Serial dual-wavelength illumination for retinal vessel oximetry using a conventional fundus camera: a proof-of-concept study.

IF 2.1 4区 医学 Q3 OPHTHALMOLOGY
Melih Tarhan, Daniel Meller, Martin Hammer
{"title":"Serial dual-wavelength illumination for retinal vessel oximetry using a conventional fundus camera: a proof-of-concept study.","authors":"Melih Tarhan, Daniel Meller, Martin Hammer","doi":"10.1186/s12886-026-05216-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Retinal vessel oximetry requires fundus images at two wavelengths with different hemoglobin absorption characteristics. Simultaneous acquisition using one-chip color cameras may be affected by incomplete spectral separation of detector channels, whereas multi-chip systems such as 3-chip CCD cameras can achieve improved spectral separation. We present a serial dual-wavelength illumination approach integrated into a conventional fundus camera and compare oxygen saturation measurements with those obtained from a commercial retinal oximeter.</p><p><strong>Methods: </strong>A custom LED illumination unit was fiber-coupled into a fundus camera to provide serial illumination at 548 nm (isosbestic hemoglobin wavelength) and 605 nm (oxygen-sensitive wavelength with differential absorption between oxy- and deoxyhemoglobin), each for 250 ms. Corneal irradiance was 33 mW/cm² at 548 nm and 7 mW/cm² at 605 nm. Sequentially acquired images were registered and analyzed using established retinal oximetry algorithms. Mean arterial and venous oxygen saturation values were compared with measurements from a commercial retinal oximeter in nine participants undergoing routine ophthalmic examination. Linear regression analysis was performed to assess the association between methods, and agreement was additionally evaluated using Bland-Altman analysis.</p><p><strong>Results: </strong>Mean arterial oxygen saturation was 99.4 ± 5.7% with serial illumination and 100.4 ± 3.0% with standard oximetry. Mean venous oxygen saturation was 71.5 ± 7.0% and 69.5 ± 11.6%, respectively. The mean absolute differences between both techniques were 3.6 ± 2.7% for arteries and 8.4 ± 9.3% for veins. Bland-Altman analysis showed a small arterial bias of 0.94% with limits of agreement from - 7.9% to + 9.8%. For venous measurements, bias was - 1.98% with wider limits of agreement (- 26.9% to + 22.9%). Arterial measurements showed borderline non-significant correlation (R = 0.606, p = 0.084). Venous measurements showed no correlation.</p><p><strong>Conclusions: </strong>Serial illumination is feasible for retinal vessel oximetry and mitigates detector-level spectral crosstalk associated with simultaneous dual-wavelength imaging using one-chip color cameras. Arterial measurements showed only a small systematic bias compared with the commercial retinal oximeter, whereas venous measurements demonstrated greater variability, indicating the need for further optimization and validation.</p>","PeriodicalId":9058,"journal":{"name":"BMC Ophthalmology","volume":"26 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2026-08-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13455296/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s12886-026-05216-7","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Retinal vessel oximetry requires fundus images at two wavelengths with different hemoglobin absorption characteristics. Simultaneous acquisition using one-chip color cameras may be affected by incomplete spectral separation of detector channels, whereas multi-chip systems such as 3-chip CCD cameras can achieve improved spectral separation. We present a serial dual-wavelength illumination approach integrated into a conventional fundus camera and compare oxygen saturation measurements with those obtained from a commercial retinal oximeter.

Methods: A custom LED illumination unit was fiber-coupled into a fundus camera to provide serial illumination at 548 nm (isosbestic hemoglobin wavelength) and 605 nm (oxygen-sensitive wavelength with differential absorption between oxy- and deoxyhemoglobin), each for 250 ms. Corneal irradiance was 33 mW/cm² at 548 nm and 7 mW/cm² at 605 nm. Sequentially acquired images were registered and analyzed using established retinal oximetry algorithms. Mean arterial and venous oxygen saturation values were compared with measurements from a commercial retinal oximeter in nine participants undergoing routine ophthalmic examination. Linear regression analysis was performed to assess the association between methods, and agreement was additionally evaluated using Bland-Altman analysis.

Results: Mean arterial oxygen saturation was 99.4 ± 5.7% with serial illumination and 100.4 ± 3.0% with standard oximetry. Mean venous oxygen saturation was 71.5 ± 7.0% and 69.5 ± 11.6%, respectively. The mean absolute differences between both techniques were 3.6 ± 2.7% for arteries and 8.4 ± 9.3% for veins. Bland-Altman analysis showed a small arterial bias of 0.94% with limits of agreement from - 7.9% to + 9.8%. For venous measurements, bias was - 1.98% with wider limits of agreement (- 26.9% to + 22.9%). Arterial measurements showed borderline non-significant correlation (R = 0.606, p = 0.084). Venous measurements showed no correlation.

Conclusions: Serial illumination is feasible for retinal vessel oximetry and mitigates detector-level spectral crosstalk associated with simultaneous dual-wavelength imaging using one-chip color cameras. Arterial measurements showed only a small systematic bias compared with the commercial retinal oximeter, whereas venous measurements demonstrated greater variability, indicating the need for further optimization and validation.

使用传统眼底相机进行视网膜血管血氧测量的串行双波长照明:概念验证研究。
目的:视网膜血管血氧测定需要两个波长的眼底图像,具有不同的血红蛋白吸收特征。单片彩色相机同时采集可能会受到探测器通道光谱分离不完全的影响,而多片系统如3片CCD相机可以实现更好的光谱分离。我们提出了一种集成到传统眼底相机中的串行双波长照明方法,并将氧饱和度测量结果与商业视网膜血氧计获得的结果进行了比较。方法:将定制的LED照明单元光纤耦合到眼底相机上,在548 nm(等吸收血红蛋白波长)和605 nm(氧敏波长,氧敏和脱氧血红蛋白吸收差异)波长上连续照明,每次照明250 ms。角膜辐照度在548 nm为33 mW/cm²,在605 nm为7 mW/cm²。使用已建立的视网膜血氧仪算法对顺序获取的图像进行配准和分析。对9名接受常规眼科检查的参与者的平均动脉和静脉氧饱和度值与商用视网膜氧饱和度仪的测量值进行比较。采用线性回归分析来评估方法之间的相关性,并使用Bland-Altman分析进一步评估一致性。结果:连续照明平均动脉血氧饱和度为99.4±5.7%,标准血氧饱和度为100.4±3.0%。平均静脉血氧饱和度分别为71.5±7.0%和69.5±11.6%。两种方法的平均绝对差异分别为动脉(3.6±2.7%)和静脉(8.4±9.3%)。Bland-Altman分析显示动脉偏差较小,为0.94%,一致性范围为- 7.9%至+ 9.8%。对于静脉测量,偏差为- 1.98%,一致性范围更广(- 26.9%至+ 22.9%)。动脉测量值呈临界无显著相关性(R = 0.606, p = 0.084)。静脉测量结果无相关性。结论:连续照明在视网膜血管血氧测量中是可行的,并且减轻了使用单片彩色相机同时进行双波长成像所带来的探测器级光谱串扰。与商用视网膜血氧计相比,动脉测量仅显示出很小的系统偏差,而静脉测量显示出更大的变异性,表明需要进一步优化和验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
BMC Ophthalmology
BMC Ophthalmology OPHTHALMOLOGY-
CiteScore
3.40
自引率
5.00%
发文量
441
审稿时长
6-12 weeks
期刊介绍: BMC Ophthalmology is an open access, peer-reviewed journal that considers articles on all aspects of the prevention, diagnosis and management of eye disorders, as well as related molecular genetics, pathophysiology, and epidemiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书