Hongkyung Kim, Oh Joo Kweon, Sumi Yoon, Yong Kwan Lim, Bohyun Kim
{"title":"Digital Morphology Analyzer-Based White Blood Cell Differentials: Practical Considerations for External Quality Assessment.","authors":"Hongkyung Kim, Oh Joo Kweon, Sumi Yoon, Yong Kwan Lim, Bohyun Kim","doi":"10.3343/alm.2026.0097","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Digital morphology (DM) analyzers are increasingly used for white blood cell (WBC) differential counting in routine hematology laboratories, necessitating tailored external quality assessment (EQA) schemes. We evaluated EQA-relevant practical considerations for DM analyzer-based WBC differentials.</p><p><strong>Methods: </strong>Fifteen clinical laboratories participated in a multicenter EQA simulation using five centrally prepared, unstained peripheral blood smear samples. Institution-level post-classification results were analyzed as the primary outcomes. Exploratory analyses were performed using pre-classification and manual microscopic differentials. Examiner-level variability within institutions was evaluated, along with the distributions of WBC-classified and unclassified images following examiner verification.</p><p><strong>Results: </strong>Most participating laboratories used DM analyzers primarily for screening WBC differentials, with predefined criteria for manual microscopic reviews, particularly when abnormal cell populations were noted. Inter-institutional and examiner-level variability were most pronounced for morphologically similar cell classes. Post-classification results were generally comparable with those of manual differentials across most cell classes. The proportion of unclassified images (including artifacts) increased, suggesting that operational and sample-related factors influenced the interpretation of the EQA results.</p><p><strong>Conclusions: </strong>We identified key practical considerations for the EQA of DM analyzer-based WBC differential counting. An appropriate EQA design should incorporate clinically relevant abnormal cell populations and challenging sample types; data interpretation should extend beyond standard deviation index-based metrics to address abnormal cell detection. The overall concordance between post-classification results and manual differentials supports the importance of periodic verification, and slide quality should be considered when implementing DM analyzer-based EQA programs.</p>","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2026.0097","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Digital morphology (DM) analyzers are increasingly used for white blood cell (WBC) differential counting in routine hematology laboratories, necessitating tailored external quality assessment (EQA) schemes. We evaluated EQA-relevant practical considerations for DM analyzer-based WBC differentials.
Methods: Fifteen clinical laboratories participated in a multicenter EQA simulation using five centrally prepared, unstained peripheral blood smear samples. Institution-level post-classification results were analyzed as the primary outcomes. Exploratory analyses were performed using pre-classification and manual microscopic differentials. Examiner-level variability within institutions was evaluated, along with the distributions of WBC-classified and unclassified images following examiner verification.
Results: Most participating laboratories used DM analyzers primarily for screening WBC differentials, with predefined criteria for manual microscopic reviews, particularly when abnormal cell populations were noted. Inter-institutional and examiner-level variability were most pronounced for morphologically similar cell classes. Post-classification results were generally comparable with those of manual differentials across most cell classes. The proportion of unclassified images (including artifacts) increased, suggesting that operational and sample-related factors influenced the interpretation of the EQA results.
Conclusions: We identified key practical considerations for the EQA of DM analyzer-based WBC differential counting. An appropriate EQA design should incorporate clinically relevant abnormal cell populations and challenging sample types; data interpretation should extend beyond standard deviation index-based metrics to address abnormal cell detection. The overall concordance between post-classification results and manual differentials supports the importance of periodic verification, and slide quality should be considered when implementing DM analyzer-based EQA programs.
期刊介绍:
Annals of Laboratory Medicine is the official journal of Korean Society for Laboratory Medicine. The journal title has been recently changed from the Korean Journal of Laboratory Medicine (ISSN, 1598-6535) from the January issue of 2012. The JCR 2017 Impact factor of Ann Lab Med was 1.916.