Application of negative pressure to the periocular microenvironment to lower IOP in patients implanted with telemetric IOP sensor: report of two early studies of the FSYX ocular pressure adjusting pump.

IF 2.4
Frontiers in ophthalmology Pub Date : 2026-07-27 eCollection Date: 2026-01-01 DOI:10.3389/fopht.2026.1887910
Michael D Greenwood, Tanner J Ferguson, Lars Choritz, Thomas W Samuelson, Jamie J Beckman, H Burkhard Dick, Iqbal K Ahmed, John P Berdahl
{"title":"Application of negative pressure to the periocular microenvironment to lower IOP in patients implanted with telemetric IOP sensor: report of two early studies of the FSYX ocular pressure adjusting pump.","authors":"Michael D Greenwood, Tanner J Ferguson, Lars Choritz, Thomas W Samuelson, Jamie J Beckman, H Burkhard Dick, Iqbal K Ahmed, John P Berdahl","doi":"10.3389/fopht.2026.1887910","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The purpose of this study was to determine the effect on intraocular pressure (IOP) of applying negative pressure (NP) to the periocular space, as measured by an intraocular IOP sensor in human eyes.</p><p><strong>Methods: </strong>Eyes with open-angle glaucoma (OAG) and previously implanted with the eyemate<sup>®</sup>-IO IOP-measuring device (Implandata Ophthalmic Products GmbH (now G-Metrics GmbH)) wore an early prototype of the FSYX™ Ocular Pressure Adjusting Pump (OPAP; Balance Ophthalmics, Inc.) in two separate studies: \"Eyemate 1\" and \"Eyemate 2.\" In Eyemate 1, study eyes received periorbital NP application at -10mmHg with OPAP for 70 to 110 minutes vs no NP in the fellow control eye. In Eyemate 2, eyes wore OPAP for 8 hours of continuous NP application set to 50% of the eye's baseline IOP vs no NP in the fellow control eye. In both studies, IOP was measured in real time using the implanted eyemate-IO device. The primary effectiveness endpoint in both studies was the change from baseline IOP in the study eye, as measured with eyemate-IO.</p><p><strong>Results: </strong>All seven study eyes demonstrated immediate and sustained reductions in IOP during NP application. In Eyemate 1, three study eyes had individual mean steady-state IOP reductions of 31%, 51%, and 23% from baseline. In Eyemate 2, four study eyes had a mean 8-hour IOP reduction of 20% from baseline. All patients tolerated goggle wear and NP application.</p><p><strong>Conclusion: </strong>These studies provided early proof-of-concept evidence that NP application to the periocular space lowers IOP, as measured by an implanted intraocular sensor, and was well tolerated over short-term use. Given the small, uncontrolled or open-label designs, these findings are hypothesis-generating; larger, controlled studies are required to confirm efficacy and characterize safety.</p>","PeriodicalId":73096,"journal":{"name":"Frontiers in ophthalmology","volume":"6 ","pages":"1887910"},"PeriodicalIF":2.4000,"publicationDate":"2026-07-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13453810/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fopht.2026.1887910","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: The purpose of this study was to determine the effect on intraocular pressure (IOP) of applying negative pressure (NP) to the periocular space, as measured by an intraocular IOP sensor in human eyes.

Methods: Eyes with open-angle glaucoma (OAG) and previously implanted with the eyemate®-IO IOP-measuring device (Implandata Ophthalmic Products GmbH (now G-Metrics GmbH)) wore an early prototype of the FSYX™ Ocular Pressure Adjusting Pump (OPAP; Balance Ophthalmics, Inc.) in two separate studies: "Eyemate 1" and "Eyemate 2." In Eyemate 1, study eyes received periorbital NP application at -10mmHg with OPAP for 70 to 110 minutes vs no NP in the fellow control eye. In Eyemate 2, eyes wore OPAP for 8 hours of continuous NP application set to 50% of the eye's baseline IOP vs no NP in the fellow control eye. In both studies, IOP was measured in real time using the implanted eyemate-IO device. The primary effectiveness endpoint in both studies was the change from baseline IOP in the study eye, as measured with eyemate-IO.

Results: All seven study eyes demonstrated immediate and sustained reductions in IOP during NP application. In Eyemate 1, three study eyes had individual mean steady-state IOP reductions of 31%, 51%, and 23% from baseline. In Eyemate 2, four study eyes had a mean 8-hour IOP reduction of 20% from baseline. All patients tolerated goggle wear and NP application.

Conclusion: These studies provided early proof-of-concept evidence that NP application to the periocular space lowers IOP, as measured by an implanted intraocular sensor, and was well tolerated over short-term use. Given the small, uncontrolled or open-label designs, these findings are hypothesis-generating; larger, controlled studies are required to confirm efficacy and characterize safety.

应用眼周微环境负压降低眼压传感器植入术患者的IOP: FSYX眼压调节泵的两项早期研究报告
简介:本研究的目的是通过人眼眼内压传感器测量眼周空间施加负压(NP)对眼内压(IOP)的影响。方法:开角型青光眼(OAG)患者的眼睛先前植入了eyeemate®-IO眼压测量装置(Implandata Ophthalmic Products GmbH(现G-Metrics GmbH)),在两项独立的研究中佩戴了FSYX™眼压调节泵(OPAP; Balance Ophthalmic, Inc.)的早期原型:“eyemate 1”和“eyemate 2”。在Eyemate 1中,研究眼接受-10mmHg的眶周NP应用和OPAP,持续70 - 110分钟,而另一只对照眼不接受NP。在eyeemate 2中,眼睛连续佩戴OPAP 8小时,NP应用设置为眼睛基线IOP的50%,而另一只对照眼没有NP。在这两项研究中,眼压都是使用植入的眼眼- io装置实时测量的。两项研究的主要有效性终点都是研究眼的基线IOP变化,用eyemate-IO测量。结果:在NP应用期间,所有7只研究眼睛的IOP均出现了立即和持续的下降。在Eyemate 1中,三个研究眼的平均稳态IOP分别比基线降低了31%、51%和23%。在Eyemate 2中,4只研究眼睛的IOP比基线平均降低了20%。所有患者都能耐受护目镜佩戴和NP应用。结论:这些研究提供了早期的概念证明,通过植入眼内传感器测量,NP应用于眼周间隙可降低IOP,并且短期使用耐受性良好。考虑到小型、非受控或开放标签设计,这些发现是假设生成的;需要更大规模的对照研究来确认有效性和安全性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.50
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书