Advancing endodontic imaging: a high-resolution comparison between CBCT and micro-CT in the danger zone.

Biomaterial investigations in dentistry Pub Date : 2026-08-05 eCollection Date: 2026-01-01 DOI:10.2340/biid.v13.46618
Gabriel de Toledo Telles-Araujo, Raimundo Sales de Oliveira Neto, Mariana Quirino Silveira Soares, Heitor Marques Honório, Viviane Almeida Sarmento, Marco Antônio Hungaro Duarte, Izabel Regina Fischer Rubira-Bullen
{"title":"Advancing endodontic imaging: a high-resolution comparison between CBCT and micro-CT in the danger zone.","authors":"Gabriel de Toledo Telles-Araujo, Raimundo Sales de Oliveira Neto, Mariana Quirino Silveira Soares, Heitor Marques Honório, Viviane Almeida Sarmento, Marco Antônio Hungaro Duarte, Izabel Regina Fischer Rubira-Bullen","doi":"10.2340/biid.v13.46618","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study evaluated the accuracy of cone-beam computed tomography (CBCT) protocols for measuring submillimetric dentin thickness in the danger zone (DZ) of mandibular molars, using micro-computed tomography (micro-CT) as the reference standard.</p><p><strong>Materials and methods: </strong>Fifty-one extracted mandibular molars were scanned using micro-CT and two CBCT devices under different acquisition protocols (Standard, High Fidelity, High Resolution [HIRE], and Imaging System [i-CAT]). After three-dimensional image registration, dentin thickness in the DZ, located 2 mm below the furcation, was measured in both mesial canals.</p><p><strong>Results: </strong>All CBCT protocols significantly underestimated dentin thickness compared with micro-CT (<i>p</i> < 0.05), with discrepancies ranging from 4.5 to 8.6%. Underestimation was more pronounced in the mesiolingual canal. The HIRE protocol (0.125 mm voxel size) showed the smallest discrepancy, whereas the i-CAT protocol showed the largest.</p><p><strong>Conclusions: </strong>CBCT systematically underestimates dentin thickness in the DZ and is not metrically equivalent to micro-CT for submillimetric measurements. Measurement accuracy is influenced by acquisition parameters and cannot be explained by voxel size alone.</p>","PeriodicalId":72378,"journal":{"name":"Biomaterial investigations in dentistry","volume":"13 ","pages":"46618"},"PeriodicalIF":0.0000,"publicationDate":"2026-08-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13454917/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomaterial investigations in dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2340/biid.v13.46618","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: This study evaluated the accuracy of cone-beam computed tomography (CBCT) protocols for measuring submillimetric dentin thickness in the danger zone (DZ) of mandibular molars, using micro-computed tomography (micro-CT) as the reference standard.

Materials and methods: Fifty-one extracted mandibular molars were scanned using micro-CT and two CBCT devices under different acquisition protocols (Standard, High Fidelity, High Resolution [HIRE], and Imaging System [i-CAT]). After three-dimensional image registration, dentin thickness in the DZ, located 2 mm below the furcation, was measured in both mesial canals.

Results: All CBCT protocols significantly underestimated dentin thickness compared with micro-CT (p < 0.05), with discrepancies ranging from 4.5 to 8.6%. Underestimation was more pronounced in the mesiolingual canal. The HIRE protocol (0.125 mm voxel size) showed the smallest discrepancy, whereas the i-CAT protocol showed the largest.

Conclusions: CBCT systematically underestimates dentin thickness in the DZ and is not metrically equivalent to micro-CT for submillimetric measurements. Measurement accuracy is influenced by acquisition parameters and cannot be explained by voxel size alone.

牙髓成像的进步:危险区域CBCT和micro-CT的高分辨率比较。
目的:以微ct (micro-computed tomography, micro-CT)为参考标准,评价锥束ct (cone-beam computed tomography, CBCT)测量下颌磨牙危险区亚毫米牙本质厚度的准确性。材料和方法:使用micro-CT和两种CBCT设备在不同采集方案(标准、高保真、高分辨率[HIRE]和成像系统[i-CAT])下对51颗拔出的下颌磨牙进行扫描。三维图像配准后,测量了位于分叉下2mm的DZ内两个中内根管的牙本质厚度。结果:与micro-CT相比,所有CBCT方案均显著低估了牙本质厚度(p < 0.05),差异范围为4.5%至8.6%。低估在中舌管中更为明显。HIRE协议(0.125 mm体素尺寸)的差异最小,而i-CAT协议的差异最大。结论:CBCT系统地低估了DZ的牙本质厚度,并且在亚毫米测量上与微ct不等效。测量精度受采集参数的影响,不能单独用体素大小来解释。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
14 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书