Sensitivity Enhancement by Dual Deflection Electrodes in an Electrospray Ionization Source

IF 1.8 3区 化学 Q4 BIOCHEMICAL RESEARCH METHODS
Motoki Date, Shun Kumano, Masuyuki Sugiyama, Hideki Hasegawa, Masaki Watanabe
{"title":"Sensitivity Enhancement by Dual Deflection Electrodes in an Electrospray Ionization Source","authors":"Motoki Date,&nbsp;Shun Kumano,&nbsp;Masuyuki Sugiyama,&nbsp;Hideki Hasegawa,&nbsp;Masaki Watanabe","doi":"10.1002/rcm.70159","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Rationale</h3>\n \n <p>In electrospray ionization–mass spectrometry (ESI-MS), ESI configurations such as orthogonal spray arrangements are often used to suppress droplet introduction into the vacuum chamber, but they can also reduce the ion flux into the vacuum region. Previous methods that electrostatically deflected ions toward the vacuum inlet using a single electrode achieved signal enhancement but could not adequately control deflection position, limiting performance. Therefore, we investigated whether adding a second deflection electrode to create a more localized electric field could further increase ion flux into the vacuum region and enhance sensitivity.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Electrostatic analysis was performed using a two-dimensional cross-sectional model for three ion-source configurations: no electrode, single-plate electrode, and dual electrodes (plate and cylindrical rod). For experimental evaluation, direct-infusion experiments (100 μL min<sup>−1</sup>) were performed using reserpine (<i>m</i>/<i>z</i> = 609.3) and a CsI/diethylamine mixture spanning <i>m</i>/<i>z</i> values of 74–912, and electrode voltages were stepped (E1, 0–5 kV; E2, 1–3 kV). Flow-injection analysis (FIA) of reserpine (0.5–20 ng mL<sup>−1</sup>) was conducted to evaluate quantitative performance.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Electrostatic analysis showed that the dual-electrode configuration concentrated the inlet-directed field near the inlet and suppressed the unwanted field above it. Experimentally, the dual-electrode configuration increased the reserpine signal by 3.3-fold compared with the configuration without electrodes. Across five chemically diverse ions, signal enhancements were obtained. For singly charged species, the gain scaled inversely with collision cross-section, indicating a specific benefit for low-ion-mobility analytes. FIA yielded a 4.0-fold increase in the calibration curve slope and halved the lower limit of quantification without degrading linearity (<i>R</i><sup>2</sup> = 0.998).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Localizing the electric field using two deflection electrodes enhances ion signal intensity, sensitivity, and quantitative limits in the ESI source configuration examined here. These results indicate that spatial control of the deflection field is an effective strategy for improving ion transmission in an orthogonal ESI source.</p>\n </section>\n </div>","PeriodicalId":225,"journal":{"name":"Rapid Communications in Mass Spectrometry","volume":"40 21","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2026-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rapid Communications in Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/rcm.70159","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

Abstract

Rationale

In electrospray ionization–mass spectrometry (ESI-MS), ESI configurations such as orthogonal spray arrangements are often used to suppress droplet introduction into the vacuum chamber, but they can also reduce the ion flux into the vacuum region. Previous methods that electrostatically deflected ions toward the vacuum inlet using a single electrode achieved signal enhancement but could not adequately control deflection position, limiting performance. Therefore, we investigated whether adding a second deflection electrode to create a more localized electric field could further increase ion flux into the vacuum region and enhance sensitivity.

Methods

Electrostatic analysis was performed using a two-dimensional cross-sectional model for three ion-source configurations: no electrode, single-plate electrode, and dual electrodes (plate and cylindrical rod). For experimental evaluation, direct-infusion experiments (100 μL min−1) were performed using reserpine (m/z = 609.3) and a CsI/diethylamine mixture spanning m/z values of 74–912, and electrode voltages were stepped (E1, 0–5 kV; E2, 1–3 kV). Flow-injection analysis (FIA) of reserpine (0.5–20 ng mL−1) was conducted to evaluate quantitative performance.

Results

Electrostatic analysis showed that the dual-electrode configuration concentrated the inlet-directed field near the inlet and suppressed the unwanted field above it. Experimentally, the dual-electrode configuration increased the reserpine signal by 3.3-fold compared with the configuration without electrodes. Across five chemically diverse ions, signal enhancements were obtained. For singly charged species, the gain scaled inversely with collision cross-section, indicating a specific benefit for low-ion-mobility analytes. FIA yielded a 4.0-fold increase in the calibration curve slope and halved the lower limit of quantification without degrading linearity (R2 = 0.998).

Conclusions

Localizing the electric field using two deflection electrodes enhances ion signal intensity, sensitivity, and quantitative limits in the ESI source configuration examined here. These results indicate that spatial control of the deflection field is an effective strategy for improving ion transmission in an orthogonal ESI source.

双偏转电极在电喷雾电离源中的灵敏度增强
在电喷雾电离-质谱(ESI- ms)中,通常采用正交喷雾排列等ESI配置来抑制液滴进入真空室,但它们也会降低进入真空区的离子通量。以前的方法是使用单个电极将离子静电偏转到真空入口,从而实现了信号增强,但不能充分控制偏转位置,从而限制了性能。因此,我们研究了添加第二个偏转电极以产生更局部化的电场是否可以进一步增加进入真空区域的离子通量并提高灵敏度。方法采用二维截面模型对无电极、单电极和双电极(板和柱棒)三种离子源配置进行静电分析。为了进行实验评估,采用利血平(m/z = 609.3)和CsI/二乙胺混合物(m/z值为74 ~ 912)进行直接输注实验(100 μL min−1),电极电压步进(E1, 0 ~ 5 kV; E2, 1 ~ 3 kV)。利血平(0.5 ~ 20 ng mL−1)进行流动注射分析(FIA)以评价定量性能。结果静电分析表明,双电极结构使入口附近的场集中,入口上方的场被抑制。实验结果表明,双电极配置比无电极配置增加了3.3倍的利血平信号。在五种化学不同的离子中,获得了信号增强。对于单电荷的物质,增益与碰撞截面成反比,表明对低离子迁移率的分析物有特殊的好处。FIA可使校准曲线斜率增加4.0倍,在不影响线性的情况下将定量下限减半(R2 = 0.998)。结论:在本研究中,使用两个偏转电极定位电场可以增强离子信号强度、灵敏度和ESI源配置的定量限制。这些结果表明,在正交ESI源中,对偏转场进行空间控制是提高离子透射率的有效策略。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.10
自引率
5.00%
发文量
219
审稿时长
2.6 months
期刊介绍: Rapid Communications in Mass Spectrometry is a journal whose aim is the rapid publication of original research results and ideas on all aspects of the science of gas-phase ions; it covers all the associated scientific disciplines. There is no formal limit on paper length ("rapid" is not synonymous with "brief"), but papers should be of a length that is commensurate with the importance and complexity of the results being reported. Contributions may be theoretical or practical in nature; they may deal with methods, techniques and applications, or with the interpretation of results; they may cover any area in science that depends directly on measurements made upon gaseous ions or that is associated with such measurements.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书