Enhancing inter-fraction set-up accuracy and margin minimization (E-SAMM): a comparison of maskless surface-guided and mask-based laser alignment positioning in pelvic radiotherapy.

IF 2.8 3区 医学 Q3 ONCOLOGY
Pramod Kumar Gupta, Sharad Singh, Rumita Singh, Sumanta Manna, Ragul T, Ritika Ranjan, Himanshu Verma, Shweta Srivastava, M L B Bhatt
{"title":"Enhancing inter-fraction set-up accuracy and margin minimization (E-SAMM): a comparison of maskless surface-guided and mask-based laser alignment positioning in pelvic radiotherapy.","authors":"Pramod Kumar Gupta, Sharad Singh, Rumita Singh, Sumanta Manna, Ragul T, Ritika Ranjan, Himanshu Verma, Shweta Srivastava, M L B Bhatt","doi":"10.1007/s00066-026-02585-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Surface-guided radiation therapy (SGRT) is a non-invasive, tattoo-less technique that enables real-time optical surface monitoring during radiotherapy. This prospective E‑SAMM study compared inter-fraction set-up accuracy, estimated PTV margin requirements derived from CBCT-based pre-correction setup deviations, and set-up time between maskless SGRT and mask-based setup in patients receiving pelvic radiotherapy.</p><p><strong>Methods: </strong>One hundred patients with pelvic malignancies (cervical, endometrial, rectal, and prostate cancers) undergoing radical or neoadjuvant radiotherapy (45-68 Gy delivered in 25-28 fractions, five fractions per week) were prospectively enrolled. A total of 2376 fractions were analysed. Patients were randomised to either maskless SGRT with AlignRT, knee support, and intrafraction monitoring (Group A, n = 50) or six-clamp mask-based laser alignment with image-guided radiotherapy (Group B, n = 50). All patients underwent online cone-beam computed tomography (CBCT) verification on a TrueBeam SVC linear accelerator with a six-degree-of-freedom couch. Inter-fraction translational (medio-lateral [ML], cranio-caudal [CC], antero-posterior [AP]) and rotational set-up errors were derived from CBCT corrections. PTV margin estimates were calculated using the van Herk formula from CBCT-recorded pre-correction setup deviations. Statistical comparisons were performed using analysis of variance and the Mann-Whitney U test.</p><p><strong>Results: </strong>SGRT demonstrated significantly improved translational accuracy in the ML (1.97 ± 1.54 mm vs 2.63 ± 2.06 mm, P = 0.03) and CC (3.13 ± 2.34 mm vs 4.03 ± 2.67 mm, P = 0.001) directions compared with mask-based alignment, whereas larger shifts were observed in the AP direction (P = 0.001). Rotational differences were not statistically significant (P = 0.14). The mean three-dimensional vector displacement was significantly lower with SGRT (1.91 ± 0.36 mm) than with mask-based positioning (2.26 ± 0.48 mm; P < 0.05). Estimated PTV margins were smaller with SGRT in all translational directions. Mean set-up time was significantly reduced with SGRT (2.51 ± 0.81 min vs 3.55 ± 0.99 min, P = 0.007).</p><p><strong>Conclusions: </strong>Maskless SGRT provided improved inter-fraction set-up accuracy, lower estimated PTV margin requirements based on setup deviations, and shorter set-up times compared with mask-based laser alignment for pelvic radiotherapy.</p>","PeriodicalId":21998,"journal":{"name":"Strahlentherapie und Onkologie","volume":" ","pages":""},"PeriodicalIF":2.8000,"publicationDate":"2026-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Strahlentherapie und Onkologie","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00066-026-02585-2","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Surface-guided radiation therapy (SGRT) is a non-invasive, tattoo-less technique that enables real-time optical surface monitoring during radiotherapy. This prospective E‑SAMM study compared inter-fraction set-up accuracy, estimated PTV margin requirements derived from CBCT-based pre-correction setup deviations, and set-up time between maskless SGRT and mask-based setup in patients receiving pelvic radiotherapy.

