Wear Rate, Particle Morphology, and Biocompatibility of Polyether-Ether-Ketone Femoral Heads Against Highly Cross-Linked Polyethylene Acetabular Cups in Hip Prosthetic Couples

IF 3 4区 医学 Q2 ENGINEERING, BIOMEDICAL
Fei Hu, Xinyue Zhang, Xiaobo Wu, Zhuocheng Lyu, Xu Zheng, Linjun Yang, Zikai Hua, Dekun Zhang, You Wang
{"title":"Wear Rate, Particle Morphology, and Biocompatibility of Polyether-Ether-Ketone Femoral Heads Against Highly Cross-Linked Polyethylene Acetabular Cups in Hip Prosthetic Couples","authors":"Fei Hu,&nbsp;Xinyue Zhang,&nbsp;Xiaobo Wu,&nbsp;Zhuocheng Lyu,&nbsp;Xu Zheng,&nbsp;Linjun Yang,&nbsp;Zikai Hua,&nbsp;Dekun Zhang,&nbsp;You Wang","doi":"10.1002/jbm.b.70139","DOIUrl":null,"url":null,"abstract":"<p>Biological reactions to wear particles from artificial joint prostheses can lead to osteolysis, ultimately resulting in aseptic loosening and implant failure, which affect the longevity of the prosthesis and the patient's quality of life. Polyether-ether-ketone (PEEK) has emerged as a key area of research in orthopedic implant materials due to its excellent wear resistance and biomechanical compatibility with human bone tissue. In this study, we conducted a 10-million-cycle simulator wear test on hip artificial joints with polyetheretherketone-on-highly cross-linked polyethylene (PEEK-on-HCLPE) bearing couplings for the first time. The results indicated the total wear loss of the highly cross-linked polyethylene (HCLPE) liners was 34.03 ± 3.2 mg (~2.96 mg/MC or 3.18 mm<sup>3</sup>/MC after linear fitting). The physicochemical properties of the mixed particles indicated a predominant particle size distribution ranging from 0.1 to 0.5 μm. Pure PEEK or HCLPE particles with the similar size range (maximum size: 0.5 μm) were generated by an EXAKT grinding system, followed by vacuum filtration with precision pore—size membranes for the biocompatibility experiments. In vitro immunological assays showed that 0.1 to 0.5 μm PEEK particles may induce a relatively weaker immune inflammatory effect than 0.1 to 0.5 μm HCLPE particles. In conclusion, the PEEK-on-HCLPE bearing couple has been demonstrated to have acceptable biosafety and a wear rate suitable for use as a potential hip replacement implant.</p>","PeriodicalId":15269,"journal":{"name":"Journal of biomedical materials research. Part B, Applied biomaterials","volume":"114 8","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2026-08-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jbm.b.70139","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biomedical materials research. Part B, Applied biomaterials","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jbm.b.70139","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Biological reactions to wear particles from artificial joint prostheses can lead to osteolysis, ultimately resulting in aseptic loosening and implant failure, which affect the longevity of the prosthesis and the patient's quality of life. Polyether-ether-ketone (PEEK) has emerged as a key area of research in orthopedic implant materials due to its excellent wear resistance and biomechanical compatibility with human bone tissue. In this study, we conducted a 10-million-cycle simulator wear test on hip artificial joints with polyetheretherketone-on-highly cross-linked polyethylene (PEEK-on-HCLPE) bearing couplings for the first time. The results indicated the total wear loss of the highly cross-linked polyethylene (HCLPE) liners was 34.03 ± 3.2 mg (~2.96 mg/MC or 3.18 mm3/MC after linear fitting). The physicochemical properties of the mixed particles indicated a predominant particle size distribution ranging from 0.1 to 0.5 μm. Pure PEEK or HCLPE particles with the similar size range (maximum size: 0.5 μm) were generated by an EXAKT grinding system, followed by vacuum filtration with precision pore—size membranes for the biocompatibility experiments. In vitro immunological assays showed that 0.1 to 0.5 μm PEEK particles may induce a relatively weaker immune inflammatory effect than 0.1 to 0.5 μm HCLPE particles. In conclusion, the PEEK-on-HCLPE bearing couple has been demonstrated to have acceptable biosafety and a wear rate suitable for use as a potential hip replacement implant.

聚醚醚酮股骨头与高交联聚乙烯髋臼杯的磨损率、颗粒形态和生物相容性。
人工关节假体磨损颗粒的生物反应可导致骨溶解,最终导致无菌性松动和假体失效,影响假体的使用寿命和患者的生活质量。聚醚醚酮(PEEK)由于其优异的耐磨性和与人体骨组织的生物力学相容性而成为骨科植入材料研究的一个重点领域。在本研究中,我们首次对采用聚醚醚酮-高交联聚乙烯(PEEK-on-HCLPE)轴承联轴器的髋关节人工关节进行了1000万次模拟器磨损试验。结果表明,高交联聚乙烯(HCLPE)衬垫的总磨损量为34.03±3.2 mg (~2.96 mg/MC,线性拟合后为3.18 mm3/MC)。混合颗粒的物理化学性质表明,粒径分布在0.1 ~ 0.5 μm之间。采用EXAKT研磨系统制备尺寸相近(最大尺寸为0.5 μm)的纯PEEK或HCLPE颗粒,并用精密孔径膜进行真空过滤,进行生物相容性实验。体外免疫实验表明,0.1 ~ 0.5 μm PEEK颗粒诱导的免疫炎症效应较0.1 ~ 0.5 μm HCLPE颗粒弱。总之,PEEK-on-HCLPE轴承偶已被证明具有可接受的生物安全性和磨损率,适合用作潜在的髋关节置换术植入物。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
7.50
自引率
2.90%
发文量
199
审稿时长
12 months
期刊介绍: Journal of Biomedical Materials Research – Part B: Applied Biomaterials is a highly interdisciplinary peer-reviewed journal serving the needs of biomaterials professionals who design, develop, produce and apply biomaterials and medical devices. It has the common focus of biomaterials applied to the human body and covers all disciplines where medical devices are used. Papers are published on biomaterials related to medical device development and manufacture, degradation in the body, nano- and biomimetic- biomaterials interactions, mechanics of biomaterials, implant retrieval and analysis, tissue-biomaterial surface interactions, wound healing, infection, drug delivery, standards and regulation of devices, animal and pre-clinical studies of biomaterials and medical devices, and tissue-biopolymer-material combination products. Manuscripts are published in one of six formats: • original research reports • short research and development reports • scientific reviews • current concepts articles • special reports • editorials Journal of Biomedical Materials Research – Part B: Applied Biomaterials is an official journal of the Society for Biomaterials, Japanese Society for Biomaterials, the Australasian Society for Biomaterials, and the Korean Society for Biomaterials. Manuscripts from all countries are invited but must be in English. Authors are not required to be members of the affiliated Societies, but members of these societies are encouraged to submit their work to the journal for consideration.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书