Performance and Synchronization of Portable Oxygen Concentrators Delivering Pulsed Oxygen Therapy.

IF 8.7 3区 医学 Q1 RESPIRATORY SYSTEM
Marius Lebret, Emeline Fresnel, Jacqueline Delrieu, Maxime Patout, Claudio Rabec
{"title":"Performance and Synchronization of Portable Oxygen Concentrators Delivering Pulsed Oxygen Therapy.","authors":"Marius Lebret, Emeline Fresnel, Jacqueline Delrieu, Maxime Patout, Claudio Rabec","doi":"10.1016/j.arbres.2026.07.009","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>This study evaluated the performance of seven POCs, focusing on FiO<sub>2</sub>, bolus triggering delay, synchronization, and technical specifications, including weight, size, noise level, and battery autonomy.</p><p><strong>Methods: </strong>A bench study was conducted using a mechanical lung model (ASL 5000) equipped with a Fast Oxygen Measurement sensor. Seven devices available on the European market were tested: FreeStyle Comfort, iGo2, Inogen One G4, Inogen One G5, Platinum Mobile, Zen-O lite, and Horizon P5. 200 ventilatory cycles were simulated per device across various respiratory rates (RR). Measurements included FiO<sub>2</sub>, bolus triggering delay, synchronization, and bolus volume. Technical specifications were also compared. All devices were tested across their full range of settings, with each test replicated on a second device to confirm results within a 10% margin of error.</p><p><strong>Results: </strong>Significant differences in FiO<sub>2</sub> delivery were observed (p<0.001), with all devices delivering at least 28% FiO<sub>2</sub> at their lowest settings. The Inogen One G5 and Horizon P5 achieved the highest FiO<sub>2</sub> levels (48.7% [38.4-56.9%] and 47.8% [35.6-56.4%], respectively). Synchronization was high across devices but declined at extreme RR (15 and 40bpm). FiO<sub>2</sub> variations were influenced by RR, while bolus delays depended on device type and RR. Technical specifications varied widely across devices.</p><p><strong>Conclusions: </strong>POCs demonstrate notable differences in performance, with RR and device type impacting FiO<sub>2</sub> delivery and synchronization. These findings may help support more informed device selection according to the intended conditions of use and patient requirements.</p>","PeriodicalId":8339,"journal":{"name":"Archivos De Bronconeumologia","volume":" ","pages":""},"PeriodicalIF":8.7000,"publicationDate":"2026-08-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archivos De Bronconeumologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.arbres.2026.07.009","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: This study evaluated the performance of seven POCs, focusing on FiO2, bolus triggering delay, synchronization, and technical specifications, including weight, size, noise level, and battery autonomy.

Methods: A bench study was conducted using a mechanical lung model (ASL 5000) equipped with a Fast Oxygen Measurement sensor. Seven devices available on the European market were tested: FreeStyle Comfort, iGo2, Inogen One G4, Inogen One G5, Platinum Mobile, Zen-O lite, and Horizon P5. 200 ventilatory cycles were simulated per device across various respiratory rates (RR). Measurements included FiO2, bolus triggering delay, synchronization, and bolus volume. Technical specifications were also compared. All devices were tested across their full range of settings, with each test replicated on a second device to confirm results within a 10% margin of error.

Results: Significant differences in FiO2 delivery were observed (p<0.001), with all devices delivering at least 28% FiO2 at their lowest settings. The Inogen One G5 and Horizon P5 achieved the highest FiO2 levels (48.7% [38.4-56.9%] and 47.8% [35.6-56.4%], respectively). Synchronization was high across devices but declined at extreme RR (15 and 40bpm). FiO2 variations were influenced by RR, while bolus delays depended on device type and RR. Technical specifications varied widely across devices.

Conclusions: POCs demonstrate notable differences in performance, with RR and device type impacting FiO2 delivery and synchronization. These findings may help support more informed device selection according to the intended conditions of use and patient requirements.

提供脉冲氧治疗的便携式氧气浓缩器的性能和同步。
目的:本研究评估了7种POCs的性能,重点关注FiO2,丸触发延迟,同步和技术规格,包括重量,尺寸,噪音水平和电池自主性。方法:采用配备快速氧测量传感器的机械肺模型(asl5000)进行台架研究。测试了欧洲市场上的七种设备:FreeStyle Comfort, iGo2, Inogen One G4, Inogen One G5, Platinum Mobile, Zen-O lite和Horizon P5。每个设备在不同呼吸速率(RR)下模拟200个呼吸周期。测量包括FiO2、丸触发延迟、同步和丸体积。技术指标也进行了比较。所有设备都在其全部设置范围内进行了测试,每次测试都在另一台设备上重复进行,以确认结果在10%的误差范围内。结果:在最低设置下,观察到FiO2交付量的显著差异(p2)。Inogen One G5和Horizon P5的FiO2水平最高,分别为48.7%[38.4-56.9%]和47.8%[35.6-56.4%]。设备间的同步度很高,但在极端RR(15和40bpm)时下降。FiO2的变化受RR的影响,而bolus延迟则取决于器件类型和RR。不同设备的技术规格差异很大。结论:POCs表现出显著的性能差异,RR和设备类型影响FiO2输送和同步。这些发现可能有助于根据预期的使用条件和患者的要求进行更明智的器械选择。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Archivos De Bronconeumologia
Archivos De Bronconeumologia Medicine-Pulmonary and Respiratory Medicine
CiteScore
3.50
自引率
17.50%
发文量
330
审稿时长
14 days
期刊介绍: Archivos de Bronconeumologia is a scientific journal that specializes in publishing prospective original research articles focusing on various aspects of respiratory diseases, including epidemiology, pathophysiology, clinical practice, surgery, and basic investigation. Additionally, the journal features other types of articles such as reviews, editorials, special articles of interest to the society and editorial board, scientific letters, letters to the editor, and clinical images. Published monthly, the journal comprises 12 regular issues along with occasional supplements containing articles from different sections. All manuscripts submitted to the journal undergo rigorous evaluation by the editors and are subjected to expert peer review. The editorial team, led by the Editor and/or an Associate Editor, manages the peer-review process. Archivos de Bronconeumologia is published monthly in English, facilitating broad dissemination of the latest research findings in the field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书