From earlier to updated quadrifocal intraocular lens designs: objective optical metrology analysis.

IF 3.7 3区 医学 Q2 OPHTHALMOLOGY
Eun Ah Jeong, Seung Pil Bang, Seth Pantanelli, Sujin Kang, Sonia H Yoo
{"title":"From earlier to updated quadrifocal intraocular lens designs: objective optical metrology analysis.","authors":"Eun Ah Jeong, Seung Pil Bang, Seth Pantanelli, Sujin Kang, Sonia H Yoo","doi":"10.1097/j.jcrs.0000000000002042","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the optical performance of the Clareon PanOptix and Clareon PanOptix Pro intraocular lenses (IOLs) using interferometric metrology and identify measurable optical differences.</p><p><strong>Setting: </strong>Physiological Optics Laboratory, Daegu, South Korea.</p><p><strong>Design: </strong>Experimental optical-bench investigation.</p><p><strong>Methods: </strong>3 +20.0-diopter samples each of the PanOptix and PanOptix Pro IOLs were analyzed using a Mach-Zehnder interferometer in a model eye incorporating +0.27 μm corneal spherical aberration (SA). Microscopic imaging, wavefront reconstruction with Zernike analysis, diffractive-step profile reconstruction, point-spread function (PSF)-based retinal image simulation, halo analysis, and through-focus modulation transfer function (MTF) analysis were performed.</p><p><strong>Results: </strong>Both IOLs showed broadly similar diffractive architecture and SA profiles. Peak-to-valley step-profile amplitudes were comparable, although the PanOptix Pro showed localized differences in the reconstructed step-height contour. PSF-based analyses showed broadly similar focal behavior, but the PanOptix Pro demonstrated clearer Sloan F contour preservation and more evident central PSF-core brightness at the far-intermediate position. In through-focus average MTF (aMTF) profiles, the PanOptix Pro showed a modestly higher far-focus peak and a less pronounced post-peak dip, particularly at 4.5 and 3.0 mm pupils. MTF50 analysis showed a similar overall pattern, with greater variability and more apparent intermodel differences at larger pupil sizes.</p><p><strong>Conclusions: </strong>The PanOptix Pro preserves the fundamental design and SA profile of the original PanOptix while showing localized diffractive step-height modifications. Together with clearer far-intermediate F-image preservation, sustained central PSF-core brightness, and a shallower post-peak aMTF dip, these findings are consistent with a relative redistribution of optical energy toward the far-intermediate region, supporting greater far-intermediate optical continuity without substantial changes in nominal add power or overall focal structure.</p>","PeriodicalId":15214,"journal":{"name":"Journal of cataract and refractive surgery","volume":" ","pages":""},"PeriodicalIF":3.7000,"publicationDate":"2026-08-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cataract and refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/j.jcrs.0000000000002042","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To compare the optical performance of the Clareon PanOptix and Clareon PanOptix Pro intraocular lenses (IOLs) using interferometric metrology and identify measurable optical differences.

Setting: Physiological Optics Laboratory, Daegu, South Korea.

Design: Experimental optical-bench investigation.

Methods: 3 +20.0-diopter samples each of the PanOptix and PanOptix Pro IOLs were analyzed using a Mach-Zehnder interferometer in a model eye incorporating +0.27 μm corneal spherical aberration (SA). Microscopic imaging, wavefront reconstruction with Zernike analysis, diffractive-step profile reconstruction, point-spread function (PSF)-based retinal image simulation, halo analysis, and through-focus modulation transfer function (MTF) analysis were performed.

Results: Both IOLs showed broadly similar diffractive architecture and SA profiles. Peak-to-valley step-profile amplitudes were comparable, although the PanOptix Pro showed localized differences in the reconstructed step-height contour. PSF-based analyses showed broadly similar focal behavior, but the PanOptix Pro demonstrated clearer Sloan F contour preservation and more evident central PSF-core brightness at the far-intermediate position. In through-focus average MTF (aMTF) profiles, the PanOptix Pro showed a modestly higher far-focus peak and a less pronounced post-peak dip, particularly at 4.5 and 3.0 mm pupils. MTF50 analysis showed a similar overall pattern, with greater variability and more apparent intermodel differences at larger pupil sizes.

Conclusions: The PanOptix Pro preserves the fundamental design and SA profile of the original PanOptix while showing localized diffractive step-height modifications. Together with clearer far-intermediate F-image preservation, sustained central PSF-core brightness, and a shallower post-peak aMTF dip, these findings are consistent with a relative redistribution of optical energy toward the far-intermediate region, supporting greater far-intermediate optical continuity without substantial changes in nominal add power or overall focal structure.

从早期到更新的四焦人工晶状体设计:客观光学计量分析。
目的:采用干涉测量法比较Clareon PanOptix和Clareon PanOptix Pro人工晶状体(iol)的光学性能,确定可测量的光学差异。单位:韩国大邱生理光学实验室。设计:实验光学台架研究。方法:在角膜球差为+0.27µm的模型眼中,采用Mach-Zehnder干涉仪对3个+20.0 D的PanOptix和PanOptix Pro iol (Alcon)样品进行分析。显微成像、Zernike分析的波前重建、衍射阶像重建、基于点扩散函数(PSF)的视网膜图像模拟、光晕分析和透焦调制传递函数(TF-MTF)分析。结果:两种iol的衍射结构和球差分布大致相似。尽管PanOptix Pro显示重建的步高轮廓存在局部差异,但峰谷步高轮廓振幅具有可比性。基于psf的分析显示出大致相似的焦点行为,但PanOptix Pro显示出更清晰的斯隆F轮廓保存和更明显的远中间位置的中央psf核心亮度。在通焦平均MTF (aMTF)剖面中,PanOptix Pro显示出略高的远焦峰值和不太明显的峰后下降,特别是在4.5和3.0 mm瞳孔处。MTF50分析显示了类似的总体模式,在较大的瞳孔尺寸下具有更大的变异性和更明显的模型间差异。结论:PanOptix Pro保留了原始PanOptix的基本设计和球差轮廓,同时显示了局部的衍射步高修改。再加上更清晰的远中程f图像保存,持续的中央psf核心亮度,以及更浅的峰后aMTF下降,这些发现与光能向远中程区域的相对重新分配相一致,支持更大的远中程光学连续性,而名义附加功率或整体焦结构没有实质性变化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.60
自引率
14.30%
发文量
259
审稿时长
8.5 weeks
期刊介绍: The Journal of Cataract & Refractive Surgery (JCRS), a preeminent peer-reviewed monthly ophthalmology publication, is the official journal of the American Society of Cataract and Refractive Surgery (ASCRS) and the European Society of Cataract and Refractive Surgeons (ESCRS). JCRS publishes high quality articles on all aspects of anterior segment surgery. In addition to original clinical studies, the journal features a consultation section, practical techniques, important cases, and reviews as well as basic science articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书