Spectral Black-Blood Photon-Counting CT for High-Resolution Carotid Vessel Wall Imaging: A Feasibility Study.

IF 7.9 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Adrienn Tóth, Muhammad Taha Hagar, Moritz C Halfmann, Sardi Hyska, Eric Bass, Pooyan Sahbaee, Markus Jürgens, Milan Vecsey-Nagy, Tilman Emrich, Pal Maurovich-Horvat, Akos Varga-Szemes, Maria Vittoria Spampinato
{"title":"Spectral Black-Blood Photon-Counting CT for High-Resolution Carotid Vessel Wall Imaging: A Feasibility Study.","authors":"Adrienn Tóth, Muhammad Taha Hagar, Moritz C Halfmann, Sardi Hyska, Eric Bass, Pooyan Sahbaee, Markus Jürgens, Milan Vecsey-Nagy, Tilman Emrich, Pal Maurovich-Horvat, Akos Varga-Szemes, Maria Vittoria Spampinato","doi":"10.1097/RLI.0000000000001309","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To evaluate the feasibility of a novel single-phase spectral black-blood CT (BBCT) technique for high-resolution carotid vessel wall imaging and explore its potential impact on plaque feature visualization.</p><p><strong>Methods: </strong>In this IRB-approved retrospective study, clinically indicated head and neck photon-counting detector (PCD)-CT angiography (CTA) examinations were used to generate high-resolution BBCT reconstructions from a single angiographic-phase data set using spectral iodine suppression. Nine iodine ratio (IR) settings were evaluated to determine the balance between intraluminal suppression and vessel wall preservation. Objective image quality was assessed using attenuation and contrast-to-noise ratio (CNR). Three readers independently assessed overall image quality, lumen visualization, wall-lumen delineation, plaque conspicuity, and high-risk plaque features, including intraplaque hemorrhage (IPH), lipid-rich necrotic core (LRNC), and thin fibrous cap, and compared BBCT with conventional CTA reconstructions (0.4 mm, Qr48 kernel). Inter-reader agreement was assessed using Gwet's agreement coefficient (AC).</p><p><strong>Results: </strong>A total of 40 patients (mean age: 64.1±11.3 y; 57.5% women) were included. Lumen attenuation progressively decreased with lower IR values (range: 51.0 to -176.6 HU), whereas vessel wall attenuation remained relatively preserved (range: 11.4 to 55.8 HU, P=0.004). Lumen-to-wall CNR decreased from 13.6±6.1 to 2.1±1.7, whereas wall-to-fat CNR increased from 7.2±3.8 to 15.3±6.8, with IR 2.0 representing the best balance between lumen suppression and wall conspicuity. Compared with conventional CTA, BBCT improved wall-lumen delineation across readers (median score: 2 to 3 vs. 4; all P<0.001) and plaque conspicuity in evaluable plaques (median score: 3 vs. 4; all P≤0.03), whereas overall image quality was preserved (all P≥0.45). Agreement improved for IPH (AC: 0.63 vs. 0.45), LRNC (0.62 vs. 0.50), and thin fibrous cap assessment (0.70 vs. 0.33). However, the number of cases classified as positive for IPH varied across readers and was lower on BBCT (1 to 9 cases) than on CTA (11 to 14 cases), suggesting possible subtraction-related sensitivity loss.</p><p><strong>Conclusions: </strong>Spectral BBCT was feasible for high-resolution carotid vessel wall imaging, with preserved overall image quality, improved wall-lumen delineation, and higher plaque conspicuity compared with conventional CTA. Improved agreement for selected plaque features supports its potential as a complementary vessel wall reconstruction; however, possible subtraction-related sensitivity loss suggests BBCT should be interpreted in conjunction with conventional CTA rather than as a standalone reconstruction.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2026-08-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001309","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To evaluate the feasibility of a novel single-phase spectral black-blood CT (BBCT) technique for high-resolution carotid vessel wall imaging and explore its potential impact on plaque feature visualization.

