{"title":"Quantitative MALDI-MSI Workflow for Lipids with a LC-MS-Based Quality Control Pipeline","authors":"Felix-Levin Hormann, Sven Heiles","doi":"10.1021/acsmeasuresciau.6c00083","DOIUrl":null,"url":null,"abstract":"<p>Mass spectrometry imaging (MSI) has established itself\r\nas a major\r\nanalytical method in the spatial analysis of biological tissue. Besides\r\nqualitative mapping of analyte distributions, recent developments\r\naim to achieve quantification of target analytes. After successful\r\nspatial quantification of pharmaceuticals, the field is now shifting\r\ntoward endogenous biomolecules such as lipids and metabolites. Here,\r\nwe present a quantitative MSI (qMSI) workflow for endogenous lipids\r\nemploying matrix-assisted laser desorption/ionization (MALDI) and\r\nstable isotope labeled standards (SILS) integrated into a quality\r\ncontrol pipeline utilizing quantitative reversed-phase liquid chromatography\r\ntandem mass spectrometry (RP-LC-MS/MS). We thoroughly characterized\r\nthe standard deposition of 13 lipid SILS with a focus on homogeneity\r\nand detection errors on mouse heart septa with intraday coefficient\r\nof variations (CVs) of ≤15%. Unlike RP-LC-MS/MS for which CV\r\nvalues increase mainly for lowly abundant lipids, variations of lipids\r\nin MALDI-MSI are affected by concentration as well as lipid class.\r\nAdditionally, we compared lipid annotations based on MALDI-MSI and\r\nRP-LC-MS/MS and correlated these annotations between the methods based\r\non quantification results. This method comparison revealed that most\r\npolar lipids such as PCs or SMs are captured in both MALDI-MSI and\r\nRP-LC-MS/MS, whereas less polar compounds such as TGs exhibit limited\r\noverlap. These results highlight that a determination of the SILS\r\namounts and lipid annotations via RP-LC-MS/MS is necessary to counter\r\nthe increase variability during quantitative MALDI-MSI. To demonstrate\r\nthe analytical capabilities of the quantitative MALDI-MSI pipeline\r\ncombined with RP-LC-MS/MS quality control steps, we analyzed mouse\r\nkidney from a two-hit heart failure with preserved ejection fraction\r\nmodel and achieved good comparability between RP-LC-MS/MS and MALDI-MSI\r\nquantitation results. Our data suggest increased ether lipid concentrations\r\nin obese mice kidney, likely to cope with increased oxidative stress\r\ninduced by the progressing disease state of the mice.</p>","PeriodicalId":29800,"journal":{"name":"ACS Measurement Science Au","volume":"6 4","pages":"1194–1205"},"PeriodicalIF":14.7000,"publicationDate":"2026-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Measurement Science Au","FirstCategoryId":"1085","ListUrlMain":"https://pubs.acs.org/doi/10.1021/acsmeasuresciau.6c00083","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/6 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Mass spectrometry imaging (MSI) has established itself
as a major
analytical method in the spatial analysis of biological tissue. Besides
qualitative mapping of analyte distributions, recent developments
aim to achieve quantification of target analytes. After successful
spatial quantification of pharmaceuticals, the field is now shifting
toward endogenous biomolecules such as lipids and metabolites. Here,
we present a quantitative MSI (qMSI) workflow for endogenous lipids
employing matrix-assisted laser desorption/ionization (MALDI) and
stable isotope labeled standards (SILS) integrated into a quality
control pipeline utilizing quantitative reversed-phase liquid chromatography
tandem mass spectrometry (RP-LC-MS/MS). We thoroughly characterized
the standard deposition of 13 lipid SILS with a focus on homogeneity
and detection errors on mouse heart septa with intraday coefficient
of variations (CVs) of ≤15%. Unlike RP-LC-MS/MS for which CV
values increase mainly for lowly abundant lipids, variations of lipids
in MALDI-MSI are affected by concentration as well as lipid class.
Additionally, we compared lipid annotations based on MALDI-MSI and
RP-LC-MS/MS and correlated these annotations between the methods based
on quantification results. This method comparison revealed that most
polar lipids such as PCs or SMs are captured in both MALDI-MSI and
RP-LC-MS/MS, whereas less polar compounds such as TGs exhibit limited
overlap. These results highlight that a determination of the SILS
amounts and lipid annotations via RP-LC-MS/MS is necessary to counter
the increase variability during quantitative MALDI-MSI. To demonstrate
the analytical capabilities of the quantitative MALDI-MSI pipeline
combined with RP-LC-MS/MS quality control steps, we analyzed mouse
kidney from a two-hit heart failure with preserved ejection fraction
model and achieved good comparability between RP-LC-MS/MS and MALDI-MSI
quantitation results. Our data suggest increased ether lipid concentrations
in obese mice kidney, likely to cope with increased oxidative stress
induced by the progressing disease state of the mice.
期刊介绍:
ACS Measurement Science Au is an open access journal that publishes experimental computational or theoretical research in all areas of chemical measurement science. Short letters comprehensive articles reviews and perspectives are welcome on topics that report on any phase of analytical operations including sampling measurement and data analysis. This includes:Chemical Reactions and SelectivityChemometrics and Data ProcessingElectrochemistryElemental and Molecular CharacterizationImagingInstrumentationMass SpectrometryMicroscale and Nanoscale systemsOmics (Genomics Proteomics Metabonomics Metabolomics and Bioinformatics)Sensors and Sensing (Biosensors Chemical Sensors Gas Sensors Intracellular Sensors Single-Molecule Sensors Cell Chips Arrays Microfluidic Devices)SeparationsSpectroscopySurface analysisPapers dealing with established methods need to offer a significantly improved original application of the method.