A Comprehensive Numerical Simulation on the Safety and Efficacy of Circular-Versus Trapezoidal-Shaped Intravascular Drug-Eluting Stents

IF 2.1 4区 医学 Q3 ENGINEERING, BIOMEDICAL
Ramprosad Saha, Akash Pradip Mandal
{"title":"A Comprehensive Numerical Simulation on the Safety and Efficacy of Circular-Versus Trapezoidal-Shaped Intravascular Drug-Eluting Stents","authors":"Ramprosad Saha,&nbsp;Akash Pradip Mandal","doi":"10.1002/cnm.70205","DOIUrl":null,"url":null,"abstract":"<div>\n \n <p>In modern days, drug-eluting stent (DES) acts as a kingpin for the treatment of coronary artery disease (CAD) and it has a dramatic reduction rate of in-stent restenosis (ISR) in compare to a bare metal stent (BMS). To replicate the drug release from therapeutic devices and its corresponding transportation in the physiological atmosphere, mathematical and computational simulations have become an effective technique. The present article is developed by a thorough comparative analysis of drug transport mechanisms between half-embedded trapezoidal-shaped and circular-shaped struts and also its optimality in stent-based drug delivery. The geometry of the implanted struts (trapezoidal-shaped and circular-shaped) is modeled in a two-dimensional axi-symmetric environment and the target lesion is considered as a single homogeneous layer with identical diffusivity. Due to the hydrostatic pressure of blood, plasma filtration is allowed through the blood-tissue interface along the transmural direction and the flow of interstitial fluid within the porous arterial wall is governed by the unsteady Navier–Stokes equation and the equation of continuity. While the drug is distributed within the arterial wall, tissue receptors grip the drug molecules, so the present study includes the binding of drug along with free drug. An unsteady convection-diffusion–reaction process demonstrates the transportation of free drug, but only reaction process manifests the transportation of bound drug. The mathematical equations of interstitial fluid flow and the transportation of drug along with pertinent initial and boundary conditions are penciled by using a two-dimensional (2D) cylindrical polar coordinate system. A staggered grid generation technique is also leveraged to discretize all the governing equations and boundary conditions, which are then successfully solved numerically by using Marker-and-Cell (MAC) method. The coefficient of variance (CV), a statistical parameter, is introduced in this present drug delivery system to access the consistency of the drug distribution. The study incorporates several necessary factors, such as drug efficacy, tissue drug content, and therapeutic effectiveness to optimize the choice of strut shape and the overall performance of the stent. To achieve the best possible outcomes, a robust sensitivity analysis of several perturbed parameters is carried out by implementing one-way ANOVA. The results highlight that, for long-term safety and efficacy, trapezoidal-shaped struts may be a good choice in compared to circular ones. Furthermore, the tailored combination of strut shape and drug delivery strategy presented herein offers a significant advancement in designing the next-generation DES.</p>\n </div>","PeriodicalId":50349,"journal":{"name":"International Journal for Numerical Methods in Biomedical Engineering","volume":"42 8","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2026-08-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal for Numerical Methods in Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cnm.70205","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