Methods: One hundred patients with pelvic malignancies (cervical, endometrial, rectal, and prostate cancers) undergoing radical or neoadjuvant radiotherapy (45-68 Gy delivered in 25-28 fractions, five fractions per week) were prospectively enrolled. A total of 2376 fractions were analysed. Patients were randomised to either maskless SGRT with AlignRT, knee support, and intrafraction monitoring (Group A, n = 50) or six-clamp mask-based laser alignment with image-guided radiotherapy (Group B, n = 50). All patients underwent online cone-beam computed tomography (CBCT) verification on a TrueBeam SVC linear accelerator with a six-degree-of-freedom couch. Inter-fraction translational (medio-lateral [ML], cranio-caudal [CC], antero-posterior [AP]) and rotational set-up errors were derived from CBCT corrections. PTV margin estimates were calculated using the van Herk formula from CBCT-recorded pre-correction setup deviations. Statistical comparisons were performed using analysis of variance and the Mann-Whitney U test.

Results: SGRT demonstrated significantly improved translational accuracy in the ML (1.97 ± 1.54 mm vs 2.63 ± 2.06 mm, P = 0.03) and CC (3.13 ± 2.34 mm vs 4.03 ± 2.67 mm, P = 0.001) directions compared with mask-based alignment, whereas larger shifts were observed in the AP direction (P = 0.001). Rotational differences were not statistically significant (P = 0.14). The mean three-dimensional vector displacement was significantly lower with SGRT (1.91 ± 0.36 mm) than with mask-based positioning (2.26 ± 0.48 mm; P < 0.05). Estimated PTV margins were smaller with SGRT in all translational directions. Mean set-up time was significantly reduced with SGRT (2.51 ± 0.81 min vs 3.55 ± 0.99 min, P = 0.007).

Conclusions: Maskless SGRT provided improved inter-fraction set-up accuracy, lower estimated PTV margin requirements based on setup deviations, and shorter set-up times compared with mask-based laser alignment for pelvic radiotherapy.

提高分数间设置精度和边缘最小化(E-SAMM):盆腔放疗中无面罩表面引导和基于面罩的激光对准定位的比较。
目的:表面引导放射治疗(SGRT)是一种无创、无纹身的技术,可在放射治疗期间实现实时光学表面监测。这项前瞻性E - SAMM研究比较了接受盆腔放疗的患者的分数间设置准确性、基于cbct的预校正设置偏差得出的PTV边缘要求,以及无掩膜SGRT和基于掩膜的设置之间的设置时间。方法:前瞻性纳入100例盆腔恶性肿瘤(宫颈、子宫内膜、直肠和前列腺癌)患者,接受根治性或新辅助放疗(45-68 Gy,分25-28次,每周5次)。共分析了2376个馏分。患者被随机分配到使用AlignRT、膝关节支持和屈光内监测的无面罩SGRT (A组, = 50)或基于六钳面罩的激光对准与图像引导放疗(B组, = 50)。所有患者均在TrueBeam SVC线性加速器和六自由度沙发上进行在线锥束计算机断层扫描(CBCT)验证。分数间平移(中外侧[ML],颅尾[CC],前后[AP])和旋转设置误差来源于CBCT校正。根据cbct记录的预校正设置偏差,使用van Herk公式计算PTV边际估计。采用方差分析和Mann-Whitney U检验进行统计学比较。结果:SGRT证明显著提高转化精度的毫升(1.97  ±1.54毫米和2.63  ±2.06毫米,P = 0.03)和CC(3.13  ±2.34毫米和4.03  ±2.67毫米,P = 0.001)的方向与mask-based对齐,而较大的变化观察方向美联社 (P = 0.001)。旋转差异无统计学意义(P = 0.14)。SGRT的平均三维矢量位移(1.91 ±0.36 mm)明显低于基于掩膜定位(2.26 ±0.48 mm); P 结论:与基于掩膜的激光对中盆腔放疗相比,无掩膜SGRT提供了更高的分段间设置精度,更低的基于设置偏差的PTV margin要求,以及更短的设置时间。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.70
自引率
12.90%
发文量
141
审稿时长
3-8 weeks
期刊介绍: Strahlentherapie und Onkologie, published monthly, is a scientific journal that covers all aspects of oncology with focus on radiooncology, radiation biology and radiation physics. The articles are not only of interest to radiooncologists but to all physicians interested in oncology, to radiation biologists and radiation physicists. The journal publishes original articles, review articles and case studies that are peer-reviewed. It includes scientific short communications as well as a literature review with annotated articles that inform the reader on new developments in the various disciplines concerned and hence allow for a sound overview on the latest results in radiooncology research. Founded in 1912, Strahlentherapie und Onkologie is the oldest oncological journal in the world. Today, contributions are published in English and German. All articles have English summaries and legends. The journal is the official publication of several scientific radiooncological societies and publishes the relevant communications of these societies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书