Comparison of TISAB II and III for salivary fluoride determination: addressing a bioanalytical challenge in fluoride analysis.

IF 2.1 4区 医学 Q3 BIOCHEMICAL RESEARCH METHODS
Bioanalysis Pub Date : 2026-07-01 Epub Date: 2026-08-01 DOI:10.1080/17576180.2026.2712771
Flavia Almeida Amorim, Maria Suzi de Sousa Lopes, Lyzia Vitoria Mendes Rezende, Evanildo Canuto Paz, Marcus Vinícius Oliveira, Glauber Campos Vale
{"title":"Comparison of TISAB II and III for salivary fluoride determination: addressing a bioanalytical challenge in fluoride analysis.","authors":"Flavia Almeida Amorim, Maria Suzi de Sousa Lopes, Lyzia Vitoria Mendes Rezende, Evanildo Canuto Paz, Marcus Vinícius Oliveira, Glauber Campos Vale","doi":"10.1080/17576180.2026.2712771","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Accurate determination of salivary fluoride (F) depends on appropriate analytical conditions, particularly the buffering system used before ion-selective electrode measurements.</p><p><strong>Research design and methods: </strong>This laboratory-based analytical study addressed the practical challenge of selecting between two total ionic strength adjustment buffers (TISAB II and TISAB III) for salivary F-determination using samples obtained from previously conducted clinical trials. Unstimulated saliva from eight participants was collected at baseline and at 5 min, 30 min, 1 h, 4 h, and 12 h after exposure to three F-treatments: F-varnish, 1,450 ppm F-dentifrice, and 5,000 ppm F-dentifrice. F-concentrations were measured using an F-ion-selective electrode after buffering with either TISAB II or TISAB III. Data were analyzed using paired statistical tests and Pearson correlation analysis (α = 0.05).</p><p><strong>Results: </strong>No significant differences were observed between the buffers across most conditions, although differences were detected at later time points following dentifrice use. A strong positive correlation was found between the two buffering systems (<i>p</i> < 0.001), and no differences were detected in the area under the concentration-time curve.</p><p><strong>Conclusions: </strong>These findings indicate that both TISAB II and TISAB III provide comparable analytical performance for salivary fluoride determination under the tested conditions, although differences may occur at lower fluoride concentrations.</p>","PeriodicalId":8797,"journal":{"name":"Bioanalysis","volume":" ","pages":"955-960"},"PeriodicalIF":2.1000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13540148/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioanalysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/17576180.2026.2712771","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/1 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Accurate determination of salivary fluoride (F) depends on appropriate analytical conditions, particularly the buffering system used before ion-selective electrode measurements.

Research design and methods: This laboratory-based analytical study addressed the practical challenge of selecting between two total ionic strength adjustment buffers (TISAB II and TISAB III) for salivary F-determination using samples obtained from previously conducted clinical trials. Unstimulated saliva from eight participants was collected at baseline and at 5 min, 30 min, 1 h, 4 h, and 12 h after exposure to three F-treatments: F-varnish, 1,450 ppm F-dentifrice, and 5,000 ppm F-dentifrice. F-concentrations were measured using an F-ion-selective electrode after buffering with either TISAB II or TISAB III. Data were analyzed using paired statistical tests and Pearson correlation analysis (α = 0.05).

Results: No significant differences were observed between the buffers across most conditions, although differences were detected at later time points following dentifrice use. A strong positive correlation was found between the two buffering systems (p < 0.001), and no differences were detected in the area under the concentration-time curve.

Conclusions: These findings indicate that both TISAB II and TISAB III provide comparable analytical performance for salivary fluoride determination under the tested conditions, although differences may occur at lower fluoride concentrations.

唾液氟化物测定TISAB II和III的比较:解决氟化物分析中的生物分析挑战。
背景:唾液氟化物(F)的准确测定取决于适当的分析条件,特别是离子选择电极测量前使用的缓冲系统。研究设计和方法:这项基于实验室的分析研究解决了在两种总离子强度调节缓冲液(TISAB II和TISAB III)之间进行选择的实际挑战,该缓冲液使用先前进行的临床试验中获得的样品进行唾液f测定。8名参与者在基线和暴露于三种f处理后5分钟、30分钟、1小时、4小时和12小时收集未受刺激的唾液:f清漆、1,450 ppm的f -牙膏和5,000 ppm的f -牙膏。用TISAB II或TISAB III缓冲后,使用f离子选择电极测量f浓度。资料采用配对统计检验和Pearson相关分析(α = 0.05)。结果:在大多数情况下,缓冲液之间没有观察到显著差异,尽管在使用牙膏后的较晚时间点检测到差异。结论:这些发现表明,TISAB II和TISAB III在测试条件下对唾液氟化物测定的分析性能相当,尽管在较低的氟化物浓度下可能会出现差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Bioanalysis
Bioanalysis BIOCHEMICAL RESEARCH METHODS-CHEMISTRY, ANALYTICAL
CiteScore
3.30
自引率
16.70%
发文量
88
审稿时长
2 months
期刊介绍: Reliable data obtained from selective, sensitive and reproducible analysis of xenobiotics and biotics in biological samples is a fundamental and crucial part of every successful drug development program. The same principles can also apply to many other areas of research such as forensic science, toxicology and sports doping testing. The bioanalytical field incorporates sophisticated techniques linking sample preparation and advanced separations with MS and NMR detection systems, automation and robotics. Standards set by regulatory bodies regarding method development and validation increasingly define the boundaries between speed and quality. Bioanalysis is a progressive discipline for which the future holds many exciting opportunities to further reduce sample volumes, analysis cost and environmental impact, as well as to improve sensitivity, specificity, accuracy, efficiency, assay throughput, data quality, data handling and processing. The journal Bioanalysis focuses on the techniques and methods used for the detection or quantitative study of analytes in human or animal biological samples. Bioanalysis encourages the submission of articles describing forward-looking applications, including biosensors, microfluidics, miniaturized analytical devices, and new hyphenated and multi-dimensional techniques. Bioanalysis delivers essential information in concise, at-a-glance article formats. Key advances in the field are reported and analyzed by international experts, providing an authoritative but accessible forum for the modern bioanalyst.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书