Giorgio Russo, Santa Gabriella Bonaccorsi, Renata Leanza, Maria Costa, Gaia Pucci, Cristiana Alberghina, Francesco Paolo Cammarata, Giovanni Li Volti, Rosalba Parenti
{"title":"Radiation protection strategies for FLASH radiotherapy accelerator installation in animal housing.","authors":"Giorgio Russo, Santa Gabriella Bonaccorsi, Renata Leanza, Maria Costa, Gaia Pucci, Cristiana Alberghina, Francesco Paolo Cammarata, Giovanni Li Volti, Rosalba Parenti","doi":"10.1093/rpd/ncag079","DOIUrl":null,"url":null,"abstract":"<p><p>This study presents a radioprotection assessment for an ultra-high dose rate accelerator installation in one of the animal housing of the Centre for Advanced Preclinical in vivo Research. The main problem concerns the absence of a bunker while limiting exposure in nearby work areas. The geometry of the accelerator, the building layouts and the distribution of the environmental dose equivalent were simulated with FLUKA.CERN 4-4.1, modeling a 9 MeV electron beam. The results confirmed compliance with the limit of 0.250 mSv/year for unrestricted areas during daytime operation. Night-time operating zone restrictions apply when no staff are present, allowing beam on for quality controls. The effective doses to the population, researchers and workers are less than 1 mSv/year, in accordance with Legislative Decree 101/2020; the accelerator operators are classified as exposed. For small animals, precautionary dose limit of 1 mSv/year was adopted. The use of lead shielding and site-specific safety measures ensure radiological compliance.</p>","PeriodicalId":20795,"journal":{"name":"Radiation protection dosimetry","volume":" ","pages":""},"PeriodicalIF":0.8000,"publicationDate":"2026-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation protection dosimetry","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncag079","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
This study presents a radioprotection assessment for an ultra-high dose rate accelerator installation in one of the animal housing of the Centre for Advanced Preclinical in vivo Research. The main problem concerns the absence of a bunker while limiting exposure in nearby work areas. The geometry of the accelerator, the building layouts and the distribution of the environmental dose equivalent were simulated with FLUKA.CERN 4-4.1, modeling a 9 MeV electron beam. The results confirmed compliance with the limit of 0.250 mSv/year for unrestricted areas during daytime operation. Night-time operating zone restrictions apply when no staff are present, allowing beam on for quality controls. The effective doses to the population, researchers and workers are less than 1 mSv/year, in accordance with Legislative Decree 101/2020; the accelerator operators are classified as exposed. For small animals, precautionary dose limit of 1 mSv/year was adopted. The use of lead shielding and site-specific safety measures ensure radiological compliance.
期刊介绍:
Radiation Protection Dosimetry covers all aspects of personal and environmental dosimetry and monitoring, for both ionising and non-ionising radiations. This includes biological aspects, physical concepts, biophysical dosimetry, external and internal personal dosimetry and monitoring, environmental and workplace monitoring, accident dosimetry, and dosimetry related to the protection of patients. Particular emphasis is placed on papers covering the fundamentals of dosimetry; units, radiation quantities and conversion factors. Papers covering archaeological dating are included only if the fundamental measurement method or technique, such as thermoluminescence, has direct application to personal dosimetry measurements. Papers covering the dosimetric aspects of radon or other naturally occurring radioactive materials and low level radiation are included. Animal experiments and ecological sample measurements are not included unless there is a significant relevant content reason.