Marius Lebret, Emeline Fresnel, Jacqueline Delrieu, Maxime Patout, Claudio Rabec
{"title":"Performance and Synchronization of Portable Oxygen Concentrators Delivering Pulsed Oxygen Therapy.","authors":"Marius Lebret, Emeline Fresnel, Jacqueline Delrieu, Maxime Patout, Claudio Rabec","doi":"10.1016/j.arbres.2026.07.009","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>This study evaluated the performance of seven POCs, focusing on FiO<sub>2</sub>, bolus triggering delay, synchronization, and technical specifications, including weight, size, noise level, and battery autonomy.</p><p><strong>Methods: </strong>A bench study was conducted using a mechanical lung model (ASL 5000) equipped with a Fast Oxygen Measurement sensor. Seven devices available on the European market were tested: FreeStyle Comfort, iGo2, Inogen One G4, Inogen One G5, Platinum Mobile, Zen-O lite, and Horizon P5. 200 ventilatory cycles were simulated per device across various respiratory rates (RR). Measurements included FiO<sub>2</sub>, bolus triggering delay, synchronization, and bolus volume. Technical specifications were also compared. All devices were tested across their full range of settings, with each test replicated on a second device to confirm results within a 10% margin of error.</p><p><strong>Results: </strong>Significant differences in FiO<sub>2</sub> delivery were observed (p<0.001), with all devices delivering at least 28% FiO<sub>2</sub> at their lowest settings. The Inogen One G5 and Horizon P5 achieved the highest FiO<sub>2</sub> levels (48.7% [38.4-56.9%] and 47.8% [35.6-56.4%], respectively). Synchronization was high across devices but declined at extreme RR (15 and 40bpm). FiO<sub>2</sub> variations were influenced by RR, while bolus delays depended on device type and RR. Technical specifications varied widely across devices.</p><p><strong>Conclusions: </strong>POCs demonstrate notable differences in performance, with RR and device type impacting FiO<sub>2</sub> delivery and synchronization. These findings may help support more informed device selection according to the intended conditions of use and patient requirements.</p>","PeriodicalId":8339,"journal":{"name":"Archivos De Bronconeumologia","volume":" ","pages":""},"PeriodicalIF":8.7000,"publicationDate":"2026-08-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archivos De Bronconeumologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.arbres.2026.07.009","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: This study evaluated the performance of seven POCs, focusing on FiO2, bolus triggering delay, synchronization, and technical specifications, including weight, size, noise level, and battery autonomy.
Methods: A bench study was conducted using a mechanical lung model (ASL 5000) equipped with a Fast Oxygen Measurement sensor. Seven devices available on the European market were tested: FreeStyle Comfort, iGo2, Inogen One G4, Inogen One G5, Platinum Mobile, Zen-O lite, and Horizon P5. 200 ventilatory cycles were simulated per device across various respiratory rates (RR). Measurements included FiO2, bolus triggering delay, synchronization, and bolus volume. Technical specifications were also compared. All devices were tested across their full range of settings, with each test replicated on a second device to confirm results within a 10% margin of error.
Results: Significant differences in FiO2 delivery were observed (p<0.001), with all devices delivering at least 28% FiO2 at their lowest settings. The Inogen One G5 and Horizon P5 achieved the highest FiO2 levels (48.7% [38.4-56.9%] and 47.8% [35.6-56.4%], respectively). Synchronization was high across devices but declined at extreme RR (15 and 40bpm). FiO2 variations were influenced by RR, while bolus delays depended on device type and RR. Technical specifications varied widely across devices.
Conclusions: POCs demonstrate notable differences in performance, with RR and device type impacting FiO2 delivery and synchronization. These findings may help support more informed device selection according to the intended conditions of use and patient requirements.
期刊介绍:
Archivos de Bronconeumologia is a scientific journal that specializes in publishing prospective original research articles focusing on various aspects of respiratory diseases, including epidemiology, pathophysiology, clinical practice, surgery, and basic investigation. Additionally, the journal features other types of articles such as reviews, editorials, special articles of interest to the society and editorial board, scientific letters, letters to the editor, and clinical images. Published monthly, the journal comprises 12 regular issues along with occasional supplements containing articles from different sections.
All manuscripts submitted to the journal undergo rigorous evaluation by the editors and are subjected to expert peer review. The editorial team, led by the Editor and/or an Associate Editor, manages the peer-review process. Archivos de Bronconeumologia is published monthly in English, facilitating broad dissemination of the latest research findings in the field.