for the client and raise concerns for specimen adulteration.

假阳性增加与干扰标志增加:2种乙基葡萄糖醛酸苷免疫测定的比较
背景:在治疗疼痛和物质使用障碍的人群中进行酒精使用筛查对于监测治疗计划的依从性很重要。与LC-MS/MS相比,通过免疫分析法进行尿液乙醇暴露筛查可以提供一种具有成本效益的测试选择,其周转时间更快。本研究的目的是比较两种乙基葡萄糖醛酸苷(EtG)免疫测定法,以确定哪种方法更适合我们的实验室工作流程和患者群体。方法:对两种EtG免疫测定法进行评估,以评估酒精生物标志物检测的假阳性结果和干扰标志。EtG酶免疫测定法在两个不同的时间段内进行验证并投入使用。结果:两种检测方法均与LC-MS/MS在阳性EtG (>500 ng/mL)患者样品中的定性比较良好。在我们的患者群体中,ARK EtG检测的推定阳性率为13.81%,假阳性率为1.35%,干扰率为0.45%。免疫分析方法的推定阳性率为9.87%,假阳性率为0.15%,干扰率为2.86%。结论:为实验室选择一种检测方法需要了解该方法的优点和局限性。对于我们的工作流程,我们实验室选择了干扰较多的检测方法,因为我们的工作流程将推定阳性的免疫检测结果和干扰反射到质谱确认中。对于仅使用免疫分析筛选方法的实验室,选择产生较高干扰标记率的免疫分析可能会使客户难以解释结果,并引起对标本掺假的担忧。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Applied Laboratory Medicine
Journal of Applied Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
3.70
自引率
5.00%
发文量
137
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书