In modern days, drug-eluting stent (DES) acts as a kingpin for the treatment of coronary artery disease (CAD) and it has a dramatic reduction rate of in-stent restenosis (ISR) in compare to a bare metal stent (BMS). To replicate the drug release from therapeutic devices and its corresponding transportation in the physiological atmosphere, mathematical and computational simulations have become an effective technique. The present article is developed by a thorough comparative analysis of drug transport mechanisms between half-embedded trapezoidal-shaped and circular-shaped struts and also its optimality in stent-based drug delivery. The geometry of the implanted struts (trapezoidal-shaped and circular-shaped) is modeled in a two-dimensional axi-symmetric environment and the target lesion is considered as a single homogeneous layer with identical diffusivity. Due to the hydrostatic pressure of blood, plasma filtration is allowed through the blood-tissue interface along the transmural direction and the flow of interstitial fluid within the porous arterial wall is governed by the unsteady Navier–Stokes equation and the equation of continuity. While the drug is distributed within the arterial wall, tissue receptors grip the drug molecules, so the present study includes the binding of drug along with free drug. An unsteady convection-diffusion–reaction process demonstrates the transportation of free drug, but only reaction process manifests the transportation of bound drug. The mathematical equations of interstitial fluid flow and the transportation of drug along with pertinent initial and boundary conditions are penciled by using a two-dimensional (2D) cylindrical polar coordinate system. A staggered grid generation technique is also leveraged to discretize all the governing equations and boundary conditions, which are then successfully solved numerically by using Marker-and-Cell (MAC) method. The coefficient of variance (CV), a statistical parameter, is introduced in this present drug delivery system to access the consistency of the drug distribution. The study incorporates several necessary factors, such as drug efficacy, tissue drug content, and therapeutic effectiveness to optimize the choice of strut shape and the overall performance of the stent. To achieve the best possible outcomes, a robust sensitivity analysis of several perturbed parameters is carried out by implementing one-way ANOVA. The results highlight that, for long-term safety and efficacy, trapezoidal-shaped struts may be a good choice in compared to circular ones. Furthermore, the tailored combination of strut shape and drug delivery strategy presented herein offers a significant advancement in designing the next-generation DES.

圆形与梯形血管内药物洗脱支架安全性和有效性的综合数值模拟。
在现代,药物洗脱支架(DES)作为治疗冠状动脉疾病(CAD)的主要手段,与裸金属支架(BMS)相比,它具有显着降低支架内再狭窄(ISR)的比率。为了复制药物从治疗装置中释放及其在生理气氛中的相应运输,数学和计算模拟已成为一种有效的技术。本文是通过对半嵌入梯形和圆形支杆的药物输送机制及其在支架药物输送中的最优性进行全面的比较分析而发展起来的。植入支柱(梯形和圆形)的几何形状在二维轴对称环境中建模,并将目标病变视为具有相同扩散率的单一均匀层。由于血液的静水压力,允许血浆沿跨壁方向通过血组织界面,多孔动脉壁内的间质液流动受非定常Navier-Stokes方程和连续性方程控制。当药物在动脉壁内分布时,组织受体抓住药物分子,因此本研究包括药物与游离药物的结合。非定常的对流-扩散-反应过程体现了游离药物的转运,而只有反应过程才能体现结合药物的转运。利用二维圆柱极坐标系,建立了间隙流体流动和药物输运的数学方程,并给出了相应的初始条件和边界条件。利用交错网格生成技术对所有控制方程和边界条件进行离散化,然后利用标记-单元(MAC)方法成功地进行数值求解。在给药系统中引入方差系数(CV)这一统计参数来评价给药系统中药物分布的一致性。本研究结合药物疗效、组织药物含量、治疗效果等几个必要因素,优化支架形状的选择和支架的整体性能。为了获得最好的结果,通过实施单向方差分析,对几个扰动参数进行了稳健的敏感性分析。结果表明,从长期安全性和有效性来看,梯形支撑可能是较好的选择,而不是圆形支撑。此外,本文提出的支架形状和药物递送策略的定制组合为设计下一代DES提供了重要的进步。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal for Numerical Methods in Biomedical Engineering
International Journal for Numerical Methods in Biomedical Engineering ENGINEERING, BIOMEDICAL-MATHEMATICAL & COMPUTATIONAL BIOLOGY
CiteScore
4.50
自引率
9.50%
发文量
103
审稿时长
3 months
期刊介绍: All differential equation based models for biomedical applications and their novel solutions (using either established numerical methods such as finite difference, finite element and finite volume methods or new numerical methods) are within the scope of this journal. Manuscripts with experimental and analytical themes are also welcome if a component of the paper deals with numerical methods. Special cases that may not involve differential equations such as image processing, meshing and artificial intelligence are within the scope. Any research that is broadly linked to the wellbeing of the human body, either directly or indirectly, is also within the scope of this journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书