Methods: In this IRB-approved retrospective study, clinically indicated head and neck photon-counting detector (PCD)-CT angiography (CTA) examinations were used to generate high-resolution BBCT reconstructions from a single angiographic-phase data set using spectral iodine suppression. Nine iodine ratio (IR) settings were evaluated to determine the balance between intraluminal suppression and vessel wall preservation. Objective image quality was assessed using attenuation and contrast-to-noise ratio (CNR). Three readers independently assessed overall image quality, lumen visualization, wall-lumen delineation, plaque conspicuity, and high-risk plaque features, including intraplaque hemorrhage (IPH), lipid-rich necrotic core (LRNC), and thin fibrous cap, and compared BBCT with conventional CTA reconstructions (0.4 mm, Qr48 kernel). Inter-reader agreement was assessed using Gwet's agreement coefficient (AC).

Results: A total of 40 patients (mean age: 64.1±11.3 y; 57.5% women) were included. Lumen attenuation progressively decreased with lower IR values (range: 51.0 to -176.6 HU), whereas vessel wall attenuation remained relatively preserved (range: 11.4 to 55.8 HU, P=0.004). Lumen-to-wall CNR decreased from 13.6±6.1 to 2.1±1.7, whereas wall-to-fat CNR increased from 7.2±3.8 to 15.3±6.8, with IR 2.0 representing the best balance between lumen suppression and wall conspicuity. Compared with conventional CTA, BBCT improved wall-lumen delineation across readers (median score: 2 to 3 vs. 4; all P<0.001) and plaque conspicuity in evaluable plaques (median score: 3 vs. 4; all P≤0.03), whereas overall image quality was preserved (all P≥0.45). Agreement improved for IPH (AC: 0.63 vs. 0.45), LRNC (0.62 vs. 0.50), and thin fibrous cap assessment (0.70 vs. 0.33). However, the number of cases classified as positive for IPH varied across readers and was lower on BBCT (1 to 9 cases) than on CTA (11 to 14 cases), suggesting possible subtraction-related sensitivity loss.

Conclusions: Spectral BBCT was feasible for high-resolution carotid vessel wall imaging, with preserved overall image quality, improved wall-lumen delineation, and higher plaque conspicuity compared with conventional CTA. Improved agreement for selected plaque features supports its potential as a complementary vessel wall reconstruction; however, possible subtraction-related sensitivity loss suggests BBCT should be interpreted in conjunction with conventional CTA rather than as a standalone reconstruction.

光谱黑血光子计数CT用于高分辨率颈动脉血管壁成像的可行性研究。
目的:评价一种新型的单相光谱黑血CT (BBCT)技术用于高分辨率颈动脉血管壁成像的可行性,并探讨其对斑块特征可视化的潜在影响。方法:在这项经irb批准的回顾性研究中,临床指示的头颈部光子计数检测器(PCD)-CT血管造影(CTA)检查使用光谱碘抑制从单一血管造影阶段数据集生成高分辨率BBCT重建。评估了9种碘比(IR)设置,以确定腔内抑制和血管壁保护之间的平衡。客观图像质量评估采用衰减和对比噪声比(CNR)。三位读者独立评估了整体图像质量、管腔可视化、壁-管腔描绘、斑块显著性和高危斑块特征,包括斑块内出血(IPH)、富含脂质坏死核心(LRNC)和薄纤维帽,并将BBCT与传统CTA重建(0.4 mm, Qr48内核)进行了比较。使用Gwet协议系数(AC)评估读者间协议。结果:共纳入40例患者,平均年龄:64.1±11.3岁,女性占57.5%。随着IR值的降低,管腔衰减逐渐减少(范围:51.0至-176.6 HU),而血管壁衰减相对保持不变(范围:11.4至55.8 HU, P=0.004)。管腔对壁的CNR从13.6±6.1下降到2.1±1.7,而壁对脂肪的CNR从7.2±3.8增加到15.3±6.8,IR 2.0代表了管腔抑制和壁显著性之间的最佳平衡。与传统CTA相比,BBCT改善了颈动脉壁腔的描绘(中位评分:2至3分vs. 4分)。结论:光谱BBCT用于高分辨率颈动脉壁成像是可行的,与传统CTA相比,它保留了整体图像质量,改善了壁腔的描绘,并且斑块的显著性更高。对选定斑块特征的改进协议支持其作为补充血管壁重建的潜力;然而,可能的减法相关的灵敏度损失表明,BBCT应与常规CTA一起解释,而不是作为单独的重建。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Investigative Radiology
Investigative Radiology 医学-核医学
CiteScore
15.10
自引率
16.40%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书