Identification of Clinically Relevant Hb Variants During HbA1c Testing Using the ARKRAY HA-8180V in Fast Mode: Reliable HbE Detection Based on Retention Time.
{"title":"Identification of Clinically Relevant Hb Variants During HbA1c Testing Using the ARKRAY HA-8180V in Fast Mode: Reliable HbE Detection Based on Retention Time.","authors":"Monthon Mayoon, Sitthichai Panyasai","doi":"10.3343/alm.2026.0183","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Hb variants are often incidentally detected during HbA1c testing and can cause interference. We evaluated the ability of the ARKRAY HA-8180V HPLC system operating in fast mode (FM) to identify Hb variants, particularly HbE, in samples submitted for HbA1c testing based on characteristic retention times (RTs).</p><p><strong>Methods: </strong>Chromatograms of samples tested for HbA1c were analyzed; those showing predefined abnormal chromatographic patterns suggestive of Hb variants were further evaluated using capillary electrophoresis (CE) and β- and α-globin gene sequencing. RTs were recorded.</p><p><strong>Results: </strong>Of 4,888 chromatograms analyzed, 588 (12.0%) were abnormal. Among these, 564 (95.9%) showed an abnormal peak at an RT of 23-25 s and were confirmed to contain HbE via CE. Identified Hb phenotypes included EA (N=526), compound heterozygotes [Constant Spring (CS) EA, N=2; CSEE, N=1], EE (N=27), EFA (N=3), and EF (N=5). Other variants included Hb CSA<sub>2</sub>ABart'sH, A<sub>2</sub>FA, A<sub>2</sub>FABart's, and A<sub>2</sub>F. Rare variants, Hb Hope (N=12) and Hb Korle-Bu (N=1), and two uncharacterized variants were also detected. Hb Hope interfered with HbA1c measurements, and all CSA<sub>2</sub>ABart'sH samples produced falsely low HbA1c values. Artificially decreased HbA1c was detected in 21/27 (77.8%) HbEE samples. ARKRAY HA-8180V FM analysis yielded reproducible RTs for HbE, achieving 100% sensitivity, specificity, and accuracy in the validation cohort.</p><p><strong>Conclusions: </strong>HbE and thalassemia variants can interfere with HbA1c quantification. ARKRAY HA-8180V FM chromatogram review allows reliable detection of HbE and prevents misinterpretation of HbA1c results.</p>","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-08-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2026.0183","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Hb variants are often incidentally detected during HbA1c testing and can cause interference. We evaluated the ability of the ARKRAY HA-8180V HPLC system operating in fast mode (FM) to identify Hb variants, particularly HbE, in samples submitted for HbA1c testing based on characteristic retention times (RTs).
Methods: Chromatograms of samples tested for HbA1c were analyzed; those showing predefined abnormal chromatographic patterns suggestive of Hb variants were further evaluated using capillary electrophoresis (CE) and β- and α-globin gene sequencing. RTs were recorded.
Results: Of 4,888 chromatograms analyzed, 588 (12.0%) were abnormal. Among these, 564 (95.9%) showed an abnormal peak at an RT of 23-25 s and were confirmed to contain HbE via CE. Identified Hb phenotypes included EA (N=526), compound heterozygotes [Constant Spring (CS) EA, N=2; CSEE, N=1], EE (N=27), EFA (N=3), and EF (N=5). Other variants included Hb CSA2ABart'sH, A2FA, A2FABart's, and A2F. Rare variants, Hb Hope (N=12) and Hb Korle-Bu (N=1), and two uncharacterized variants were also detected. Hb Hope interfered with HbA1c measurements, and all CSA2ABart'sH samples produced falsely low HbA1c values. Artificially decreased HbA1c was detected in 21/27 (77.8%) HbEE samples. ARKRAY HA-8180V FM analysis yielded reproducible RTs for HbE, achieving 100% sensitivity, specificity, and accuracy in the validation cohort.
Conclusions: HbE and thalassemia variants can interfere with HbA1c quantification. ARKRAY HA-8180V FM chromatogram review allows reliable detection of HbE and prevents misinterpretation of HbA1c results.